1 minute read

Bath Fluids

Bath Fluids

In the case of silicone oil, carefully read the SDS and wear the appropriate PPE

Specified fluid

Silicone bath fluid, 10cSt Silicone bath fluid, 20cSt Water – soft, decalcified (filtered)*

Bath Fluids

Part Ref

20417/1 20418/1 n/a Temperature range

20 to 50°C 50 to 100°C 20 to 60°C

NOTICE Changes to the specified bath liquids not expressly agreed by Poulten Selfe & Lee Ltd may void the warranty

Water

* For use with manual viscometers only. Do not use with automated viscometer systems. The quality of water depends upon local conditions. Only filtered water shall be used with the EVSTCB bath.

The following types of water should not be used: • Hard water – due to the high concentration of lime this will lead to the calcification in the bath

• Ferrous water – can cause corrosion

• Distilled or deionized water – special properties may cause corrosion in the bath

This article is from: