Viscometer Control module – USER MANUAL
Check list prior to Start-Up of the Viscometer Control Module Wear PPE, including safety glass and gloves when connecting solvents to the control module
Step
Check
1
Viscometer Control Module is connected to the PC using the 78 way communication cable
Yes/No (tick)
2
RS232 cable is also connected to the PC
3
24VDC power supply is plugged in at rear
4
Check measuring heads are plugged in to each position
5
Check filling stations are plugged in to each position
6
Solvent 1 “cleaning” solvent is plugged in.
7
Solvent 2 “drying” solvent is plugged in (or optional second cleaning solvent)
8
Solvent 3 is plugged in if applicable
9
Waste line – position 1 is connected – from the control module to waste bottle 1
10
Waste line – position 2 is connected – from the control module to waste bottle 2
11
Vac valve breather 1 is connected – from the control module to breather bottle 1
12
Vac valve breather 2 is connected – from the control module to breather bottle 2
13
Waste vessels are empty
14
Vacuum is connected – from the vacuum manifold (located on the vacuum source) to each waste bottle.
15
Vacuum gauges are connected to each waste bottle and are reading a vacuum level below -0.7 bar e.g. 0.8 or lower
16
If applicable – check auto sampler fill lines are connected
POULTEN SELFE & LEE LTD
RPV-3_VCM_V1.1
31