Shipping the Instrument
Shipping the Instrument
Before shipping the Viscometer Control Module, follow these steps:
Step 1: Carry out a user activated clean and dry of the system.
Step 1B: Carry out a second user activated clean with the drying solvent connected to Solvent 1 –this will flush the cleaning solvent path and will completely dry it, leaving no residue solvent in the system.
Step 2: Carefully remove the solvent lines from the solvent bottles and place in a clean beaker
Step 3: Carry out a second clean and dry – this time not sucking any new solvent into the system –this is designed to purge the system
Step 4: Turn off the vacuum (if external)
Step 5: Turn off the Viscometer Control Module and remove the power lead
Step 6: Dis-connect the waste lines
Step 7: Disconnect the solvent lines
Step 8: Disconnect any other communication cables
Step 9: Ensure that the thumbscrews holding the top lid on are secure and tight
Step 10: Lift the module using the two hand holds on either side
Step 11: Use the original packaging or a strong shipping crate
Step 12: Ensure plenty of foam or packaging material is in the bottom of the case
Step 13: Place the instrument carefully into the packing case
Step 14: Pack with plenty of foam on the sides and top