RPV-3 software – USER MANUAL
Application configuration checks Note these checks will normally be performed by an installation or service engineer.
Parameter
Check:
Measuring head - settings
Measuring head voltages are set to 2.5V System Diagnostics – enter password – Sensor and Valves Note measuring head voltages can be re-trimmed from the front of the Viscometer Control module
Sample push up
Check push-up pressure: Activate capillary valve, turn on head pump and observe push up – adjust push up pressure to obtain a smooth rate. Check measuring head voltages detect meniscus on push up. System Diagnostics – enter password – Sensor and Valves
Ad-hoc clean
Ensure that a viscometer and a method is configured for each position. Carry out an ad-hoc clean – ensure tubes are clean and dry Home Page – Adhoc Clean
PSL RHEOTEK VERSION 2.03
12