Changing the Transmitter Settings continued
Mass low-flow cutoff
You can change the mass low-flow cutoff with a HART Communicator or ProLink II software.
With a HART Communicator To change the mass low-flow cutoff with a HART Communicator: 1. Press 4, 2, 1. 2. Select “Mass flo cutoff.” 3. Type the new mass low-flow cutoff. 4. Press F4 “ENTER.” 5. Press F2 “SEND.”
With ProLink II software To change the mass low-flow cutoff with ProLink II software: 1. Click the Flow tab. 2. Type the new mass low-flow cutoff in the Mass Flow Cutoff box. 3. Click Apply.
Volume low-flow cutoff
You can change the volume low-flow cutoff with a HART Communicator or ProLink II software.
With a HART Communicator To change the volume low-flow cutoff with a HART Communicator: 1. Press 4, 2, 1. 2. Select “Vol flo cutoff.” 3. Type the new volume low-flow cutoff. 4. Press F4 “ENTER.” 5. Press F2 “SEND.”
With ProLink II software To change the volume low-flow cutoff with ProLink II software: 1. Click the Flow tab. 2. Type the new volume low-flow cutoff in the Vol Flow Cutoff box. 3. Click Apply.
60
Series 1000 and 2000 Transmitters with Intrinsically Safe Outputs