Temperature Controlled Bath – USER MANUAL
Bath Fluids In the case of silicone oil, carefully read the SDS and wear the appropriate PPE
Bath Fluids Specified fluid
Part Ref
Temperature range
Silicone bath fluid, 10cSt Silicone bath fluid, 20cSt Water – soft, decalcified (filtered)*
20417/1 20418/1 n/a
20 to 50°C 50 to 100°C 20 to 60°C
NOTICE
Changes to the specified bath liquids not expressly agreed by Poulten Selfe & Lee Ltd may void the warranty
Water * For use with manual viscometers only. Do not use with automated viscometer systems. The quality of water depends upon local conditions. Only filtered water shall be used with the EVSTCB bath. The following types of water should not be used: •
Hard water – due to the high concentration of lime this will lead to the calcification in the bath
•
Ferrous water – can cause corrosion
•
Distilled or deionized water – special properties may cause corrosion in the bath
POULTEN SELFE & LEE LTD
EVS_TCB_V1.04
11