Product catalogue

Page 1

Dr. Födisch Umweltmesstechnik AG

PRODUCT CATALOGUE

Dr. Födisch Umweltmesstechnik AG Zwenkauer Straße 159 D - 04420 Markranstädt Tel. Fax. Mail. Web.

+49-34205-755-0 +49-34205-755-40 sales@foedisch.de http://www.foedisch.de

© Dr. Födisch Umweltmesstechnik AG - subject to change


Content 1

Contact

1.1 1.2 1.2.1

Headquarter................................................................................................................4 Sales Offices...............................................................................................................5 Germany .....................................................................................................................5

1.2.1.1 1.2.1.2 1.2.1.3

Germany | Sales Office Rhine/Main ......................................................................................... 5 Germany | Western Sales Office.............................................................................................. 6 Germany | Northern Sales Office ............................................................................................. 6

1.2.2

Worldwide ...................................................................................................................7

1.2.2.1

Worldwide | China .................................................................................................................... 7

1.3

Partners ......................................................................................................................8

2

Company Profile

2.1 2.2

Performance profile ..................................................................................................10 Shareholdings...........................................................................................................11

3

Product overview

3.1 3.2 3.3 3.3.1 3.3.2 3.4 3.5

Measuring devices....................................................................................................12 Emission measurement / Process and Analysis measuring technology...................16 Services ....................................................................................................................16 Maintenance & Service (Devices and EMI-Systems) ...............................................16 Engineering...............................................................................................................16 Research/Development ............................................................................................16 Publications ..............................................................................................................16

4

Specifications

4.1 4.2 4.2.1

Selection of measuring point ....................................................................................17 Technical data ..........................................................................................................18 Filter controller ..........................................................................................................18

4.2.1.1 4.2.1.2 4.2.1.3 4.2.1.4 4.2.1.5 4.2.1.6 4.2.1.7 4.2.1.8 4.2.1.9 4.2.1.10 4.2.1.11 4.2.1.12

Measuring principle of triboelectric dust measurement.......................................................... 18 Applications ............................................................................................................................ 19 Filter controller in comparison ................................................................................................ 20 Filter controller PFM 92 .......................................................................................................... 21 Filter controller PFM 92 T....................................................................................................... 22 Filter controller PFM 92 C ...................................................................................................... 23 Filter controller PFM 02 .......................................................................................................... 24 Filter controller PFM 02 Ex..................................................................................................... 25 Filter controller PFM 02 C ...................................................................................................... 26 Filter controller PFM 92 K....................................................................................................... 27 Filter monitorino system ......................................................................................................... 28 References ............................................................................................................................. 29

4.2.2

Dust measuring devices ...........................................................................................30

4.2.2.1 4.2.2.2 4.2.2.3 4.2.2.4 4.2.2.5 4.2.2.6 4.2.2.7

Measuring principle of triboelectric dust measurement.......................................................... 30 Measuring principle optic dust measurement......................................................................... 32 Applications ............................................................................................................................ 33 Dust measuring devices in comparison ................................................................................. 34 Dust measuring device PFM 97 ............................................................................................. 35 Dust measuring device PFM 97 W......................................................................................... 36 Dust measuring device PFM 97 ED ...................................................................................... 37

2

4

10

12

17

4th Edition, Conditions March 2010


4.2.2.8 Dust measuring device PFM 06 ED........................................................................................38 4.2.2.9 Dust measuring device PFM 02 V ..........................................................................................39 4.2.2.10 References..............................................................................................................................40

4.2.3

Gas Analysers .......................................................................................................... 41

4.2.3.1 4.2.3.2 4.2.3.3 4.2.3.4 4.2.3.5 4.2.3.6 4.2.3.7 4.2.3.8 4.2.3.9 4.2.3.10 4.2.3.11 4.2.3.12 4.2.3.13 4.2.3.14 4.2.3.15 4.2.3.16 4.2.3.17

Measuring principle MGA 23...................................................................................................41 Measuring principle MGA 09 / BGM 06 ..................................................................................43 Measuring principle MCA 04 ...................................................................................................43 Measuring principle O2MD 04 .................................................................................................45 Measuring principle HMD 06...................................................................................................46 Applications.............................................................................................................................47 Gas analysers overview ..........................................................................................................48 Gas analyser MGA 23.............................................................................................................49 Gas analyser MGA 09.............................................................................................................50 Gas analyser MCA 04 .............................................................................................................51 Gas analysis system MCA 04 .................................................................................................52 Mobile gas analysis system MCA 04 ......................................................................................53 Gas analyser O2MD 04 ...........................................................................................................54 Gas analyser HMD 06.............................................................................................................55 Bio gas monitor BGM 06 .........................................................................................................56 Gas analyser AMD 08 .............................................................................................................57 References..............................................................................................................................58

4.2.4

Flow measuring devices ........................................................................................... 59

4.2.4.1 4.2.4.2 4.2.4.3 4.2.4.4 4.2.4.5 4.2.4.6

Measuring principle .................................................................................................................59 Applications.............................................................................................................................60 Flow measuring devices in comparison ..................................................................................61 Flow measuring device FMD 02 .............................................................................................62 Flow measuring device FMD 99 .............................................................................................63 References..............................................................................................................................64

4.2.5

Emission evaluation computer ................................................................................. 65

4.2.5.1

Emission evaluation computer DAS 05 & UmweltOffice 2005................................................65

4.2.6

Accessories .............................................................................................................. 66

4.2.6.1 4.2.6.2 4.2.6.3 4.2.6.4 4.2.6.5

Applications.............................................................................................................................66 Gravimetric dust measuring system GMD 06 .........................................................................67 Cascade impactor FIF 7..........................................................................................................69 Probe cleaning system PCS 03 ..............................................................................................71 Condensate detector KM 2 .....................................................................................................72

4.3 4.3.1 4.4 4.4.1 4.5 4.5.1 4.6

Emission measuring systems ................................................................................... 73 References ............................................................................................................... 74 Maintenance and Service......................................................................................... 76 References ............................................................................................................... 78 Engineering .............................................................................................................. 79 References ............................................................................................................... 81 Research/Development ............................................................................................ 82

5

Others

5.1 5.2 5.3

Publications .............................................................................................................. 83 Terms of delivery ...................................................................................................... 84 How to find us ?........................................................................................................ 86

4th Edition, Conditions March 2010

83

3


Product catalog Dr. Födisch AG :: Contact

1

Contact

1.1

Headquarter

CONTACT

Dr. Födisch Umweltmesstechnik AG Zwenkauer Straße 159 D-04420 Markranstädt

Tel.: Fax: E-Mail: Internet:

+49(0)34205-755-0 +49(0)34205-755-40 info@foedisch.de http://www.foedisch.de

Contact persons in the headquarter: Name

Telephone number

e-Mail

0049-(0)34205-755-0

info@foedisch.de

Mr Kaczmarek

0049-(0)34205-755-13

kaczmarek@foedisch.de

Ms Kern

0049-(0)34205-755-60

kern@foedisch.de

Mr Mäder

0049-(0)34205-755-26

maeder@foedisch.de

Mr Sauer

0049-(0)34205-755-39

sauer@foedisch.de

Ms Schramm

0049-(0)34205-755-54

schramm@foedisch.de

Mr Westphal

0049-(0)34205-755-32

westphal@foedisch.de

Ms Hanke

0049-(0)34205-755-57

hanke@foedisch.de

Mr Kronberg

0049-(0)34205-755-15

kronberg@foedisch.de

Mr Pieper

0049-(0)34205-755-30

pieper@foedisch.de

Mr Amboldt

0049-(0)34205-755-31

amboldt@foedisch.de

Mr Fröhlich

0049-(0)34205-755-67

froehlich@foedisch.de

Mr Gläßer

0049-(0)34205-755-48

glaesser@foedisch.de

Ms Gabenstein

0049-(0)34205-755-74

grabenstein@foedisch.de

Mr Grahneis

0049-(0)34205-755-44

grahneis@foedisch.de

Mr Groß

0049-(0)34205-755-45

gross@foedisch.de

Mr Grüttner

0049-(0)34205-755-37

gruettner@foedisch.de

Mr Pehnert

0049-(0)34205-755-19

pehnert@foedisch.de

Mr Kelm

0049-(0)34205-755-37

kelm@foedisch.de

Mr Krauß

0049-(0)172-6636673

krauss@foedisch.com

Ms Marx

0049-(0)34205-755-33

marx@foedisch.de

Mr Römisch

0049-(0)34205-755-67

roemisch@foedisch.de

Mr Steude

0049-(0)34205-755-36

steude@foedisch.de

Ms Wicklein

0049-(0)34205-755-62

wicklein@foedisch.de

0049-(0)34205-755-18

dietrich@foedisch.de

Mrs Krüger

0049-(0)34205-755-28

krueger@foedisch.de

Mr Krummsdorf

0049-(0)34205-755-25

krummsdorf@foedisch.de

Mr Schulz

0049-(0)34205-755-41

schulz@foedisch.de

Mr Wagner

0049-(0)34205-755-49

wagner@foedisch.de

0049-(0)34205-755-16

riemann@foedisch.de

0049-(0)34205-755-34

lehmann@foedisch.de

Management Dr. Födisch Sales

System Technology

Maintenance and Service

Engineering Ms Dietrich Research and Development

Production Mr Riemann Accounting Mrs Lehmann

4

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Contact

1.2

Sales Offices

1.2.1

Germany Headquarter / Markranstädt Sales Office Rhine/Main / Hünstetten Western Sales Office / Mülheim a.d. Ruhr

Locations

Northern Sales Office / Großburgwedel DFU Service GmbH / Kriftel

DFU Analysenservice GmbH / Duisburg

Sales Offcie Rhine/Main

1.2.1.1 Germany | Sales Office Rhine/Main Our local sales office in Rhine/Main area is responsible for the following regions with postal codes: 34xxx - 36xxx, 54xxx - 56xxx, 6xxxx, 7xxxx and 97xxx

Germany Dr. Födisch AG Sales Office Rhine/Main Mr Thomas Lambertz Rabenweg 18 65510 Hünstetten Tel.: +49-34205-755-71 Fax.: +49-34205-755-68 Mobil: +49-177-3461083 E-Mail : lambertz@foedisch.de

4th Edition, Conditions March 2010

5


Product catalog Dr. Födisch AG :: Contact

Westerm Sales Office

1.2.1.2 Germany | Western Sales Office Our local sales office in the Western area is responsible for the following regions with postal codes: 32xxx - 33xxx, 4xxxx, 50xxx - 53xxx and 57xxx - 59xxx

Germany Dr. Födisch AG Western Sales Office Mr Dirk Roloff Kesselbruchweg 70 45478 Mühlheim an der Ruhr Tel.: +49-208-5941587 Fax.: +49-208-5941588 Mobil: +49-174-3344354 E-Mail : roloff@foedisch.de

Northern Sales Office

1.2.1.3 Germany | Northern Sales Office Our local sales office in the Northern area is responsible for the following regions with postal codes: 2xxxx, 30xxx - 31xxx und 37xxx - 38xxx

Germany Dr. Födisch AG Northern Sales Office Mr Reiner Krogbeumker Am Schützenplatz 11 30938 Großburgwedel Tel.: +49-5139-6974040 Fax.: +49-5139-6979000 Mobil: +49-172-7572884 E-Mail: krogbeumker@foedisch.de

6

4th Edition, Conditions March 2010


Product catalog Dr. Fรถdisch AG :: Contact

1.2.2

Worldwide

1.2.2.1 Worldwide | China

Rep offices China

Our representative offices in the PR China

Office Hangzhou

Office Hangzhou

Office Beijing

Dr. Fรถdisch AG Office Hangzhou Ms Sharon Gu 16 D/F, Zhong Tian Mansion, No. 173 Yugu Road 310007 Hangzhou P.R. China Tel.: +86-571-8763 5670 Fax.: +86-571-8763 5690 E-Mail : gu@foedisch.com

Dr. Fรถdisch AG Office Hangzhou Ms Brenda Ying 16 D, Zhong Tian Mansion, No. 173 Yugu Road 310007 Hangzhou P.R. China Tel.: +86-571-8763 5670 Fax.: +86-571-8763 5690 E-Mail: ying@foedisch.com

Dr. Fรถdisch AG Office Beijing Mr Shihong Zhang Ocean Express Tower A, Office 1806 Xia Guangli 66, Chaoyang District Beijing

4th Edition, Conditions March 2010

E-Mail: beijing@foedisch.com

7


Product catalog Dr. Fรถdisch AG :: Contact

Partners

Worldwide

1.3

EUROPE FINLAND

UNITED KINGDOM

Oy Anatec Instruments AB Kumitehtaankatu 5F 04260 Kerava

Casella Measurement Regent House Wolseley Road Kempston Bedford MK42 7JY

Tel: Fax: E-Mail: Web:

+358 9 6899 5250 +358 9 6899 5250 sales@anatec.net http://www.anatec.net/

Tel: Fax: Web:

+44 (0)1234 844100 +44 (0)1234 841490 www.casellameasurement.com

FRANCE Solstice SAS Jean-Philippe MERLINO 2, route de la Noue F-91190 Gif sur Yvette Tel: +33 (0)1 64 86 58 55 Fax: +33 (0)1 64 46 33 95 Mobile: +33 (0)6 77 77 82 15 Web: www.solstice-analyse.com

MIDDLE EAST KAZAKHSTAN

KINGDOM OF SAUDI ARABIEN

BKKS Ltd. Batys Kazakhstan Kuat Service 418440 Republic of Kazakhstan West Kazakhstan oblast, Burlinskii Raion, Aksai-2 Industrial zone, Mail box No.10

Projects System Group P.O. Box 5117 31422 Dammam Kingdom of Saudi Arabia

Tel.: Fax: E-Mail: Web:

8

+7 31133 9 30 67 +7 31133 9 31 82 info@bkks.kz www.bkks.kz

Tel. Fax. Email : Web :

00966-3-8334 100 00966-3-8324 750 info@projectssystemgroup.com foedisch@projectssystemgroup.com www.projectssystemgroup.com

4th Edition, Conditions March 2010


Product catalog Dr. Fรถdisch AG :: Contact

ASIA SOUTH KOREA

TAIWAN

Eroom Technology Co. Ltd. 897-1, HOKYE-DONG, DONGAN-GU 431-082 Anyang-City, KYEONGGI-DO

Atlas Technology Corporation 22 F., No. 1 Bausheng Rd., Yunghe City 234 Taipeh

Tel: Fax: E-Mail: Web:

+82-31-453-4006 +82-31-453-4007 mbox@eroomtech.co.kr www.eroomtech.co.kr

4th Edition, Conditions March 2010

Tel: Fax: E-Mail: Web:

+886-2-2232-0556 +886-2-2231-6657 atlas@atlasgroup.com.tw http://www.atlasgroup.com.tw/

9


Product catalog Dr. Födisch AG :: Company Profile

Company Profile Individual solutions from one source this is the guiding principle of Dr. Födisch Umweltmesstechnik AG. By hiving off the research department from the Umwelttechnik / Luftreinhaltung Leipzig GmbH in Markranstädt in 1991 it started as Dr. Födisch Umweltmesstechnik GmbH at first and changed to Dr. Födisch Umweltmesstechnik AG in 2000.

Company

2

Long-time experiences of the staff in the field of development and test of process- and environmental measuring technologies combined with excellent knowledge in the field of environmental laws and regulations allow to create optimal customised solutions. Our engineers and technicians solve even difficult measuring problems favourably-priced and care for the customer from the idea to the realisation of the measuring task and also in the After-Sales-Service. This includes self-developed measuring devices and complete emission measuring facilities (continuous measurements of dusts and gases as well as other exhaust gas parameters) according to TI Air, 13th and 17th, 27th as well as 30th BImSchV and for operational measuring tasks. Different environmental and engineering services complete the company’s performance profile.

2.1

Performance profile

Process and analysis measuring technology

Dust concentration measuring devices and filter controller Measuring devices for exhaust gas velocity/ exhaust gas flow Analysis measuring technology for process gases Odour measuring technology

Service / Environmental services

Maintenance and service (incl. on call duty in Germany) for emission measuring facilities Approval procedures according to BImSchG Environmental impact analysis Support on immission protection rights Engineering services and optimisation investigations at industrial plants

10

Emission measuring facilities according to TI Air, 13th/17th/ 27th/30th BImSchV Consulting Project planning Engineering Manufacturing (cabinets, analysis containers) Assembly Commissioning After-Sales-Services

Research / Development

Gas measurement Continuous dust measuring technology Continuous odour measuring technology Measuring technology for exhaust gas parameters Tailor-made solutions Pilot plants

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Company Profile

Shareholdings

Company

2.2

At the moment 80 employees are engaged in the group to which various subsidiaries in the fields of automation technology, maintenance and services, water/waste water and civil engineering/reconstruction belong – almost half of them work for the Dr. Födisch Umweltmesstechnik AG. Since 2002 a 100 %subsidiary, the DFU Service GmbH, has been active in the area of Rhine-Main (and since 2009 additionally the DFU Analysenservice GmbH in Ruhr) offering maintenance and service for emission measuring facilities as well as environmental services.

GROUP Dr. Födisch Umweltmesstechnik AG

Dr. Födisch Umweltmesstechnik AG Markranstädt Development, production and sales of environmental, process and analysis measuring technology as well as services

Investments

Subsidiaries DFU Service GmbH Kriftel (100 %)

Fanalmatic Gesellschaft für Umwelttechnik und Industrieautomation Weißandt-Gölzau (100 %)

IMASYS Gesellschaft für Integrative Mess- und Automationssysteme mbH Markranstädt (75 %)

Peter Nitschke Dosieranlagen und Service GmbH Markranstädt (67 %)

Supplier for assembly, maintenance and service works at measuring devices and systems for emission and process monitoring

Supplier for electric and electronic control and automation facilities incl. process visualisation

supplier of engineering services, maintenance and service

supplier for technical equipments and services in the field of water and waste water

4 employees

16 employees

4th Edition, Conditions March 2010

7 employees

7 employees

GIBA Gesellschaft für Ingenieurbau, Bauwerksinstandhaltung und Anlagenmanagement mbH Markranstädt (40 %)

AllTec Automatisierungsund Kommunikationstechnik GmbH Borna (26 %)

PROCEMS Asia Pacific Pte Ltd Centre of Competence for South East Asia Singapur (10 %)

Supplier for planning and other services in the field of engineering and pipe construction

Supplier of automation technology, control room construction and information technology

Supplier of process and emission measuring technology as well as services

3 employees

150 employees

5 employees

11


Product catalog Dr. Födisch AG :: Product overview

Product overview

3.1

Measuring devices

FILTER CONTROLLER

3

12

PFM 92 [in-situ]

Filter controller for in-situ dust measurement Consists of probe and control unit On-site – display of the measuring value at the control unit Approved according to TI Air

PFM 92 T [in-situ]

Special version of PFM 92 Filter controller for in-situ dust measurement Consists of probe and control unit High-temperature configuration up to 500 °C

PFM 92 C [in-situ]

Filter controller for in-situ dust measurement Compact device Suitable for the construction of filter monitoring systems Approved according to TI Air

PFM 02 [in-situ]

Filter controller for in-situ dust measurement Compact device Display for providing the measuring value in mg/m³ or as line diagram Approved according to TI Air

PFM 02 Ex [in-situ]

Filter controller for in-situ dust measurement according to ATEX-category 2/3D Compact device Display for providing the measuring value in mg/m³ or as line diagram Approved according to TI Air

PFM 02 C [in-situ]

Basic version of the PFM 02 without display or keyboard Compact device Suitable for the construction of filter monitoring systems

PFM 92 K [in-situ]

Mobile filter controller for in-situ dust measurement With low efforts many measuring points reachable Online – line diagram (digital recorder) Variable probe length

Filter monitoring system

For plants with several or even many single emission sources Anticipatory monitoring of all main filter systems High efficiency of preventive filter maintenance

4th Edition, Conditions March 2010


DUST MEASURING DEVICES

Product catalog Dr. Födisch AG :: Product overview

PFM 97 [in-situ]

Combined measuring device for dust, flow and temperature In-situ – dust measurement Approved according to TI Air, 13th and 27th BImSchV

PFM 97 W [in-situ]

Combined measuring device for dust, flow and temperature In-situ – dust measurement Approved according to TI Air, 13th , 17th and 27th BImSchV

PFM 97 ED [extractive]

Extractive dust concentration measuring device Dust measurement in wet and adhesive exhaust gases (e.g. behind wood-dryers) Triboelectric dust sensor Approved according to TI Air, 13th , 17th and 27th BImSchV

PFM 06 ED [extractive]

Extractive dust concentration measuring device Dust measurement in wet and adhesive exhaust gases (e.g. behind wood-dryers) Optic dust sensor

PFM 02 V [in-situ]

In-situ – dust measurement e.g. in combination with the flow measuring device FMD 02 or other approved velocity measuring devices Approved according to TI Air, 13th , 17th and 27th BImSchV

4th Edition, Conditions March 2010

13


GAS ANALYSERS

Product catalog Dr. Födisch AG :: Product overview

14

MGA 23 [extractive]

MultiGasAnalyser for extractive gas components’ measurement Cold measurement behind gas cooler Approved according to TI Air, 13th and 27th BImSchV

MGA 09 [extractive]

MultiGasAnalyser for extractive gas components’ measurement Cold measurement behind gas cooler

MCA 04 [extractive]

MultiComponentAnalyser for extractive gas components’ measurement Hot measurement (use without gas cooler)

MCA 04 [extractive]

Analysis system based on the MultiComponentAnalyser MCA 04 for extractive gas components’ measurement Hot measurement (use without gas cooler) Approved according to TI Air, 13th , 17th and 27th BImSchV

MCA 04 mobile [extractive]

Mobile analysis system based on multi-component analyser MCA 04 for extractive gas measurement Hot measurement (use without gas cooler)

O2MD 04 [in-situ]

Continuous in-situ – oxygen measurement Zirkonium oxide probe

HMD 06 [extractive]

HumidityMonitoringDevice for continuous extractive humidity measurement in flue gases Very efficient in combination with extractive cold gas analysis

BGM 06 [extractive]

Biogasmonitor Dilution principle Measuring components: CH4, CO2, O2, H2S

AMD 08 [extractive]

Continuous extractive acid dew point measurement Optimisation of the power plant’s operations

4th Edition, Conditions March 2010


ACCESSORIES

DATA SYSTEM

FLOW MEASURING DEVICES

Product catalog Dr. Fรถdisch AG :: Product overview

FMD 99 [in-situ]

Continuous measurement of: Exhaust gas velocity resp. flow and Exhaust gas temperature as well as Pressure (option) Use also in case of high dust contents in the exhaust gas Option: back-purging

FMD 02 [in-situ]

Continuous measurement of: Exhaust gas velocity resp. flow and Exhaust gas temperature Compact device Approved according to TI Air, 13th , 17th and 27th BImSchV

DAS 05

Emission evalation PC Approved according to TI Air, 13th , 17th and 27th BImSchV

GMD 06 [extractive]

Manual gravimetric dust content determination according to VDI 2066, page 1, 2, 3, 7 in streaming gases

FIF 7

Cascade impactor with 7 separation stages for manual gravimetric dust content determination according to VDI 2066, page 5

PCS 03

Automatic probe cleaning system for PFM 97 and PFM 97 W as well as FMD 99

KM 2

Condensate monitoring in measuring gases for continuous emission measurement behind gas coolers

4th Edition, Conditions March 2010

15


Product catalog Dr. Födisch AG :: Product overview

Emission measurement / Process and Analysis measuring technology Measuring systems for operational and official monitoring of pollutants For fixed installation or for mobile use High-quality complete solutions for emission monitoring Support for implementing authority requirements (incl. determination of the measuring points) Delivery of systems being proper in function and low in maintenance Compact construction of multi-component measuring systems

CEMS

3.2

3.3

Services

3.3.1

Maintenance & Service (Devices and EMI-Systems)

SERVICE

Periodic maintenance, repair of measuring facilities (cycles: depending on customer and plant) 24 h – response time (resp. on call duty after agreement) Training of operator’s staff Remote maintenance and remote diagnosis

Engineering Authority engineering, e.g.: Approval procedure according to BImSchG Environmental impact analyses Immission prognoses Process services Execution of seminars

ENGINEERING

3.3.2

Research/Development Development of devices / optimisation Dust channel Research for industry

F&E

3.4

Publications

Book

3.5

16

Diverse reports and publications in magazines Book: Dust emission measuring technology / Author: Dr. Holger Födisch

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

4

Specifications

4.1

Selection of measuring point

In order to get representative results with regard to the dust concentration of plants whose emissions shall be monitored continuously, the correct installation of the dust measuring devices is a basic condition. This is also valid for the application of measuring devices within the frame of process monitoring. (e.g. qualitative measurement – monitoring of the clean gas content of bag filters). The probe’s place of installation should fulfil the requirements of valid guidelines (e.g. EN 13284-1, VDI 2066 page 1). In case of doubts it is recommended to let have determined the measuring point by a responsible measuring institute (in Germany according to §§ 26/28 BImSchG). We recommend to realise as entry and exit section at least 5 times the diameter of the exhaust gas channel.

VDI 2066 pg. 1 EN 13284-1 5xD

Right!

VDI 2066 pg. 1 EN 13284-1 5xD

Wrong! Right! <5xD

2xD

5xD

VDI 2066 pg. 1 EN 13284-1

Basically it has to be considered that at the measuring point the dust and smoke gas distribution must prevail as homogeneously as possible in order to get a representative registration of the dust load over the channel cross-section.

4th Edition, Conditions March 2010

17


Product catalog Dr. FĂśdisch AG :: Specifications

4.2

Technical data

4.2.1

Filter controller

4.2.1.1 Measuring principle of triboelectric dust measurement

PFM 92

PFM 92 T

PFM 92 C

PFM 02 C

PFM 92 K

Filter monitoring system

PFM 02

PFM 02 Ex

In order to determine the dust content in industrial exhaust gases there are numerous non-continuous and continuous measuring methods. The Dr. FÜdisch Umweltmesstechnik AG mainly concentrates on the measuring principle of triboelectricity – a quite young method with lots of possibilities to realise in concrete measuring technology. Probe

Insulator Probe rod

Probe

If two bodies are contacted by friction or touch a charge transfer occurs. The arising charge difference is the basis for the filter controllers and dust measuring devices after the triboelectric measuring principle whereby the charge exchange between probe and nearby streaming and colliding dust particles is used.

The triboelectric signal depends on different process conditions and the mechanical and electrical properties of the dusts.

Advantages of the triboelectric dust measurement: Measurement of the dust content (in-situ or extractively) High measuring precision Simple assembly without special tools Low maintenance demand (long operation times) Minimum operation costs No purging air required Very low influence of probe soiling on the measuring result Several operational voltages for selection

18

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

4.2.1.2 Applications The qualitative dust measurement includes the monitoring of the clean gas dust content behind de-dusting facilities and at the same time the evaluation of the flue gas purification plant’s state.

Device

Applications

Measuring ranges after gravimetric calibration

PFM 92 [in-situ]

Monitoring of filtering precipitators (bag or cartridge filters) and cyclones Tendency measurements of raw gas dust concentrations

0,1 ... 1.000 mg/m³, 0 ... 100 %

PFM 92 T [in-situ]

Monitoring of dust emissions at high exhaust gas temperatures

0,1 ... 1.000 mg/m³, 0 ... 100 %

PFM 92 C [in-situ]

Monitoring of filtering precipitators (bag or cartridge filters) and cyclones Construction of complex monitoring systems

0,1 ... 1.000 mg/m³, 0 ... 100 %

PFM 02 [in-situ]

Monitoring of filtering precipitators (bag or cartridge filters) and cyclones Construction of complex monitoring systems

0 ... 10 (1.000) mg/m³, 0 ... 100 %

PFM 02 Ex [in-situ]

Monitoring of filtering precipitators (bag or cartridge filters) and cyclones Construction of complex monitoring systems

0 ... 10 (1.000) mg/m³, 0 ... 100 %

PFM 02 C [in-situ]

Monitoring of filters und construction of filter monitoring systems

0,1 ... 1.000 mg/m³, 0 ... 100 %

PFM 92 K [in-situ]

Monitoring of filtering precipitators, especially pulse-jet-filters

0,1 ... 1.000 mg/m³, 0 ... 100 %

4th Edition, Conditions March 2010

TÜV -approval

TI Air

19


Product catalog Dr. Födisch AG :: Specifications

Measuring arrangement

In-situ

Measuring arrangement

Extractive

Voltage supply

230/110 VAC, 50/60 Hz

Voltage supply

24 VAC

Voltage supply

24 VDC

PFM 92 K

PFM 02 C

PFM 02 Ex

PFM 02

PFM 92 C

PFM 92 T

PFM 92

4.2.1.3 Filter controller in comparison

1)

Probe Control unit Media temperature

≤ 280 °C

Media temperature

> 280 °C

Flow velocity

from 3 m/s

Analogue output

4 ... 20 mA

Digital output

Limit value 1

Digital output

Limit value 2

Digital output

Limit value 2 / Maintenance request

Digital output

Error

Digital output

Error/Maintenance

Approval

TI Air

1) only 230 VAC

20

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

FILTER CONTROLLER

4.2.1.4 Filter controller PFM 92

TÜV-approved according to TI Air

Basic device consisting of probe and control unit Simple assembly: Control unit: wall assembly Probe: 1"-sleeve (DIN 2986) optionally flange DN 25 PN 6 Analogue output: Dust: 4 ... 20 mA Digital output (potential-free): Limit value 1 and 2 (adjustable) Failure (Power failure) 4 measuring ranges adjustable: in relation 0,33 : 1 : 3,33 : 10

Application and function The filter controller PFM 92 serves the qualitative monitoring of dusty emissions. In particular it is often used for the registration of the clean gas dust content behind de-dusting facilities. The measuring principle is based on the use of the triboelectric effect (charge transfer at colliding or passing of particles on leading surfaces). The filter controller consists of an isolated probe which is installed in the clean gas channel. The charge transferred by contact and triboelectric processes is derived as current, converted in the evaluation unit, amplified and provided as norm signal 4 ... 20 mA. Via a limit value detector the exceeding of the allowable emission limit value can be signalised optically or acoustically. Technical data Control unit:

Weather-proof aluminium case IP 65, Dimensions: 210 x 240 x 280 mm (H x W x D)

Probe:

Stainless steel probe rod with pre-amplifier in the probe head, probe length 300 mm (adjustments possible), IP 65

