EVS-TCB User Manual

Page 27

Temperature Controlled Bath – USER MANUAL

Check list on Start Up

Wear PPE, including safety glass and gloves when topping up the bath fluid and coolant

Step 1

Check BATH SET POINT – temperature is set to required setpoint

Yes/No (tick)

2

Toggle OFF the chiller to speed up the bath warm up

(Optional)

3

Toggle ON the chiller when the bath temperature is close to the required set-point

(Optional)

4

Once the required bath temperature is reached, check the bath liquid level. If necessary top up with liquid.

5

Coolant level – unscrew the cap and top up with coolant (liquid level should be just below the threaded part on the neck of the reservoir)

6

The bath becomes STABLE (S) when it reaches the required temperature

7

Heating % is between 10 to 20% when the bath is stable

8

If more heat input is required refer to the section on regulating the Cooling power

POULTEN SELFE & LEE LTD

EVS_TCB_V1.04

26