1 minute read
Service Center
In order to assure vaporizer accuracy, most manufacturers recommend each vaporizer be serviced and calibrated annually - with the exceptions of the D19 & D20, which are calibrated for life.
Internal components involved with vaporization, bypass, pressure compensation, temperature compensation, and wicking are all potential problems if not serviced on a regular schedule.
The expense to maintain an accurate vaporizer is easily justified:
• Economic savings through reduced anesthetic waste • Health concerns are addressed by eliminating waste anesthetic gas leaks • Anesthetic accuracy is improved and confidence in the equipment leads to patient safety
Our Vaporizer Services Include:
• Entire vaporizer disassembly & cleaning • Internal parts cleaning & inspection • Replacement of wicking materials (as needed) • Calibration of thermal controls • Reassembly with new seals & lubricant • Perform pressure test • External cleaning & polishing • Calibration testing with anesthetic agent as labeled • Calibration test certification (with certificate) • Vetland Medical technical expertise & product support • 3 Year Warranty Exception: Ohio Style (90-day warranty)
*Vaporizers are sophisticated medical devices & should ONLY be serviced by qualified service organizations.

SCAN TO LEARN MORE