Measuring principle:

Dust: measurement with 1 triboelectric sensor

Measuring range:

0,1 ... 1.000 mg/m³ (special measuring ranges on demand)

Calibration:

By gravimetric reference measurements

Display:

LCD-Display (0 ... 100 %)

Media temperature:

max. 280 °C

Ambient temperature:

-20 ... +50 °C

Dew point difference:

min. +5 K

Flow velocity:

From appr. 3 m/s

Analogue signals:

1 x 4 ... 20 mA (dust)

Digital signals:

3 potential-free contacts (failure, limit value 1 and 2)

Power supply:

110 VAC, 230 VAC or 24 VAC, 24 VDC

TÜV-Approval:

TI-Air

4th Edition, Conditions March 2010

21


Product catalog Dr. Födisch AG :: Specifications

FILTER CONTROLLER

4.2.1.5 Filter controller PFM 92 T

TÜV-approved according to TI Air

High temperature design of the PFM 92 consisting of probe and control unit (max. 500 °C exhaust gas temperature) Simplest assembly: Control unit: wall assembly Probe: flange DN 25 PN 6 Analogue output: Dust: 4 ... 20 mA Digital outputs (potential-free): Limit value 1 and 2 (adjustable) Failure (power failure) 4 measuring ranges adjustable: in relation 0,33 : 1 : 3,33 : 10

Application and function The filter controller PFM 92 T serves the qualitative monitoring of dusty emissions in hot exhaust gases (up to 500 °C). The measuring principle is based on the use of the triboelectric effect (charge transfer at colliding or passing of particles on leading surfaces). The filter controller consists of an isolated probe which is installed in the clean gas channel. The charge transferred by contact and triboelectric processes is derived as current, converted in the evaluation unit, amplified and provided as norm signal 4 ... 20 mA. Via a limit value detector the exceeding of the allowable emission limit value can be signalised optically or acoustically. Technical data Control unit:

Weather-proof aluminium case IP 65, Dimensions: 210 x 240 x 280 mm (H x W x D)

Probe:

Stainless steel probe rod with pre-amplifier in the probe head, probe length 300 mm (adjustments possible), IP 65

Measuring principle:

Dust: measurement with 1 triboelectric sensor

Measuring range:

0,1 ... 1.000 mg/m³ (special measuring ranges on demand)

Calibration:

By gravimetric reference measurements

Display:

LCD-Display (0 ... 100 %)

Media temperature:

max. 500 °C

Ambient temperature:

-20 ... +50 °C

Dew point difference:

min. +5 K

Flow velocity:

From appr. 3 m/s

Analogue signals:

1 x 4 ... 20 mA (dust)

Digital signals:

3 potential-free contacts (failure, limit value 1 and 2)

Power supply:

110 VAC, 230 VAC or 24 VAC, 24 VDC

TÜV-Approval:

TI Air

22

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

4.2.1.6 Filter controller PFM 92 C

FILTERWÄCHTER

Compact device consisting of probe Simplest assembly: Probe: 1"-sleeve (DIN 2986) optionally flange DN 25 PN 6 Analogue output: Dust: 4 ... 20 mA Digital outputs (potential-free): Limit value 1 and 2 (adjustable) Failure (power failure) 4 measuring ranges adjustable: in relation 0,33 : 1 : 3,33 : 10 TÜV-approved according to TI Air

Application and function The filter controller PFM 92 C serves the qualitative monitoring of dusty emissions. In particular it is often used for the registration of the clean gas dust content behind de-dusting facilities. The measuring principle is based on the use of the triboelectric effect (charge transfer at colliding or passing of particles on leading surfaces). The filter controller consists of an isolated probe which is installed in the clean gas channel. The complete electronic is integrated in the probe head. No additional control unit is required. The charge transferred by contact and triboelectric processes is derived as current, converted in the evaluation unit, amplified and provided as norm signal 4 ... 20 mA. Via a limit value detector the exceeding of the allowable emission limit value can be signalised optically or acoustically. Due to the compact design the measuring device is especially suited for the construction of complex monitoring systems with several emission sources. Technical data Control unit:

No control unit required

Probe:

Round stainless-steel probe rod at the probe head, complete electronic in the probe head – no separate control unit, probe length 300 mm (adjustments possible), IP 65

Measuring principle:

Dust: measurement with 1 triboelectric sensor

Measuring range:

0,1 ... 1.000 mg/m³ (special measuring ranges on demand)

Calibration:

By gravimetric reference measurements

Display:

none

Media temperature:

max. 280 °C

Ambient temperature:

-20 ... +50 °C

Dew point difference:

min. +5 K

Flow velocity:

From appr.. 3 m/s

Analogue signals:

1 x 4 ... 20 mA (dust)

Digital signals:

3 potential-free contacts (failure, limit value 1 and 2)

Power supply:

110 VAC, 230 VAC or 24 VAC, 24 VDC

TÜV-Approval:

TI Air

4th Edition, Conditions March 2010

23


Product catalog Dr. Födisch AG :: Specifications

FILTER CONTROLLER

4.2.1.7 Filter controller PFM 02

TÜV-approved according to TI Air Limit value

Limit value

Limit value

2

On Off

1

On

2

5s

mg m³

1,5

Off

1

On Off

2

0.5s

5s

100%

1

%

3,7

0

1min

%: 12.9

0

1

2

3

Gain

0

1

2

3

Gain

0

1

2

3

Gain

Compact device consisting of probe Simplest assembly: Probe: 1"-sleeve (DIN 2986) optionally flange DN 25 PN 6 Analogue output: Dust: 4 ... 20 mA Digital outputs (potential-free): Limit value 1 (adjustable) Limit value 2 (adjustable)/ maintenance request Failure/Maintenance 4 measuring range adjustable Display in mg/m³ or as online line diagram Target value calibration Simplest probe rod adjustment to the measuring conditions on site (probe rod turnable and exchangeable)

Application and function The dust measuring device PFM 02 is a perfect device in order to determine effectively damages at filtering precipitators on basis of the triboelectric measuring principle. The device is simple in construction and very user-friendly. The integrated micro-controller allows numerous possibilities to use the PFM 02 as well as a timely monitoring of the dust emissions. Special features Target value calibration: Enter a known, averaged dust content in operation state - The PFM 02 determines automatically appropriate calibration factors and shows the quantitative dust content in mg/m³. Technical data Control unit:

complete electronic in the probe head – no separate control unit), IP 65

Probe:

Stainless-steel probe rod (turnable and exchangeable) at the probe head, probe length 300 mm (adjustments possible), IP 65

Measuring principle:

Dust: measurement with 1 triboelectric sensor

Measuring range:

0 ... 100 % bzw. 0 ... 10 (1.000) mg/m³

Calibration:

By gravimetric reference measurements

Display:

Graphic display with online line diagram

Media temperature:

max. 280 °C

Ambient temperature:

-20 ... +50 °C

Dew point difference:

min. +5 K

Flow velocity:

From appr. 3 m/s

Analogue signals:

1 x 4 ... 20 mA, galvanically isolated

Digital signals:

3 potential-free contacts (failure/maint., limit value 1 and 2/maint. request)

Power supply:

110 VAC, 230 VAC bzw. 24 VDC

TÜV-Approval:

TI Air

24

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

FILTER CONTROLLER

4.2.1.8 Filter controller PFM 02 Ex

TÜV-approved according to TI Air Limit value

Limit value

Limit value

2

On Off

1

On

2

5s

mg m³

1,5

Off

1

On Off

2

0.5s

5s

100%

1

%

3,7

0

1min

%: 12.9

0

1

2

3

Gain

0

1

2

3

0

Gain

1

2

3

Gain

Compact device consisting of probe Filter controller as device of category 1/3D and 3G Simplest assembly: Probe: 1"-sleeve (DIN 2986) optionally flange DN 25 PN 6 Analogue output: Dust: 4 ... 20 mA Digital outputs (potential-free): Limit value 1 (adjustable) Limit value 2 (adjustable)/ maintenance request Failure/Maintenance 4 measuring range adjustable Display in mg/m³ or as online line diagram Target value calibration Simplest probe rod adjustment to the measuring conditions on site (probe rod turnable and exchangeable)

Application and function The dust measuring device PFM 02 Ex is a perfect device in order to determine effectively damages at filtering precipitators on basis of the triboelectric measuring principle. The device is simple in construction and very userfriendly. The integrated micro-controller allows numerous possibilities to use the PFM 02 Ex as well as a timely monitoring of the dust emissions. Special features Target value calibration: Enter a known, averaged dust content in operation state - The PFM 02 Ex determines automatically appropriate calibration factors and shows the quantitative dust content in mg/m³. Technical data Control unit:

complete electronic in the probe head – no separate control unit), IP 65

Probe:

Stainless-steel probe rod (turnable and exchangeable) at the probe head, probe length 300 mm (adjustments possible), IP 65

Measuring principle:

Dust: measurement with 1 triboelectric sensor

Measuring range:

0 ... 100 % bzw. 0 ... 10 (1.000) mg/m³

Calibration:

By gravimetric reference measurements

Display:

Graphic display with online line diagram

Media temperature:

max. 280 °C

Ambient temperature:

-20 ... +50 °C

Dew point difference:

min. +5 K

Flow velocity:

From appr. 3 m/s

Analogue signals:

1 x 4 ... 20 mA, galvanically isolated

Digital signals:

3 potential-free contacts (failure/maint, limit value 1 and 2/maint. request)

Power supply:

230 VAC

ATEX- Guideline:

Filter controller as device of category 1/3D and 3G

TÜV-Approval:

TI-Air

4th Edition, Conditions March 2010

25


Product catalog Dr. Födisch AG :: Specifications

4.2.1.9 Filter controller PFM 02 C

FILTER CONTROLLER

Compact device consisting of probe Simplest assembly: Probe: 1"-sleeve (DIN 2986) Analogue output: Dust: 4 ... 20 mA Digital outputs (potential-free): Limit value 1 (60 %) Limit value 2 (80 %) / maintenance request Failure/maintenance

Application and function The dust measuring device PFM 02 C is a perfect device in order to determine effectively damages at filtering precipitators on basis of the triboelectric the measuring principle. The in-situ device is simple in construction and very user-friendly. It allows a timely monitoring of dust emissions. As it was concentrated on substantial parameters the PFM 02 C is especially suited for the application in complex filter monitoring systems. Due to the compact design the measuring device has also well-proven for the construction of complete monitoring systems with several emission sources.

Technical data Control unit:

complete electronic in the probe head – no separate control unit), IP 65

Probe:

Round stainless-steel probe rod at the probe head, probe length 300 mm (adjustments possible), IP 65

Measuring principle::

Dust: measurement with 1 triboelectric sensor

Measuring range:

0,1 ... 1.000 mg/m³ (special measuring ranges on demand)

Calibration:

By gravimetric reference measurements

Display:

none

Media temperature:

max. 280 °C

Ambient temperature:

-20 ... +50 °C

Dew point difference:

min. +5 K

Flow velocity:

From appr. 3 m/s

Analogue signals:

1 x 4 ... 20 mA, galvanically isolated

Digital signals:

3 potential-free contacts (failure/maint, limit value 1 and 2/maint. request)

Power supply:

24 VDC

TÜV-Approval:

none

26

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

FILTER CONTROLLER

4.2.1.10 Filter controller PFM 92 K

TÜV-approved according to TI Air

Portable device Simple assembly: Probe: 1"-sleeve (DIN 2986) Analogue output: Dust: 4 ... 20 mA RS 232 Analogue input: Optionally max. 2 analogue inputs possible 4 Measuring ranges adjustable: in relation 0,33 : 1 : 3,33 : 10 Measuring value display on digital recorder (ECOGraph) – line diagram PC-evaluation software Simplest probe rod adjustment to the measuring conditions on site (probe rod exchangeable

Application and function The mobile filter controller PFM 92 K is a perfect device in order to determine effectively damages at filtering precipitators (e.g. defective filter bags, inactive membrane valves, insufficient filter element fixings, not tightly-closing poppet valves…). The filter diagnosis unit PFM 92 K serves the qualitative monitoring of dusty emissions. In particular it is often used for the quick registration of the clean gas dust content behind de-dusting facilities. The measuring principle is based on the use of the triboelectric effect (charge transfer at colliding or passing of particles on leading surfaces).The filter controller consists of an isolated probe which is installed in the clean gas channel. The charge transferred by contact and triboelectric processes is derived as current, converted in the evaluation unit, amplified and provided as norm signal 4 ... 20 mA. The measuring values resp. the filter diagram can be seen and evaluated immediately via the integrated electronic recorder. Technical data

Control unit:

Portable aluminium suit-case, 370 x 490 x 215 mm (H x W x D), IP 20

Probe:

Stainless-steel probe rod with pre-amplifier in the probe head, probe length: 30 ... 500 mm (adjustments possible), IP 65

Measuring principle:

Dust: measurement with 1 triboelectric sensor

Measuring range:

0,1 ... 1.000 mg/m³ (special measuring ranges on demand)

Calibration:

By gravimetric reference measurements

Display:

LCD-Display (0 ... 100 %) resp. recorder

Media temperature:

max. 280 °C

Ambient temperature:

-20 ... +50 °C

Dew point difference:

min. +5 K

Flow velocity:

From appr. 3 m/s

Analogue signals:

1 x 4 ... 20 mA (dust)

Analogue inputs:

Optionally max. 2 x 4 ... 20 mA

Digital signals:

Power supply:

230 VAC (optionally accumulator operation)

Recorder:

Electronic recorder with LCD-screen

TÜV-Approval:

TI-Air

4th Edition, Conditions March 2010

27


Product catalog Dr. Födisch AG :: Specifications

4.2.1.11 Filter monitorino system

FILTER CONTROLLER

Central monitoring for any number of measuring points Compact wall cabinet for registration and presentation of measuring valus Simple installation: Wall assembly Manipulation-proof recording optic /acoustic signalling of limit value exceedings by means of lamp / buzzer, to be confirmed

Application and function

The qualitative dust measurement guarantees not only the monitoring of the clean gas dust content behind filters, but allows the evaluation of the filters’ state as well. The advantage of the filter monitoring at single filters is undoubted. Moreover it is possible to monitor numerous plants in an efficient way by coupling several devices and installing an evaluation software. The target is to form the demand for maintenance and repair as well as the energy consumption as efficiently as possible by means of a timely filter diagnosis on the one hand and an optimisation of the filter cleaning cycles on the other hand. A filter monitoring system provides an overview about the state of all filter plants in a factory anytime. Depending on the quantity of measuring points you can chose between a system with integrated electronic recorder and a PC-based system. Both alternatives include an optic respectively acoustic limit value monitoring apart from the data recording. Technical data Control unit:

Wall cabinet

Probes:

one triboelectric probe per filter – depending on case of application PFM 92 C / PFM 02 / PFM 02 Ex

Ambient temperature:

-20 ... +50 °C

Power supply:

230 VAC, integrated voltage supply for all filter controllers

Signalling:

Buzzer or light in case of limit value exceeding

Version A: recorder Number of measuring points

1 ... 6 Measuring points per recorder (max. 18 measuring points in total)

Data storage

Internal memory and Compact Flash Card

Evaluation

Standard software for installation on customer’s PC

Data transfer

Serial interface / USB Ethernet connection (integration into internal network possible)

