1 minute read
Water Treatment
Next Article
Water Treatment
The user should carry out their own risk assessment regarding using water as a bath fluid.
This option will allow the user to carry out a water treatment heating cycle to ensure that water does not stagnate which could provide an opportunity for Legionella bacteria to proliferate.
Select this option from the MAIN MENU.
The bath will automatically heat to 65°C and hold this temperature for one hour.
Toggle the Cool Mode OFF before starting this cycle.
NOTICE
When working with manual viscometers it may be preferable to use water as the bath medium if the Bath Set Point is 50°C or lower. Only use water as specified in the bath fluids section.