INFECTION CONTROL SOLUTIONS

TESTING YOUR ULTRASONIC BATH CORRECTLY AND SAFELY

As part of our daily testing requirements as required by the AS/NZ 4815:2006 standard it specifies that daily performance testing e.g. foil test or pencil load test is required. The most common method we see in practice is the foil test, however, there seems to be varied techniques and types of foil being used. As a general rule you should always refer to the manufacture of the ultrasonic cleaner to determine how a foil test is to be performed. But also become familiar from a safety perspective how to handle and operate the machine and tests. Safety First! No part of your hands should be submerged into the cleaner during operation as the ultrasonic waves are enough to cause damage to joint tissues and could result in long-term arthritic conditions. The lid should also be closed in order to prevent excess emission of noise and aerosols. Lastly the product you’re using should be handled carefully and diluted correctly to prevent skin irritations and workplace health and safety concerns.

Step 1: Fill tank with tempered water

Step 2: Add recommended dose of solution

Step 3: Degas

Step 4: Run foil test or pencil load test

Step 5: Add basket and instruments

Step 6: Close lid

Step 7: Run Ultrasonic Cleaner

Step 8: Remove basket and set on unit to drain

Step 9: Rinse instruments off with water

Step 10: End of day– empty tank

Step 11: Rinse the tank and wipe clean with a soft cloth moistened with a mineral deposit remover and then with a mild disinfectant

aner Ultrasonic Cle de Reference Gui

2

1 Ultrasonic Clea Reference Guid ner e nded dose of

recomme Step 2: Add

with Step 1: Fill tank

tempered water

solution

3

Step 7: Run Ultrasonic Cleaner

foil test Step 4: Run

or pencil load

5

Step 3: Degas

4

7

8

test

6

Step 8: Remove basket

9

and set on unit

to drain

10

Step 9: Rinse instrument

s off with water

lid Step 6: Close basket and Step 5: Add

instruments

11

Step 10: End of day–

empty tank

6 0

Coltène/Whaled Raiffeisenstrasse ent GmbH + Co. KG 89129 Langenau 30 Tel. +49 (0)7345 / Germany Fax +49 (0)7345 805 0 info.de@coltene.c805 201 om

P/N31629A

Step 11: Rinse the tank and moistened with wipe clean with a soft cloth a then with a mild mineral deposit remover and disinfectant Coltène/Whaled ent Inc. 235 Ascot Parkway Cuyahoga Falls, OH 44223/USA Tel. USA and Canada +1 800 +1 330 916 221 3046 8800 Fax +1 330 645 8704 info.us@coltene.c www.coltene.com om

Scan here to view the full Coltene Ultrasonic Cleaner Reference Guide


Turn static files into dynamic content formats.

Create a flipbook
Issuu converts static files into: digital portfolios, online yearbooks, online catalogs, digital photo albums and more. Sign up and create your flipbook.
August - September Dental Solutions by Henry Schein ANZ - Issuu