Version B: PC - System Number of measuring points

any

Data storage

PC

Evaluation

Customised software

Data transfer

BUS-connection possible

28

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

4.2.1.12 References Device

References (extract)

Number of installed devices

PFM 92

Bayer AG, Krefeld Degussa AG, Hüls Kali und Salz, Unterbreizbach Lafarge Zement, Karsdorf Matsushima, Kitakyushu (Japan) Metaleurop, Nordenham Norddeutsche Affinerie, Hamburg Readymix Zementwerk, Rüdersdorf

>1.000

PFM 92 C

Bayer AG, Krefeld Dubai Aluminium (Dubai) Euro Science Co Ltd., Seoul (Korea) Heraeus Quarzglas GmbH, Bitterfeld Intensiv Filter, Velbert Krematorium Görlitz, Halle, Dessau, Weimar ... Lafarge Zementwerk, Adriasebina (Italien) Nanjing Electric Power, Nanjing (China) Nestlé AG, Weiding Norddeutsche Affinerie, Hamburg Rheinkalk Solvay Sodaproduktion, Rheinberg Sulzer Metco, Salzgitter Wacker Chemie GmbH, Burghausen VW AG, Braunschweig, Kassel

> 2.000

PFM 92 K

IMMB Opole (Polen) DK Science, Seoul (Korea) Unicem S.p.A. (Italien) CT Umwelttechnik (Schweiz) ABB Umwelttechnik, Butzbach Daimler AG TFT Filtertechnik Luft- und Umwelttechnik GmbH Rheinkalk GmbH K+S Kali GmbH

> 50

PFM 02

Dillinger Hütte Saarstahl VARTA automotive Systems GmbH Treibacher Schleifmittel ALCOA Trasformazioni SrL (Italien) Bremer Rolandmühle Erling GmbH & Co. KG Buderus, Wetzlar TWEER Gießerei Grace Silica GmbH, Düren Harpen EKT GmbH Krematorium Gröbers Norddeutsche Affinerie, Hamburg Paul Hartmann AG, Heidenheim Sentek, Glostrup (Dänemark) Südchemie, Moosburg UniverStar Science & Technology, Shenzhen City (China)

> 350

4th Edition, Conditions March 2010

29


Product catalog Dr. Födisch AG :: Specifications

4.2.2

Dust measuring devices

4.2.2.1 Measuring principle of triboelectric dust measurement

PFM 97

PFM 97 W

PFM 97 ED

PFM 06 ED

PFM 02 V

In order to determine the dust content in industrial exhaust gases there are numerous non-continuous and continuous measuring methods. The Dr. Födisch Umweltmesstechnik AG mainly concentrates on the measuring principle of triboelectricity – a quite young method with lots of possibilities to realise in concrete measuring technology. If two bodies are contacted by friction or touch a charge transfer occurs. The arising charge difference is the basis for the filter controllers and dust measuring ∆p-Transmitter devices after the triboelectric measuring principle whereby the charge exchange between probe and nearby streaming and colliding dust particles is used. Valves The triboelectric signal depends on different process conditions and the mechanical and electrical properties of Pt100 Dust probes the dusts. Flow profile

Terminal case

It is said that after the dust concentration the velocity takes the biggest influence on the charge transfer.

Influencing variables on the triboelectric dust measurement The mathematic correlation of these variables is shown in the following:

30

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

ci.B. = A ⋅ cal ⋅ v

Exp

+D

Exp = B ⋅ v 2 + C ⋅ v + E

ci.B.

Dust concentration [mg/m³]

A, D

Constants from the dust calibration

cal

Measuring signal [V]

v

Velocity of exhaust gas [m/s]

Exp

Exponent of velocity

B, C, E

Constants from the velocity correlation

The exponent contains the mathematic compensation of the velocity influence on the triboelectric signal. The factors are pre-adjusted at factory.

Advantages of the triboelectric dust measurement: Measurement of dust content (in-situ or extractive) High measuring precision Simple assembly without special tools Low maintenance efforts (long operation times) Minimum operational costs No purging air required Very low influence of the probe soiling on the measuring result Several operational voltages for selection

4th Edition, Conditions March 2010

31


Product catalog Dr. Födisch AG :: Specifications

4.2.2.2 Measuring principle optic dust measurement The scattered light measurement by means of laser diodes can be done via a combination of emitter and receiver unit but also in form of a measuring rod. The measuring volume is defined by the intersection of the laser diode’s emitter beam with the receiver. Modulled light in a visible range is emitted, scattered by the existing particles and registered by a highlysensitive detector. This measuring arrangement allows a relatively simple assembly without complicate adjustments.

Empfänger

Sender

Sonde

Spülluft

In case of the PFM 06 ED the measuring rod is arranged outside of the stack, that means it is an extractive measurement.

Sondenelektronik

∆p

Sondenheizung

+

Laserstrahl

T

Injektor

M

Kugelhahn

∆p

-

Stellventil

Messzelle Gebläse Injektorluft

Magnetventil

+

T

Abluft Heizung Verdünnungsluft

32

Gebläse Verdünnungsluft

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

4.2.2.3 Applications

PFM 97 [in-situ]

Monitoring of dust emissions in all industrial branches with dry exhaust gas

0 … 20 mg/m³ to max. 500 mg/m³

PFM 97 W [in-situ]

Monitoring of dust emissions in all industrial branches with dry exhaust gas

0 … 15 mg/m³ to max. 500 mg/m³

PFM 97 ED [extractive]

Monitoring of: Wet scrubbers (e.g. in waste incinerations) Fertilizer production Soil sanitation plants Monitoring of exhaust gases in the woodprocessing industry

0 … 15 mg/m³ to max. 500 mg/m³

PFM 06 ED [extractive]

Monitoring of: Wet scrubbers (e.g. in waste incinerations) Fertilizer production Soil sanitation plants Monitoring of exhaust gases in the woodprocessing industry

0 … 15 mg/m³ to max. 500 mg/m³

PFM 02 V [in-situ]

Monitoring of dust emissions in all industrial branches with dry exhaust gas Combined with FMD 02

0 … 10 mg/m³ to max. 1.000 mg/m³

4th Edition, Conditions March 2010

th

27 BImSchV

th

17 BImSchV

Measuring ranges

th

Applications

13 BImSchV

Device

TI AIr

TÜV -Approval

33


Product catalog Dr. Födisch AG :: Specifications

Measuring arrangement

In-situ

Measuring arrangement

Extractive

Measuring principle

Triboelectric

Measuring principle

Optic

Power supply

400 VAC, 50/60 Hz

Power supply

230/110 VAC, 50/60 Hz

Power supply

24 VDC

PFM 02 V

PFM 06 ED

PFM 97 ED

PFM 97 W

PFM 97

4.2.2.4 Dust measuring devices in comparison

probe Control unit Media temperature

≤ 280 °C

Flow velocity

from 3 m/s

Analogue output

4 ... 20 mA

Digital output

Limit value 1

Digital output

Limit value 2 / Maintenance request

Digital output

Failure

Digital output

Maintenance

Digital output

Failure/Maintenance

Digital output

Measuring range

Approval

TI Air

Approval

13 BImSchV

Approval

17 BImSchV

Approval

27 BImSchV

2x

2x

1)

1)

1x

1x

th

th

th

1) separately arranged frame with control unit and 2 blowers. 2) separaely arranged frame with terminal box and blower.

34

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

DUST MEASURING DEVICE

4.2.2.5 Dust measuring device PFM 97

TÜV-approved according to TI Air, 13th and 27th . BImSchV

Fields of application The dust concentration measuring device PFM 97 is used for the continuous measurement of dust contents, especially for the emission measurement according to TI Air, 13th and 27th BImSchV as well as for the dust concentration measurement in process monitoring. Parallel to the dust measurement flow and temperature of the smoke gas are registered.

Basic device consisting of probe and control unit Combined device – measurement of dust, exhaust gas velocity or flow as well as temperature (option pressure) at the same time Simple assembly: Control unit: wall assembly Probe: special flange DN 80 PN 6, Di=100 mm Analogue outputs: 4 ... 20 mA Digital outputs (potential-free): Limit value 1 (adjustable) Limit value 2 (adjustable) / maintenance request Failure Maintenance Measuring range (automatic measuring range change-over) Display dust content in mg/m³ i.N.dr. possible

Function principle The special probe of the PFM 97 consists of 2 tribo probes and a dynamic pressure probe: The tribo probes register redundantly the raw signal of the dust concentration. In order to correct the velocity influence at the triboelectric measurement the gas velocity is measured by means of a dynamic pressure probe. The synchronous determination of the gas temperature allows the calculation of the dust concentration in norm state. In case of high dust contents the dynamic pressure probe of the PFM 97 can be cleaned with the PCS 03. Technical data Control unit:

Weather-proof aluminium case, dimensions 305 x 240 x 300 mm (W x H x D), 3 kg, IP 65

Probe:

GRP weather protection hood, dimensions 300 x 400 x 1000 mm (W x H x D), ca. 10 kg

Measuring principle:

Dust: redundant measurement with 2 triboelectric probes; flow: differential pressure; temperature: Pt 100

Measuring ranges:

Temperature: 0 ... 300 °C, velocity: 0 ... 30 m/s flow: 0 ... 1.000.000 m³/h, dust i.O.: 0 ... 20 (max. 500) mg/m³, Dust i.N.: 0 ... 20/45/150/500 mg/m³

Calibration:

By gravimetric dust measurements

Display:

4-lines LCD-Display

Media temperature:

max. 280 °C (higher temperatures on request)

Ambient temperature:

-20 ... +50 °C

Dew point difference:

min. +5 K

Flow velocity:

From appr. 3 m/s

Analogue signals:

5 x 4 ... 20 mA, (including 2 x dust, temperature, flow) , galvanically isolated

Digital signals:

6 potential-free contacts (failure, maintenance, limit value 1 and 2 / maintenance request, measuring range)

Power supply:

110 VAC, 230 VAC or 24 VDC

TÜV-approval:

TI Air, 13th and 27th BImSchV

4th Edition, Conditions March 2010

35


Product catalog Dr. Födisch AG :: Specifications

DUST MEASURING DEVCIE

4.2.2.6 Dust measuring device PFM 97 W Basic device consisting of probe and control unit Combined device – measurement of dust, exhaust gas velocity or flow as well as temperature (option pressure) at the same time Use in corrosive gases (preferably in waste incinerations) Simple assembly: Control unit: wall assembly

TÜV-approved according to TI Air, 13th , 17th and 27th BImSchV

Fields of application

The dust concentration measuring device PFM 97 W is used for the continuous measurement of dust contents, especially for the emission measurement according to TI Air, 13th, 17th and 27th BImSchV as well as for the dust concentration measurement in process monitoring. Parallel to the dust measurement flow and temperature of the smoke gas are registered.

Probe: special flange DN 80, PN 6, Di=100 mm Analogue outputs: 4 ... 20 mA Digital outputs (potential-free): Limit value 1 (adjustable) Limit value 2 (adjustable) / maintenance request Failure Maintenance range (automatic measuring range change-over) Measuring

Display of dust content in mg/m³ i.N.dr. possible

Function principle

The special probe of the PFM 97 consists of 2 tribo probes and a dynamic pressure probe: The tribo probes register redundantly the raw signal of the dust concentration. In order to correct the velocity influence at the triboelectric measurement the gas velocity is measured by means of a dynamic pressure probe. The synchronous determination of the gas temperature allows the calculation of the dust concentration in norm state. In case of high dust contents the dynamic pressure probe of the PFM 97 can be cleaned with the PCS 03.

Technical data Control unit:

Weather-proof aluminium case, dimensions 305 x 240 x 300 mm (W x H x D), 3 kg, IP 65

Probe:

GRP weather protection hood, dimensions 300 x 400 x 1000 mm (W x H x D), ca. 10 kg

Measuring principle:

Dust: redundant measurement with 2 triboelectric probes; flow: differential pressure; temperature: Pt 100

Measuring ranges:

Temperature: 0 ... 300 °C, velocity: 0 ... 30 m/s flow: 0 ... 1.000.000 m³/h, dust i.O.: 0 ... 15 (max. 500) mg/m³, Dust i.N.: 0 ... 15/45/150/500 mg/m³

Calibration:

By gravimetric dust measurements

Display:

4-lines LCD-Display

Media temperature:

max. 280 °C (higher temperatures on request)

Ambient temperature:

-20 ... +50 °C

Dew point difference:

min. +5 K

Flow velocity:

From appr. 3 m/s

Analogue signals:

5 x 4 ... 20 mA, (including 2 x dust, temperature, flow) , galvanically isolated

Digital signals:

6 potential-free contacts (failure, maintenance, limit value 1 and 2 / maintenance request, measuring range)

Power supply:

110 VAC, 230 VAC or 24 VDC

TÜV-Approval:

TI Air, 13th, 17th and 27th BImSchV

36

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

DUST MEASURING DEVICE

4.2.2.7 Dust measuring device PFM 97 ED Special device consisting of probe and control unit Extractive dust measurement in wet and sticky gases (Measurement of a diluted and heated partial flow) Assembly: Control unit: wall assembly or floor mounting / frame Probe: special flange DN 80, PN 6, Di=100 mm Analogue outputs: 4 ... 20 mA Digital outputs (potential-free): Limit value 1 (adjustable) Limit value 2 (adjustable) / maintenance request Failure Maintenance Measuring range (automatic measuring range change-over)

TÜV-approved according to TI Air, 13th , 17th and 27th BImSchV

Fields of application The dust concentration measuring device PFM 97 ED is used for the continuous extractive measurement of dust contents in wet and sticky exhaust gases. Function principle By means of a temperature-regulated probe the measuring gas is extracted from the process and led to the measuring chamber where the triboelectric dust sensors are placed. In order to produce evaluable measuring signals and to protect the triboelectric measuring cell respectively the gas lines of the measuring device the measuring gas sucked off is continuously heated and diluted with dry and dust-free ambient air. The active principle of the dust measurement is based on the triboelectric effect. Therefore 2 electrically isolated half-shells are arranged in a cylindric chamber (measuring cell) and are passed by the conditioned measuring air. The triboelectric signals of the sensors are converted to equivalent dust signals in the electronic of the control unit. Technical Data Control unit:

Steel-sheet case mounted on profile frame (incl. Gebläse)

Probe:

Extractive sampling with GRP weather protection hood

Measuring principle:

Dust: redundant measurement with 2 triboelectric sensors

Measuring ranges:

Dust i.O.: 0 ... 15 (max. 500) mg/m³

Calibration:

By gravimetric dust measurements

Display:

4-lines LCD-Display

Media temperature:

max. 280 °C (higher temperatures on request)

Ambient temperature:

-20 ... +50 °C

Dew point difference:

min. +5 K

Flow velocity:

independent

Analogue signals:

5 x 4 ... 20 mA, (including 2 x dust, temperature, flow) , galvanically isolated

Digital signals:

6 potential-free contacts (failure, maintenance, limit value 1 and 2 / maintenance request, measuring range)

Power supply:

400 VAC, 50 Hz, 3~

TÜV-approval:

TI Air, 13th, 17th and 27th BImSchV

4th Edition, Conditions March 2010

37


Product catalog Dr. Födisch AG :: Specifications

4.2.2.8 Dust measuring device PFM 06 ED

DUST MEASURING DEVICE

Special device consisting of probe and control unit Extractive dust measurement in wet and sticky gases (Measurement of a diluted and heated partial flow) Assembly: Control unit: wall assembly or floor mounting / frame Probe: special flange DN 80, PN 6, Di=100 mm Measuring principle: laser optics Analogue outputs: 4 ... 20 mA Failure Maintenance Maintenance request Limit value 1 (adjustable) Limit value 2 (adjustable) Measuring range (automatic measuring range change-over)

