RPV-3 software – USER MANUAL
Bath Temperature
Select System Configurations – bath settings
In normal operation, the bath temperature is automatically set by the Method. There is no need to use bath settings, unless the temperature offset needs to be reset, or a new bath temperature setpoint is required, prior to running aa new method.
The bath temperature settings are normally set as part of the Method Configuration. If a method is not assigned to the measurement positions, the bath temperature can still be set independently. For example, on initial start up the Bath settings can be used to set the bath temperature set-point (this can also be set via the keypad control on the bath itself). When a position has been enabled for measurement, the temperature in the preconfigured method will used. For example, this enables one position to be used at one temperature and another position at a different temperature.
Clear Offset – clears the offset (generated by a reference temperature) to 0.000
PSL RHEOTEK VERSION 2.03