Fields of application The dust concentration measuring device PFM 06 ED is used for the continuous extractive measurement of dust contents in wet and sticky exhaust gases. Function principle By means of a temperature-regulated probe the measuring gas is extracted from the process and led to the measuring chamber where the optic dust sensor is placed. In order to produce evaluable measuring signals and to protect the measuring cell respectively the gas lines of the measuring device the measuring gas sucked off is continuously heated and diluted with dry and dustfree ambient air. The active principle of the dust measurement is based on scattered light. Therefore a laser rod is arranged in a cylindric chamber (measuring cell) and passed by the conditioned measuring gas. The signal of the optic unit is converted into an equivalent dust signal in the control unit. Technical Data Control unit:

Steel-sheet case mounted on profile frame (incl. Gebläse)

Probe:

Extractive sampling with GRP weather protection hood

Measuring principle:

Dust: measurement with 1 optic sensor

Measuring ranges:

Dust i.O.: 0 ... 15 (max. 500) mg/m³

Calibration:

By gravimetric dust measurements

Display:

4-lines LCD-Display

Media temperature:

max. 280 °C (higher temperatures on request)

Ambient temperature:

-20 ... +50 °C

Dew point difference:

min. +5 K

Flow velocity:

independent

Analogue signals:

5 x 4 ... 20 mA, (including 2 x dust, temperature, flow) , galvanically isolated

Digital signals:

6 potential-free contacts (failure, maintenance, limit value 1 and 2 / maintenance request, measuring range)

Power supply:

400 VAC, 50 Hz, 3~

TÜV-approval:

none

38

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

DUST MEASURING DEVICE

4.2.2.9 Dust measuring device PFM 02 V

TÜV-approved according to TI Air, 13th , 17th and 27th BImSchV Limit value 1

On Off

2

100%

0.5s

0

1min

%: 12.9

0

1

2

3

Gain

Compact device consisting of probe Dust measurement in combination with flue gas velocity measurement ,e.g. FMD 02 Simplest assembly: Probe: 1"-sleeve (DIN 2986) optionally flange DN 25 PN 6 Analogue outputs: Dust: 4 ... 20 mA Flow: 4 ... 20 mA Digital outputs (potential-free): Limit value 1 (adjustable) Limit value 2 (adjustable)/ maintenance request Failure/Maintenance 4 measuring range adjustable Display in mg/m³ or as online line diagram Target value calibration Simplest probe rod adjustment to the measuring conditions on site (probe rod turnable and exchangeable)

Application The dust concentration measuring device PFM 02 V is used for the continuous measurement of dust contents ,

especially for the emission measurement according to TI Air, 13th, 17th and 27th BImSchV as well as for the dust concentration measurement in process monitoring.. Parallel to the dust measurement flow and temperature of the smoke gas are registered (in combination e.g. with FMD 02).

Special features Target value calibration: Enter a known, averaged dust content in operation state - The PFM 02 V determines automatically appropriate calibration factors and shows the quantitative dust content in mg/m³. Technical data Control unit:

complete electronic in probe head (no separate control unit required), IP 65

Probe:

Stainless-steel probe rod (turnable and exchangeable) at the probe head, probe length 300 mm (adjustments possible), IP 65

Measuring principle:

Dust: Measurement with 1 triboelectric sensor

Measuring ranges:

0 ... 100 % respectively 0 ... 10 (1.000) mg/m³

Calibration:

By gravimetric reference measurements

Display:

Graphic display with online line diagram

Media temperature:

max. 280 °C (higher temperatures on request)

Ambient temperature:

-20 ... +50 °C

Dew point difference:

min. +5 K

Flow velocity:

From appr. 3 m/s

Analogue signals:

2 x 4 ... 20 mA, galvanically isolated

Digital signals:

3 potential-free contacts (failure/maint, limit value 1 and 2/maint. request)

Power supply:

110 VAC, 230 VAC or 24 VDC

TÜV-approval:

TI Air, 13th , 17th and 27th BImSchV (in combination with the velocity measurement)

4th Edition, Conditions March 2010

39


Product catalog Dr. Födisch AG :: Specifications

4.2.2.10 References Device

References (Extract)

Number of installed devices

PFM 97 bzw. PFM 97 W

Alstom Power Boiler GmbH, Delitzsch, Emden Biomasse HKW, Monopoli (Italien) Biomasse HKW, Malchin EVO AG, Offenbach EKO Stahl GmbH, Eisenhüttenstadt Fritz Egger GmbH, Wismar Fritz Egger GmbH, Unterradlberg (Österreich) Heraeus Quarzglas, Bitterfeld HKW Großräschen HKW Recklinghausen HKW Stavenhagen HKW Wicker Krematorium, Heidelberg Krematorium, Meißen Krematorium, Ulm Krupp Thyssen Nirosta, Krefeld Lühr, Stadthagen MEAG HKW, Wolfen Nanjing Electric Power, Nanjing (China) Norddeutsche Affinerie, Hamburg Pfleiderer AG, Rheda-Wiedenbrück Pfleiderer AG, Gütersloh Resch und Söhne, Schwerte Wacker Chemie GmbH, Burghausen WESTAG + GETALIT, Rheda-Wiedenbrück

> 200

PFM 97 ED

Kali + Salz AG Salzdetfurth Berliner Wasserbetriebe Bilfinger + Berger, Bremen BASF AG, Ludwigshafen BBP AE Energietechnik AG, Mainz EGN Entsorgungsgesellschaft Niederrhein mbH, WSAA Neuss E.ON Kraftwerke GmbH, Wilhelmshaven Kronospan GmbH, Bischweier KW Maritza II, (Bulgarien) Magmalor GmbH, Colditz MHKW, Mainz NIER, Seoul (Korea) Oy Anatec Instruments AB, Kerava (Finnland) Pfleiderer AG, Gütersloh Stadtentwässerung, Stuttgart Südchemie AG, Moosburg UHDE, Doha (Quatar) URSA GmbH, Delitzsch Wacker Chemie GmbH, Nünchritz Zellstoff Stendal GmbH

> 60

PFM 02 V

DK Jethwa, Calcutta (Indien) MBA Osnabrück MBA Rennerod MBS Vogtland, Oelsnitz Norddeutsche Affinerie, Hamburg Heye Glas GmbH Waste Tec, Trier

> 25

40

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

4.2.3

Gas Analysers

MGA 23

MGA 05

MCA 04

MCA 04 analysis cabinet

O2MD 04

HMD 06

BGM 06

AMD 08

MCA 04 mobil

4.2.3.1 Measuring principle MGA 23 By means of the gas analyser MGA 23 up to 4 gas components can be measured at the same time: maximum 3 infrared-active gases as e.g. CO, CO2, NO, SO2, CH4, R22 (Frigen CHCIF2) as well as O2 with an electrochemical oxygen measuring cell. Basic versions of MGA 23 for: 1 Infrared-gas component with/without oxygen measurement 2 Infrared-gas components with/without oxygen measurement 3 Infrared-gas components with/without oxygen measurement Smallest TÜV-approved and permitted measuring ranges: 1- and 2-component-analyser CO: 0 to 150 mg/m³ NO: 0 to 250 mg/m³ SO2: 0 to 400 mg/m³ 3-component-analyser CO: 0 to 250 mg/m³ NO: 0 to 400 mg/m³ SO2: 0 to 400 mg/m³ All higher measuring ranges are also allowed. Special configurations: The MGA 23 with 2 IR-components without pump and with or without oxygen is also available with 2 separate gas ways. This allows the measurement of 2 measuring points or e.g. at the NOx-measurement the state before and after converter.

4th Edition, Conditions March 2010

41


Product catalog Dr. Fรถdisch AG :: Specifications Two independent, selectively working measuring principles are used in the MGA 23: 1. Infrared measurement NDIR Infrared measurement This spectroscopic method is based on the absorption of non-dispersive IR radiation. The attenuation in the radiation which depends on the wavelength is a measure of the respective concentration of the gas.

12 1

10

2 3 9 11 9 Sample gas outlet

4 Sample gas inlet

9

5

8

6

M

7 1 2 3 4 5 6

Capillary Second detector layer Microflow sensor Sample cell Chopper wheel Chopper motor

7 8 9 10 11 12

IR source Reflector Window Slide First detector layer Third detector layer

2. Electro-chemically O2 Sample gas

6

1

5

2

Oxygen measurement The oxygen sensor operates according to the principle of a fuel cell. The oxygen is converted at the boundary layer between the cathode and electrolyte; the resulting current is proportional to the concentration of oxygen.

e-

U 4 1 Gold cathode 2 Electrolythe (acetic acid) 3 Thermistor and load resistor for temperature compensation

42

3 4 Signal output 5 Lead anode 6 Oxygen diffusion membrane made of FEP

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

4.2.3.2 Measuring principle MGA 09 / BGM 06 The MultiGasAnalyser MGA 09 is a highly-sensitive system for continuous measurement of gas components in exhaust gases of the industry. It can measure up to 4 components simultaneoulsy. The optic bench for the measurement of the infrared-active components consists primarily of an infrared source, a measuring cell and a detector. In the MGA 09 two independent, selectively working measuring principles are used:

Gasaustritt (... zum O2-Sensor)

Gaseintritt

Küvette Grundplatte Strahler

R1 R2 Empfänger mit Thermostat

Vorverstärkerelektronik

1. Infrared measurement NDIR Infrared absorption This spectroscopic method is based on the absorption of non-dispersive IR radiation. The attenuation in the radiation which depends on the wavelength is a measure of the respective concentration of the gas.

2. Electro-chemically O2 Oxygen measurement (Example) The oxygen sensor operates according to the principle of a fuel cell. The oxygen is converted at the boundary layer between the cathode and electrolyte; the resulting current is proportional to the concentration of oxygen.

4.2.3.3 Measuring principle MCA 04 By means of the gas analyser MCA 04 8 gas components can be measured at the same time: maximum 7 infrared-active gases (e.g. CO, NO, NO2, SO2, HCl, NH3, H2O, CO2) and O2 with an extractive zirconium oxide cell. The analysis system MCA 04 is an extractive, continuous measuring Smallest measuring ranges: Max. 7 infrared components + O2 system with an excellent cost-performance ratio. It is not only suited for the use on raw and clean gas side but also as process measuring + TOC (optional): device in the same manner. 4th Edition, Conditions March 2010

43


Product catalog Dr. Födisch AG :: Specifications By means of the compact gas analyser MCA 04 as basis the analysis CO 0 ... 75 mg/m³ system has all necessary pre-conditions for that. Moreover the NO 0 ... 200 mg/m³ analysis system can be extended at the flue gas component total 0 ... 100 mg/m³ NO2 organic carbon (option). 0 ... 75 mg/m³ SO2 The analysis system MCA 04 extracts a partial gas flow from the flue HCl 0 ... 15 mg/m³ gas which is led to the analyser in a heated way (all heated 0 ... 15 mg/m³ NH3 components of the measuring system are regulated on 185 °C ). 0 ... 40 Vol% H2O The proven gas sampling as well as the gas conditioning (flow rate 0 ... 20 Vol% CO2 200 ... 600 l/h) and most advanced photometer technology grant a 0 ... 21 Vol% O2 high reliability and long operating times with short maintenance intervals. The measuring values as well as all status and operational messages are shown on the display of the analyser module. All required adjustments of the system can be made by means of the keyboard. Via integrated modem a remote diagnosis of the analysis system done by the service staff is possible. Two independent, selectively working measuring principles are used in the MCA: Single-Beam Dual-Wavelength Method Meas. filter

IR Source

The spectral ranges are selected byalternately inserting the interference filters for absorption (measuring filter I) and non-absorption (reference filter Io). The electronics of the MCA 04 calculates I and Io with the absorbance value A and determines the concentration value.

Meas. cell Reference filter

Gas filter correlation Gas filter

IR Source

Meas. cell

Interference filter Free aperture

44

When the Gas Filter correlation method is employed, the reference signal Io, which is independent from the concentration, is generated by inserting a gas filter in the light path. This gas filter is a miniature cell filled with the measuring component under high partial pressure. Measuring signal I, which is dependent on the concentration, is obtained by inserting an empty filter wheel aperture into the beam path. The influence of maybe interfering components is almost identical on both channels. Cross-sensitivities are, therefore effectively eliminated.

4th Edition, Conditions March 2010


Product catalog Dr. FĂśdisch AG :: Specifications

4.2.3.4 Measuring principle O2MD 04 The oxygen measuring device O2MD 04 consists of an in-situ-probe with evaluation electronics and measuring value display in the probe head. The measuring gas flows through the measuring chamber of the probe rod. It can be applied for the measurement of the oxygen concentration in smoke, process and cover gases as well as for the mixture and production of special inert gases. Then the concentration of the free oxygen is measured. The core piece of the oxygen measuring device O2MD 04 is a potentiometric ZrO2-sensor. The probe is equipped with a regulated sensor heating and a high-quality electronic which provides the sensor signal (= oxygen concentration) as standard signal.

The oxygen measuring device O2MD 04 is designed for the in-situ – measurement in gases. The advantage of the direct measurement without sampling compared to extractive measuring methods with gas sampling and gas conditioning lies in the signal reception being virtually without delay (t90 < 1 seconds), by that many control and automation procedures become only possible. Another advantage of the direct measurement results from the omission of any maintenance for the gas sampling and gas conditioning. Massive board electrodes and their ceramic covers lead to high life-times as the re-crystallisation occurrences do not limit the electrodes’ lifetime as for layers and on the other hand the volatilisation of the platinum is reduced. A further advantage is that the probe does not have to be calibrated.

4th Edition, Conditions March 2010

45


Product catalog Dr. Fรถdisch AG :: Specifications

4.2.3.5 Measuring principle HMD 06 The HMD 06 uses the physical correlation between electric power of a peltier element and the humidity content of the media to be cooled. Therefore diverse electric and thermal parameters are registered at a modified measuring gas cooler (designed for flue gases in industry): Temperature at gas entry, Temperature of heat sink, Ambient temperature, Temperature at gas exit, Flow and Current-voltage-curve. The evaluation electronic of the device determines indirectly the humidity content of the gas to be cooled. Qe

Qel

Balance limit IV

Qcool Heat exchanger

QC

Peltier element

QH

Heat sink Qa

Balance limit I

Balance limit II

Balance limit III

Apart from the function as measuring device the HMD 06 can be used in the gas conditioning for a cold emission measurement.

46

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

4.2.3.6 Applications

MGA 23 [extractive]

Multi-component gas analyser for the application in official and operational emission measuring systems as well as for: Optimisation of small firing systems Monitoring of exhaust gas concentration from firing systems with different types of fuel (oil, gas and coal) as well as operational measurements in thermal incineration plants Monitoring of process control functions

MGA 05 [extractive]

CO: 0 ... 1.000 mg/m³ Multi-component gas analyser for the use in operational emission measuring NO: 0 ... 100 ppm/ systems 2.000 mg/m³ Max. 3 Infrared components + O2 SO2: 0 ... 200 mg/m³ CO2: 0 … 20 Vol% O2: 0 ... 25 Vol%

MCA 04 [extractive]

O2MD 04 [in-situ]

Multi-component gas analyser for the application in official and operational emission measuring systems e.g. in: Waste incinerations Biomass power plant Refineries Process measurements

Oxygen measuring system for operational measuring tasks

1- and 2-comp. analyser: CO: 0 ... 150 mg/m³ NO: 0 ... 250 mg/m³ SO2: 0 ... 400 mg/m³ O2: 0 ... 25 Vol% 3-comp.-analyser: CO: 0 ... 250 mg/m³ NO: 0 ... 400 mg/m³ SO2: 0 ... 400 mg/m³ O2: 0 ... 25 Vol%

CO: 0 ... 75 mg/m³ NO: 0 ... 200 mg/m³ NO2: 0 … 100 mg/m³ SO2: 0 ... 75 mg/m³ HCl: 0 … 15 mg/m³ NH3: 0 … 15 mg/m³ H2O: 0 … 40 Vol% CO2: 0 … 20 Vol% O2: 0 ... 21 Vol% O2: 0 … 20,6 Vol%

Continuous humidity measuring device for the use in CEMS and for process measurements In addition the measuring device can be used as gas conditioning for a cold emission measurement.

Dew point temperature: 25 ... 74 °C Spezific humidity: 20 ... 350 g/kg Volume portion H2O: 3 ... 36 Vol% Relative humidity: 0 ... 100 %

BGM 06 [extraktiv]

Continuous bio gas measurement e.g. for: Process monitoring and optimisation Determination of energy content (CH4-measurement) Protection of engines (H2S measurement) Bio gas plants, dry fermentation

CO2: CH4: H2S O2:

AMD 08 [extraktiv]

Continuous acid dew point measuring STP: 0 ... 300 °C device for operational tasks like process optimisation e.g. in coal power plants

HMD 06 [extraktiv]

4th Edition, Conditions March 2010

27th BImSchV

Smallest measuring ranges

17th BImSchV

Applications

13th BImSchV

Device

TI Air

TÜV -Approval

0 ... 50 Vol% 0 ... 100 Vol% 0 ... 10.000 ppm 0 ... 25 Vol%

47


Product catalog Dr. Fรถdisch AG :: Specifications

Feeding

230 VAC, 50/60 Hz

Feeding

230/110 VAC, 50/60 Hz

Feeding

24 VDC

Analogue output

4 ... 20 mA

4

4

8

1

2

Digital outputs

8

max. 16

Option

Digital outputs

3

max. 12

Option

Approval

TI Air

Aproval

13 BImSchV

Approval

17 BImSchV

Approval

27 BImSchV

Approval

MCERTS

48

AMD 08

400 VAC, 3~

BGM 06

Feeding

HMD 06

Extractive

O2MD 04

Measuring arrangement

MCA 04

In-situ

MGA 05

Measuring arrangement

MGA 23

4.2.3.7 Gas analysers overview

4

4

3

3

max. 11

6

max. 4

1

th th th

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

GAS ANALYSER

4.2.3.8 Gas analyser MGA 23

TÜV-approved according to TI Air, 13th and 27th BImSchV

Field of application By means of the gas analyser MGA 23 up to 4 gas components can be measured at the same time: maximum 3 infrared-active gases as e.g. CO, CO2, NO, SO2, CH4, R22 (Frigen CHCIF2) as well as O2 with an electrochemical measuring cell.

19“-rack for the application in measuring systems for the continuous emission measurement after gas cooler Cold measurement Simultaneous measurement of max. 3 Infrared components (CO, CO2, NO, SO2, CH4, R22) and oxygen O2 Simple assembly in the analysis cabinet Analogue output: max. 4 x 4 ... 20 mA Digital outputs (potential-free): Limit value 1 (adjustable) Limit value 2 (adjustable) / maint. request Failure/Maintenance Measuring range adjustable Low in maintenance due to automatic calibration with ambient air Check with test gas only every 6 to 12 months

Application The gas analyser MGA 23 can be used in emission measuring facilities as well as in systems of process and analysis measuring technology. For measurements of CO, NO, SO2 and O2 according to 13th respectively 27th BlmSchV and TI Air TÜVapproved versions of the MGA 23 are available. Technical data Case:

19"-rack, 4 HU, dimensions 177 x 483 x 339 mm (W x H x D), 10 kg, IP 21

Measuring principle:

CO, CO2, NO, SO2, CH4, R22 (Frigen CHCIF2): Infrared absorption O2: electro-chemical measuring cell

Smallest measuring ranges:

1- and 2-components-analysers: CO: 0 ... 150 mg/m³, NO: 0 ... 250 mg/m³, SO2: 0 ... 400 mg/m³, O2: 0 ... 25 Vol%

Calibration:

Automatic calibration with ambient air resp. N2

Display:

LCD with LED-back-light and contrast regulation, functional keys, 80 characteristics (4 lines/20 characteristics)

Media temperature:

0 ... 50 °C

Ambient temperature:

+5 ... +45 °C

Interfaces:

RS 485, optionally Profibus

Analogue signals:

max. 4 x 4 ... 20 mA

Digital signals:

8 potential-free contacts

Digital inputs:

3 potential-free contacts

Power supply:

110 VAC, 230 VAC

TÜV-Approval:

TI Air, 13th and 27th BImSchV

4th Edition, Conditions March 2010

49


Product catalog Dr. Födisch AG :: Specifications

4.2.3.9 Gas analyser MGA 09

GAS ANALYSER

19“-rack for use in measuring systems for continuous emission measurement behind gas cooler Cold measurement Simultaneous measurement of max. 3 infrared components (CO, CO2, NO, SO2, H2O behind cooler) and 2 electro-chemical cells (oxygen O2 resp. NO) Simple assembly in an analysis cabinet Analogue outputs: max. 5 x 4 ... 20 mA Digital outputs (potential-free): Limit value 1 (adjustable) Limit value 2 (adjustable) / maintenance request Failure/Maintenance 2 measuring ranges selectable with digital contact

Field of application By means of the gas analyser MGA 09 up to 4 gas components can be measured at the same time: maximum 3 infrared-active gases e.g.. CO, CO2, NO, SO2 , H2O behind cooler as well as O2 resp. NO with an electrochemical measuring cell. Application The gas analyser MGA 09 can be applied in emission measuring systems as well as systems for process and analysis measuring technology. Technical data Case:

19"-rack, 3 HU, dimensions 132,6 x 482,6 x 354 mm (H x B x T), ca. 5,6 kg, IP 20

Measuring principle:

CO, CO2, NO, SO2, H2O: Infrared absorption O2, NO: electro-chemical measuring cell

Smallest measuring ranges:

CO: 0 ... 200 mg/m³, NO: 0 ... 600 mg/m³, SO2: 0 ... 600 mg/m³, CO2: 0 ... 1.000 ppm H2O (residual moisture behind cooler): 0 ... 3 Vol%, O2: 0 ... 25 Vol%

Calibration:

Manual calibration with ambient air respectively N2 and calibration gases

Display:

LCD with LED-backlight and contrast control, function keys, (240 x 128 Pixel)

Media temperature:

0 ... 25 °C

Cell:

Heated at 55°C No mechanically moved parts

Ambient temperature:

+5 ... +40 °C

Interfaces:

RS 232 (subject to service)

Analogue signals:

max. 4 x 4 ... 20 mA

Digital signals:

Max. 16 potential-free contacts

Digital inputs:

Max. 12 potential-free contacts

Power supply:

110 VAC, 230 VAC

TÜV-Approval:

none

50

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

GAS ANALYSER

4.2.3.10 Gas analyser MCA 04

TÜV-approved according to TI Air, 13th , 17th and 27th BImSchV

Field of application By means of the gas analyser MCA 04 up to 7 gas components can be measured at the same time - as e.g. z.B. CO, NO, NO2, SO2, HCl, NH3, H2O, CO2 – as well as O2.

19"-rack for application in measuring systems for continuous emission measurement Hot measurement (without gas cooler) Simultaneous measurement of max. 7 Infrared components (e.g. CO, NO, NO2, SO2, HCl, NH3, H2O, CO2) and oxygen O2 Simple assembly in the analysis cabinet Analogue outputs: max. 8 x 4 ... 20 mA Digital outputs (potential-free): Adjustable for failure and maintenance Remote maintenance possible if in analysis system applied

Application The gas analyser MCA 04 can be used in emission measuring facilities as well as in systems of process and analysis measuring technology. The MCA 04 is for measurements according to 13th, 17th and 27th BlmSchV and TI Air suited. Technical data Case:

19"-rack, 4 HU, dimensions 665 x 440 x 360 mm (B x H x T), ca. 40 kg, IP 52

Measuring principle:

CO, NO, NO2, SO2, HCl, NH3, H2O, CO2: Infrared absorption (Single-Beam Dual-Wavelength Method respectively gas filter correlation) O2: ZrO2-cell

Smallest measuring ranges:

CO: 0 ... 75 mg/m³, NO: 0 ... 200 mg/m³, NO2: 0 … 100 mg/m³, SO2: 0 ... 75 mg/m³, HCl: 0 … 15 mg/m³, NH3: 0 … 15 mg/m³, H2O: 0 … 40 Vol%; CO2: 0 … 20 Vol%, O2: 0 ... 21 Vol%

Calibration:

Zero point: automatically with ambient air; sensitivity: with calibration gas

Display:

7.4" black/white LC-Display (640*480 Pixel)

Media temperature:

Max. 200 °C

Ambient temperature:

+5 ... +35 °C

Interfaces:

RS 232, optionally MOD-Bus

Analogue signals:

max. 8 x 4 ... 20 mA

Digital signals:

D/O: for failure, maintenance, maintenance request and measuring range signalisation available (230 V, 1 A)

Digital inputs:

Optionally for analogue and digital signals

Power supply:

110 VAC, 230 VAC

TÜV-approval:

TI Air, 13th, 17th and 27th BImSchV

4th Edition, Conditions March 2010

51


Product catalog Dr. Födisch AG :: Specifications

GAS ANALYSER

4.2.3.11 Gas analysis system MCA 04

With TÜV-approved analysers according to TI Air, 13th , 17th and 27th BImSchV

Field of application The analysis system MCA 04 is an extractive continuous measuring system with an excellent cost-performance ratio. It is not only suited for the use on raw and clean gas side but also as process measuring device in the same manner.

Analysis cabinet for installation in measuring systems for continuous emission measurement Hot measurement (without gas cooler) Simultaneous measurement of max. 7 Infrared components (e.g. CO, NO, NO2, SO2, HCl, NH3, H2O, CO2) and oxygen O2 Optional measurement of TOC (total organic carbon) via FID Simple installation in the analysis container or at the measuring point Analogue outputs: max. 8 x 4 ... 20 mA Digital outputs (potential-free): for failure and maintenance adjustable Remote maintenance via modem possible

Application Multi-component gas analysis system for the application in official and operational emission measuring systems e.g. for: Monitoring of exhaust gas concentration from firing systems with different types of fuel (oil, gas, coal biomass, substitute fuels etc.) as well as in thermal incineration plants Optimisation of small firing systems

Monitoring of processes Technical data Case:

Control cabinet, dimensions 800 x 2.100 x 600 mm 200 ... 300 kg depending on equipment, IP 54

Measuring principle:

CO, NO, NO2, SO2, HCl, NH3, H2O, CO2: Infrared absorption (Single-Beam Dual-Wavelength Method respectively gas filter correlation) O2: ZrO2-cell

Smallest measuring ranges:

CO: 0 ... 75 mg/m³, NO: 0 ... 200 mg/m³, NO2: 0 … 100 mg/m³, SO2: 0 ... 75 mg/m³, HCl: 0 … 15 mg/m³, NH3: 0 … 15 mg/m³, H2O: 0 … 40 Vol%; CO2: 0 … 20 Vol%, O2: 0 ... 21 Vol%

Calibration:

Zero point: automatically with ambient air; sensitivity: with calibration gas

Display:

7.4" black/white LC-Display (640*480 Pixel)

Media temperature:

Max. 200 °C

Ambient temperature:

+5 ... +35 °C

Interfaces:

RS 232, optionally MOD-Bus

Analogue signals:

max. 8 x 4 ... 20 mA

Digital signals:

D/O: for failure, maintenance, maintenance request and measuring range signalisation available (230 V, 1 A)

Digital inputs:

Optionally for analogue and digital signals

Power supply:

400 VAC, 50 Hz, 3~

TÜV-approval:

TI Air, 13th, 17th and 27th BImSchV

52

(W x H x D),

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

GASANALYSATOR

4.2.3.12 Mobile gas analysis system MCA 04

With TÜV-approved analysers according to TI Air, 13th , 17th and 27th BImSchV

Field of application The mobile gas analysis system MCA 04 is an extractive continuous measuring system with an excellent costperformance ratio. It is not only suited for the use on raw and clean gas side but also as process measuring device in the same manner.

Mobile gas analysis system for application in measuring systems for continuous emission measurements Hot measurement (without cooler) Simultaneous measurement of max. 7 Infrared components (e.g. CO, NO, NO2, SO2, HCl, NH3, H2O, CO2) and oxygen O2 Simple installation at the measuring point Analogue outputs: 8 x 4 ... 20 mA (optionally further analogue outputs) Digital outputs (potential-free): for failure, maintenance, maitenance request and measuring ranges adjustable Use at different sites respectively as exchange unit Integrated electronic recorder for measuring value recording and transfer to PC

Application Multi-component gas analysis system for the application in official and operational emission measuring systems e.g. for: Monitoring of exhaust gas concentration from firing systems with different types of fuel (oil, gas, coal biomass, substitute fuels etc.) as well as in thermal incineration plants Optimisation of small firing systems

Monitoring of processes Technical data Case:

Mobile case, dimensions 530 x 450 x 565 mm (W x H x D, ca. 59 kg depending on configuration, IP 54

Measuring principle:

CO, NO, NO2, SO2, HCl, NH3, H2O, CO2: Infrared absorption (Single-Beam Dual-Wavelength Method respectively gas filter correlation) O2: ZrO2-cell

Smallest measuring ranges:

CO: 0 ... 75 mg/m³, NO: 0 ... 200 mg/m³, NO2: 0 … 100 mg/m³, SO2: 0 ... 75 mg/m³, HCl: 0 … 15 mg/m³, NH3: 0 … 15 mg/m³, H2O: 0 … 40 Vol%; CO2: 0 … 20 Vol%, O2: 0 ... 21 Vol%

Calibration:

Zero point: automatically with ambient air; sensitivity: with calibration gas

Display:

7.4" black/white LC-Display (640*480 Pixel)

Media temperature:

Max. 200 °C

Ambient temperature:

+5 ... +35 °C

Interfaces:

RS 232, optionally MOD-Bus

Analogue signals:

8 x 4 ... 20 mA (optionally further analogue outputs)

Digital signals:

D/O: for failure, maintenance, maintenance request and measuring range signalisation available (230 V, 1 A)

Digital inputs:

Optionally for analogue and digital signals

Power supply:

400 VAC, 50 Hz, 3~

TÜV-approval:

TI Air, 13th, 17th and 27th BImSchV

4th Edition, Conditions March 2010

53


Product catalog Dr. Födisch AG :: Specifications

4.2.3.13 Gas analyser O2MD 04

GAS ANALYSER

Compact device consisting of probe Simple assembly: Probe: 1"-sleeve (DIN 2986) Analog outputs: Oxygen: 4 ... 20 mA Digital outputs (potential-free): Limit value 1 (adjustable) Limit value 2 (adjustable) maintenance request Failure/Maintenance Online line diagram Display in Vol% Limit value

Limit value 1

2

On Off

1

2

20,6%

O2 [Vol%]

On Off 0.5s

5s

12.9

/

0

1min

%: 12.9

Field of application The O2MD 04 can be applied for the measurement of the oxygen concentration in smoke, process and cover gases as well as for the mixture and production of special inert gases. Then the concentration of the free oxygen is measured. Application and function The oxygen measuring device O2MD 04 consists of an in-situ-probe with evaluation electronics and measuring value display in the probe head. The measuring gas flows through the measuring chamber of the probe rod. The core piece of the oxygen measuring device O2MD 04 is a potentiometric ZrO2-sensor. The probe is equipped with a regulated sensor heating and a high-quality electronic which provides the sensor signal (= oxygen concentration) as standard signal. Technical data Case:

Control unit in the probe head (no separate control unit required), IP 65

Probe:

Stainless steel ZrO2-probe at the probe head, complete electronic in the probe head – no separate control unit, IP 65

Measuring principle:

Potentiometric ZrO2-cell (heated)

Measuring range:

0 ... 20,6 Vol%

Calibration:

calibration with test gas (2 ... 3 Vol% O2) with optional zero tube

Display:

Graphic display with online line diagram

Media temperature:

max. 300 °C

Ambient temperature:

-20 ... +50 °C

Analogue signal:

1 x 4 ... 20 mA

Digital signals:

3 potential-free contacts (failure/maint, limit value 1 and 2/maint. request)

Power supply:

24 VDC

TÜV-approval:

none

54

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

4.2.3.14 Gas analyser HMD 06

GAS ANALYSER

Compact analysis cabinet for use in measuring systems for continuous emission measurement (for installation in thermically stable environment) Measurement of humidity content in flue gases Simultaneous use as gas conditioning for a cold emission measurement 2 analogue outputs for humidity content in [g/kg], [Vol%], [°C] bzw. [Rel%]: 2 x 4 ... 20 mA Digital outputs (potential-free): Limit value 1 (adjustable) Limit value 2 (adjustable) / maintenance request Failure/maintenance

Field of application The humidity measuring device HMD 06 is a highly-sensitive device for continuous measurement of humidity concentration in the exhaust gas of combustion and industrial plants. Application and function The measuring gas is led to the cooler of the HMD 06 by means of a heated measuring gas line. The power of the cooler necessary for cooling the measuring gas is measured and calculated. In this way the humidity content can be registered continuously. The absolute humidity content is provided as 4 ... 20 mA – signal. It can be chosen between the output of the humidity signal in [g/kg], [Vol%], [°C] and [Rel%]. Technical data Case:

Compact cabinet, dimensions 800 x 600 x 210 mm (W x H x D) (without fixture), ca. 35 kg depending on configuration, IP 54

Probe:

Optional gas sampling probe and heated line

Measuring principle:

Electric power of a peltier element

Measuring range:

Dew point: 25 ... 74 °C, spez. humidity: 20 ... 350 g/kg, Volume portion: 3 ... 36 Vol%, rel. humidity: 0 ... 100 Vol%

Calibration:

Zero point: automatic (electric zerp point peltier element), reference point: manually with water steam

Display:

graphic display

Media temperature:

max. 300 °C

Ambient temperature:

+5 ... +35 °C

Analogue signal:

2 x 4 ... 20 mA

Digital signals:

3 potential-free contacts (failure/maintenance, limit value 1 and 2/ maintenance request)

Power supply:

230 VAC, 50 Hz, ca. 150 VA (excl. Measuring gas line and probe)

TÜV-approval:

none

4th Edition, Conditions March 2010

55


Product catalog Dr. Födisch AG :: Specifications

4.2.3.15 Bio gas monitor BGM 06

GAS ANALYSER

Compact cabinet with gas conditioning for the continuous bio gas measurement Cold measurement Simultaneous measurement of 2 infrared components (CH4 and CO2) and 2 electro-chemical cells (oxygen O2 resp.. H2S) Simple installation Analogue outputs: max. 4 x 4 ... 20 mA Digital outputs (potential-free): Limit value 1 (adjustable) Limit value 2 (adjustable) / maintenance request Failure/maintenance 2 Measuring ranges adjustable

Field of application 4 components can be measured simultaneously by means of the bio gas monitor BGM 06: max. 3 infrared active gases like e.g. CH4, CO2 as well as O2 and H2S with electro-chemical measuring cells. Application The bio gas monitor BGM 06 can be used in emission and process measuring systems in plants of bio gas generation. It is even possible to determine continuously aggressive gas components (H2S and CH4). Technical data Case:

Wall case, dimensions 610 x 750 x 480 mm (W x H x D), ca. 25 kg,

Probe:

Dilution probe for cold sampling, ca. 3 kg

Measuring principle:

CO2, CH4: Infrared absorption O2, H2S: electrochemical measuring cell

Probe:

Gas sampling via flame barrier incl. dilution of measuring gas with inert gas (1:50)

Smallest range:

CO2: 0 ... 100 Vol%, CH4: 0 ... 100 Vol%, H2S: 0 ... 10.000 ppm, O2: 0 ... 25 Vol%

Calibration:

Manual calibration with ambient air resp. N2 and calibration gases

Display:

LCD with LED-backlight and contrast control, function keys, (240 x 128 Pixel)

Media temperature:

5 ... 40 °C

Ambient temperature:

+5 ... +35 °C

Interfaces:

RS 232 (subject to service)

Analogue signal:

max. 4 x 4 ... 20 mA

Digital signals:

Max. 11 potential-free contacts

Digital inputs:

Max. 4 potential-free contacts

Power supply:

230 VAC

TÜV-approval:

none

56

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

4.2.3.16 Gas analyser AMD 08

GAS ANALYSER

Special device consisting of probe and control unit Extractive Acid dew point measurement (STP) Assembly: Control unit: wall assembly or installation on ground / frame Probe: spezial flange DN 80, PN 6, Di=100 mm Analogue outputs: 4 ... 20 mA Digital outputs (potential-free): Limit value 1 (adjustable) Limit value 2 (adjustable) Failure Maintenance Maintenance request

Field of application

The acid dew point measuring device AMD 08 offers an online monitoring of the acid dew point (STP) in flue gases of large combustion plants. Due to the continuous STP-measurement it is possible to reduce the flue gas temperature of power plants in a technologically and economically reasonable way. So far the exhaust gas temperature has been kept with a safety margin of up to 20 K above the expected acid dew point. By means of its measurement this safety margin can be minimised without having to expect plant damages by corrosion. The reduction of the exhaust gas temperature leads directly to an increase of the plant’s net efficiency.

Application e.g. as : to control the exhaust gas temperature to avoid acid-related corrosion in channels before flue gas cleaning systems to increase the efficiency and therefore to save fuel and reduce CO2-emissions Technical data Control unit:

Steel sheet case, mounted on frame (incl. blower)

Probe:

Extractive sampling with GRP protection hood

Measuring principle:

Acid dew point condensation in triboelectric measuring cell

Measuring ranges:

Acid dew point: 0 ... 300 °C

Calibration:

By reference measurements

Display:

4-line LCD-Display

Media temperature:

max. 200 °C (higher temperatures on request)

Ambient temperature:

-20 ... +50 °C

Dew point difference:

min. +5 K

Flow velocity:

Independent

Analogue signals:

4 x 4 ... 20 mA, (therefrom 1 x STP), galvanically isolated

Digital signals:

6 potential-free contacts (failure, maintenance, limit value 1 and 2 / maintenance request)

Power supply:

400 VAC, 50 Hz, 3~

TÜV-approval:

None

4th Edition, Conditions March 2010

57


Product catalog Dr. Födisch AG :: Specifications

4.2.3.17 References Device

References (Extract)

Number of installed devices

MGA 23

BERNT Messtechnik Biomasse HKW, Malchin Beijing Monitor Environment Technology, Beijing (China) Dalkia Energie Service Leinefelde DSK Kokerei Bottrop Energie und Wasser Potsdam GmbH, Potsdam Ekomesure (France) Ekosis (Tukey) HKW UNI Münster Ibemo GmbH (Kasakhstan) Kraftwerk Gera HKW Nord HKW Süd Krematorium: Berlin Braubach Fürstenzell Gröbers Lüdenscheid Rostock Sauerland Tornesch MATTERSTEIG & Co. mbH, Markranstädt Papierfabrik, Schwedt Raba SWT, Zella Mehlis Shenzhen Zhongxin Co. (China) Standard Fasel - Lentjes B.V., Southhampton (UK) TÜV Rheinland, Köln TÜV Süd

> 250

MGA 05

China National Precision Machinery Import Export Corp., Beijing (China) China Aerospace Greatwall Trading Company, Ltd. , Beijing (China) Universtar Science & Technology, Shenzhen, (China)

> 30

MCA 04

ATZ Entwicklungszentrum, Sulzbach-Rosenberg AVA Frankfurt Nordweststadt Baumgarte Boiler Systems GmbH, MHKW Stavenhagen Casella, Bedford (UK) Harpen Energie Contracting GmbH, Bergkamen GKW Bitterfeld-Wolfen Guardian Flachglas GmbH, Thalheim HEXAL Pharma GmbH, Radebeul KW Maritza III (Bulgaria) NIER, Seoul (Korea) Steag Saar Energie AG, HKW Dresden Stevens Croft Wood Burning, Power Plant Lockerbie (UK) TREA MVV Leuna GmbH

> 50

O2MD 04

58

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

4.2.4

Flow measuring devices

4.2.4.1 Measuring principle

FMD 02

FMD 99

The measuring gas is measured in the exhaust gas flow by means of a dynamic pressure probe. Hereby the differential pressure is continuously registered. The differential pressure over the dynamic pressure probe is a degree for the velocity of the exhaust gas. The flow of the exhaust gas can be determined in relation to the cross-section of the exhaust gas channel at the measuring point. The micro controller integrated in the control unit produces a signal being in proportion to velocity respectively flow. This is provided as 4 … 20 mA-signal.

v = A⋅

2 ⋅ ∆p +D Zeta ⋅ ρ

v& = v ⋅ A quer

4th Edition, Conditions March 2010

v

Velocity of the exhaust gas in [m/s]

A, D

Calibration constants

∆p

Differential pressure over pressure probe in [Pa]

Zeta

Probe variable

ρ

Density of exhaust gas in [kg/m³]

v&

Flow of exhaust gas in [m³/h]

v

Velocity of exhaust gas in [m/s]

Aquer

Cross-section of the exhaust gas channel at the measuring point in [m²]

the

dynamic

59


Product catalog Dr. Födisch AG :: Specifications

4.2.4.2 Applications

FMD 02 [in-situ]

Continuous monitoring of the exhaust gas flow (or velocity) as well as temperature e.g. at emission measurements Use at dust contents ≤ 50 mg/m³, no corrosive gases

Velocity: 0 ... 30 m/s, Flow: 0 ... 1.000.000 m³/h, Temperature: 0 ... 300 °C, Differential pressure: 0 ... 10 mbar

FMD 99 [in-situ]

Continuous monitoring of the exhaust gas flow (or velocity) as well as temperature and pressure (optional) e.g. at emission measurements Use under rough conditions (high dust content > 50 mg/m³, corrosive gases, ...)

Velocity: 0 ... 30 m/s, Flow: 0 ... 1.000.000 m³/h, Temperature: 0 ... 300 °C, Differential pressure: 0 ... 10 mbar Exhaust gas pressure: (optional)

60

th

27 BImSchV

th

17 BImSchV

Measuring ranges

th

Applications

TI Air

Device

13 BImSchV

TÜV -Approval

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

Measuring arrangement

In-situ

Measuring arrangement

Extractive

Power supply

230/110 VAC, 50/60 Hz

Power supply

24 VDC

FMD 99

FMD 02

4.2.4.3 Flow measuring devices in comparison

Probe Control unit Media temperature ≤ 280 °C Media temperature > 280 °C Dust content in exhaust gas

≤ 50 mg/m³

Dust content in exhaust gas

> 50 mg/m³

Flow velocity

from 3 m/s

Analogue output

4 ... 20 mA

Digital output

Limit value 1

Digital output

Limit value 2 / Maintenance request

Digital output

Failure

Digital output

Maintenance

Digital output

Failure/Maintenance

Digital output

Measuring range

Approval

TI Air

Approval

13 BImSchV

Approval

17 BImSchV

Approval

27 BImSchV

th

th

th

4th Edition, Conditions March 2010

61


Product catalog Dr. Födisch AG :: Specifications

FLOW MEASURING DEVICE

4.2.4.4 Flow measuring device FMD 02

TÜV-approved according to TI AIr, 13th , 17th and 27th BImSchV Limit value

Limit value

Limit value On Off

2

1

1

On Off

2

5s

On Off

2

2s

2s

FiB [Tm³/h]

v [m/s]

30

1

1min

T: 125.6

dp: 3.9

T: 125.6

dp: 3.9

v: 12.9

T: 125.7

Compact device only consisting of a probe Combined device – simultaneous measurement of exhaust gas velocity or flow as well as exhaust gas temperature Simple assembly: Probe: 1"-sleeve (DIN 2986) optional flange DN 25 PN 6 Analogue outputs: 4 ... 20 mA Digital outputs (potential-free): Limit value 1 (adjustable) Limit value 2 (adjustable) / maintenance request Failure/maintenance Display in m³/h or m/s or as online line diagram

Application and function For the operation of a facility with streaming gases (e.g. hall exhaust air, exhaust gases etc.) the continuous registration of the exhaust gas velocity respectively the flow as well as the temperature are often of substantial importance. In case of continuous emission measurements the mass of pollutants has to be disclosed additionally (mass flow [kg/h]). The flow measuring device FMD 02 is a measuring device for the continuous registration of gas velocity and temperature of gas flows in pipelines. Moreover it is possible to display respectively provide the flow in operational or norm state. The use of the back-pressure and Pt100-measuring principle guarantees a device simply to install and handle as well as a timely monitoring of the measuring parameters. Technical data Control unit:

Electronic and display in the probe head (no separate control unit required), IP 65

Probe:

1 dynamic pressure probe with integrated thermometer (500 mm) 160 x 130 x 550 mm (W x H x D), weight 2,1 kg

Measuring principle:

Differential pressure (dynamic pressure), Pt100

Measuring ranges:

Differential pressure: 0 ... 10 mbar, velocity: 0 ... 30 m/s, Flow: 0 ... 1.000.000 m³/h, Temperature: 0 ... 300 °C

Display:

Graphic display with online line diagram

Media temperature:

max. 280 °C

Ambient temperature:

-20 ... +50 °C

Dew point difference:

min. +5 K

Flow velocity:

From appr. 3 m/s

Analogue signals:

2 x 4 ... 20 mA

Digital signals:

3 potential-free contacts (failure/maint, limit value 1 and 2/maint. request)

Power supply:

110 VAC, 230 VAC or 24 VDC

TÜV-Approval:

TI Air, 13th, 17th and 27th BImSchV

62

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

FLOW MEASURING DEVICE

4.2.4.5 Flow measuring device FMD 99

Limit value

Limit value

Limit value On Off

2

1

1

On Off

2

5s

On Off

2

2s

2s

FiB [Tm³/h]

v [m/s]

30

1

1min T: 125.6

dp: 3.9

T: 125.6

dp: 3.9

v: 12.9

T: 125.7

Compact device only consisting of a probe Combined device – simultaneous measurement of exhaust gas velocity or flow as well as exhaust gas temperature (optional absolute pressure) Simple assembly: Probe: special flange DN 80 PN 6, Di=100 mm Analogue outputs: 4 ... 20 mA Digital outputs (potential-free): Limit value 1 (adjustable) Limit value 2 (adjustable) / maintenance request Failure/maintenance Display in m³/h or m/s or as online line diagram

Application and function For the operation of a facility with streaming gases (e.g. hall exhaust air, exhaust gases etc.) the continuous registration of the exhaust gas velocity respectively the flow as well as the temperature are often of substantial importance. In case of continuous emission measurements the mass of pollutants has to be disclosed additionally (mass flow [kg/h]). The flow measuring device FMD 02 W is a measuring device for the continuous registration of gas velocity and temperature of gas flows in pipelines. Moreover it is possible to display respectively provide the flow in operational or norm state. The use of the back-pressure and Pt100-measuring principle guarantees a device simply to install and handle as well as a timely monitoring of the measuring parameters. Technical data Control unit:

Electronic and display in the probe head (no separate control unit required), IP 65

Probe:

1 dynamic pressure probe with integrated thermometer (500 mm) 420 x 650 x 1.200 mm (W x H x D), weight 10 kg

Measuring principle:

Differential pressure (dynamic pressure), Pt100

Measuring ranges:

Differential pressure: 0 ... 10 mbar, velocity: 0 ... 30 m/s, Flow: 0 ... 1.000.000 m³/h, Temperature: 0 ... 300 °C, Absolut pressure: 800 …1.200 mbar

Display:

Graphic display with online line diagram

Media temperature:

max. 280 °C

Ambient temperature:

-20 ... +50 °C

Dew point difference:

min. +5 K

Flow velocity:

From appr. 3 m/s

Analogue signals:

3 x 4 ... 20 mA

Digital signals:

3 potential-free contacts (failure/maint, limit value 1 and 2/maint. request)

Power supply:

110 VAC, 230 VAC or 24 VDC

TÜV-Approval:

none

4th Edition, Conditions March 2010

63


Product catalog Dr. Födisch AG :: Specifications

4.2.4.6 References Devices FMD 02 and FMD 99

64

References (Extract) Compac tec Danpower Deepak (Indien) FHW Lahnbergen KPO, Kasachstan Steag, Michelin Dorst Technologies Biokraftwerk Schkölen Yara Rostock Enro Biomassekraftwerk Ludwigsfelde Berliner Wasserbetriebe TREA Leuna Sita Herne Wintershall ABB Kraftwerke AG AFRISO-EURO-INDEX GmbH, Güglingen Bilfinger + Berger Bremen Babcock Borsig Power Environment GmbH, Sandreuth BBP AE Energietechnik AG, HKW Mainz DK Jethwa (Indien) DOMO Caproleuna GmbH, Leuna EGN Entsorgungsgesellschaft Niederrhein mbH, WSAA Neuss Ekosis Ltd., Ankara (Türkei) EURO Science Co. Ltd. (Korea) Heraeus Tenova AG HKW Sandreuth Industrial Services & Machinery Corporation (Indien) InfraServ, Wiesbaden ISET, Hanau KEU GmbH, Krefeld Krematorium Braubach Hanau Linz Lühr Filter GmbH, Stadthagen Mahler AGS GmbH, Böhringer-Ingelheim Mahler IGS GmbH, Merck-Gernsheim Magmalor GmbH, Colditz Mertech s.r.o., Prievidza (Slowakei) Michaelis GmbH & Co. KG, MBS Vogtland, Oelsnitz Nickelhütte, Aue OPAL Environmental Pty Ltd., Smeaton Grange (Australien) Stadtentwässerung, Stuttgart Sundwig, Andritz Group, Hemer Volkswagen AG Wacker Chemie GmbH, Nünchritz Zweckverband Abfallbehandlung Nuthe – Spree, Niederlehme

Number of installed devices > 150

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

4.2.5

Emission evaluation computer

DAS 05

EMI-RECHNER

4.2.5.1 Emission evaluation computer DAS 05 & UmweltOffice 2005 DAS 05 registers process signals and used them according to the guidelines about evaluation of continuous emission measurements Simple installation: 19“-rack up to 8 DAS 05 can be connected to a system with max. 128 measuring points 7 - 64 analogue inputs 0/4 ….20 mA 12 - 96 digital inputs 24VDC 4 - 32 analogue outputs 0/4 ….20 mA 7 - 96 digital outputs (contact) optional Ethernet (RJ45)

DAS 05 The emission of specific substances into the atmosphere - e.g. dust, SO2, CO – is limited by immission acts. The kind of measuring data registration, the data processing, the data transfer to environmental agencies and the kind of reporting are stipulated in the guidelines published by responsible authorities. Emission evaluation computer must be approved by TUV. DAS 05 with the PC-system UmweltOffice2005 is a complete network-compatible system for the continuous registration, storage and evaluation of emission data in all industrial branches according to TA Luft, 13th , 17th , 27th and 30th BImSchV. Moreover the system provides important information for an environmentally-friendly and economic operation of the production facilities. In order to reduce maintenance costs the system is configures for remote maintenance done by Dr. Födisch AG. UmweltOffice 2005 is the programme for administration, evaluation and remote transfer of emission data. The programmes correspond to the requirements of the valid 13th and 17th BImSchV as well as to the federal guidelines. UmweltOffice 2005 shows the user interface in a window of the Internet Explorer and requires the Internet Explorer 6+ or higher and for the graphics a Java Plug-in. All reports are made as PDF files and offered as download respectively are shown via Adobe Acrobat Reader on the monitor. For some reports the data pool can be exported as CSV-file for further use. Moreover the data rows can be presented via Excel on the monitor.

4th Edition, Conditions March 2010

65


Product catalog Dr. Fรถdisch AG :: Specifications

4.2.6

Accessories

GMD 06

FIF 7

PCS 03

KM 2

4.2.6.1 Applications Device

Applications

GMD 06 [extractive]

Manual gravimetric dust content determination according to VDI 2066, page 1, 2, 3, 7 in streaming gases

FIF 7

Cascade impactor with 7 separation stages for manual gravimetric dust content determination according to VDI 2066, page 5

PCS 03

Back-flushing of the dynamic pressure probes in PFM 97, PFM 97 W or FMD 99 for high exhaust gas dust contents

KM 2

Condensate monitoring in the gas conditioning unit of cold emission monitoring systems

66

4th Edition, Conditions March 2010


Product catalog Dr. Fรถdisch AG :: Specifications

4.2.6.2 Gravimetric dust measuring system GMD 06

ACCESSORIES

manual gravimetric dust content determination according to VDI 2066, page 1, 2, 3, 7 in streaming gases

[GMD 06 Measuring case]

The GMD 06 is a semi-automatic sampling unit for gravimetric dust measurements. These measurements are used for the determination of dust concentration in streaming gases/air and as reference measuring method for the calibration of continuous dust measuring devices (in Germany e.g. VDI 2066). The GMD 06 is able to record independently all parameters being necessary for the dust measurement (e.g. humidity of the measuring gas, velocity in the exhaust gas channel as well as temperature and pressure). Under observance of the exhaust gas parameters the GMD 06 sucks off automatically a partial gas flow from the stack in isocinetic way, whereby the position of the sampling system is manually fixed respectively adjusted according to the valid regulations. This partial gas flow is sucked via a filter in which the dust contained is precipitated. Before and after the measurement the filter is conditioned (equilibrated) and weighed. After the weighed filter mass has been entered manually, the GMD 06 calculates the dust content of the probe in operational and standard state. Components

Accessory case

Pump case (optional)

Probe case with probe for flow and dust measurement and probe for humidity measurement

4th Edition, Conditions March 2010

67


Product catalog Dr. FĂśdisch AG :: Specifications

In preparation of the measurement the measurement should be planned following the VDI 2448 page 1. Moreover it is necessary to pretreat the filter sleeve and the plan filter from the measurement equipment before beginning and to weigh them subsequently. The principal construction of the sampling is shown in the upper picture. The extractive sampling for collecting the particles should be done iso-cinetically. That means, the sampling of the partial flow has to be done with the same velocity as the one of the main gas flow in order to prevent segregation occurences at the sampling. Fields of application The manual dust measurement in streaming gases serves judging the dust content per volume unit. This measurement is generally known as gravimetry. On the one hand this measuring method serves the determination of the dust concentrations and on the other hand it is the reference measuring method for the calibration of continuously measuring dust concentration measuring devices. The most important bases for this measuring method are stipulated in the VDI guidelines 2066 page 1, 2, 3, 7 and DIN EN 13284-1. General technical data Case of measuring case

Portable device (control unit integrated)

Dimensions of measuring case

400 x 370 x 200 mm (Wx H x D)

Case of accessory case

Portable device

Dimensions of accessory case

400 x 370 x 200 mm (W x H x D)

Case of probe case

Portable device

Dimensions of probe case

1.570 x 120 x 230 mm (W x H x D)

Case of pump case

Portable device

Dimensions of pump case

320 x 190 x 220 mm (W x H x D)

Immersion depth humidity probe

max. 650 mm

Immersion depth dust probe

max. 1.350 mm

Media temperature

max. 400 °C

Ambient temperature

-20 ... +50 °C

Dew point difference

min. +5 K

Power supply

230 VAC / 50 Hz, 200 W

Data storage

125 measurements

68

4th Edition, Conditions March 2010


Product catalog Dr. Fรถdisch AG :: Specifications

ACCESSORIES

4.2.6.3 Cascade impactor FIF 7 Cascade impactor with 7 separation stages for manual gravimetric dust content determination according to VDI 2066, page 5 Easy, handy but robust design Low wall losses

For the dust measurement by means of cascade impactors the different inertia of particles is used. Thereby several combinations of nozzle plates and deflectors are switched successively. If the dust loaded gas flow meets the deflector the bigger (inerter) particles are separated, whereas the smaller onces follow the gas flow.

4th Edition, Conditions March 2010

69


Product catalog Dr. FĂśdisch AG :: Specifications The nozzle diameter reducing from stage to stage leads to an increase of the gas velocity and therefore to separation of the single corn fractions. Before the exit there is an absoulte filter which absorbs the remaining dust content of the gas. Similar to the gravimetric measurement the cascade impactor has to be pretreated before executing the measurement in order to get respective faultless measuring values (see also GMD 06). The absolute filter is dried, equilibrised (exsiccator) and weighed. In addition it must be secured that before the measurement all elements are cleaned. Then the net weight of the nozzle plates and deflectors belonging together of each stage should be determined and documented. The structure of the sampling can be basically compared with the gravimetric dust measurement. The cascade impactor FIF 7 is a device with 7 separation stages. In the single stages there are hole circles (nozzles) which have per stage the same diameter reducing towards the flow direction. By that the gas velocity increases from separation stage to separation stage in a defined way.

Figure 1: Cascade impactor FIF 7 A circular dust collecting room formed as blind hole on the deflector is assigned to each opening of the nozzle exit. There the particles are deposited according to their separation characteristics. The decisive advantage of the device construction is not only the low wall losses whereby the collection of dust separated in total is relieved. A use of filter or adhesive mediums is not necessary so that no material changes occur during the measurements. In addition the device features a simple and handy, but robust design. So it is especially suitable for the measurement directly in the exhaust gas stream (In-stackarrangement). The FIF 7 can be combined with common measuring technology for the gravimetric dust measurement. In order to meet the different measuring gas conditions, different materials e.g. stainless steel of titan are available. The suction rate of the FIF 7 is limited by a critical nozzle at max. 3 mÂł/h. The sampling times to be realised with the device orientate at conventional dust measuring devices. The evaluation of the measurements is done according to VDI 2066 page 5.

70

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

4.2.6.4 Probe cleaning system PCS 03

ACCESSORIES

Periodic back-purging of differential or dynamic pressure probes with compressed air Application in dust and flow measuring devices at high exhaust dust contents (> 100 mg/m³) Simple assembly: Operational device: Wall assembly Assembly in the weather protection case (in case of application with PFM 97 or PFM 97 W as well as FMD 99) Immediately ready for operation

Combination with PFM 97, PFM 97 W and FMD 99

Application and function The probe cleaning system PCS 03 is a system for periodic cleaning of differential and dynamic pressure probes with compressed air. For this the probe of the measuring device connected is linked to the transmitter via the integrated valve unit PCS 03. During the back-purging procedure the transmitter is separated from the probe by means of the valve unit and the probe is purged. The PCS 03 consists of a control unit with integrated valve unit and the display. The control unit is designed as wall assembly case. In the control unit the complete electric connection, the switching of the valve unit and the control of the valves in the measuring and purging mode are done. Technical data Feeding:

230 V, 50 Hz...60 Hz

Operational readiness:

immediately

Dimensions (W x H x D):

240 x 160 x 120 mm

Impulse time:

3 seconds (adjustable)

Weight:

ca. 5 kg

Degree of protection:

IP 65

Ambient temperature:

-20°C...+50°C

Break time:

1,0 bis 99,0 Stunden

Operational pressure:

max. 6 bar

Case material:

Polycarbonat glas-fibre reinforced

Hose connection transmitter:

pressure

Hose clip 6mm

Hose connections probe:

Hose clip 9 mm

Connection for compressed air:

Hose clip 13 mm

4th Edition, Conditions March 2010

71


Product catalog Dr. Födisch AG :: Specifications

4.2.6.5 Condensate detector KM 2

ACCESSORIES

continuous condensate monitoring in measuring gases for continuous emission measurment after gas cooler Simple assembly: Evaluation unit: wall assembly Sensor: 6 mm screwed fitting Simple use via potential-free contact Almost free of maintenance

Tasks The condensate detector KM 2 is a highly sensitive system for condensate monitoring in measuring gases of continuous emission monitoring systems. Thereby the exhaust gas is monitored for condensate after the measuring gas cooler. The condensate could damage the analyser being subsequent in the measuring gas way (e.g. defective measuring cuvettes, defective measuring gas pumps,...). In order to protect the high-quality measuring technology in these emission monitoring systems the measuring gas pump can be switched off by means of the condensate detector in case of arising condensate. For the a potential-free contact is switched in the condensate detector KM 2. The function of the condensate detector KM 2 is based on the measurement of the conductivity, that means this function is only given in case of conductive gases. The sensor is designed as plug-in sensor with an outer diameter of 6 mm and can be integrated in the gas ways of analysers by means of suitable T-pieces. Technical data Feeding:

230 VAC

Operational readiness:

after 1 to 2 minutes

Connection box:

114 x 85 x 55 mm (W x H x D), ca. 0,5 kg

Sensor:

6 x 25 mm (D x T)

Degree of protection:

IP 65

Ambient temperature:

-20°C...+50°C

72

4th Edition, Conditions March 2010


Product catalog Dr. Fรถdisch AG :: Specifications

Emission measuring systems

EMISSION MEASURING SYSTEMS

4.3

Stationary and mobile emission measuring systems:

The Dr. Fรถdisch Umweltmesstechnik AG plans, installs and services measuring systems for operational and authorityrequired monitoring of air pollutants. Based on the technical competence and the long-time experience of our employees we are able to offer advanced complete solutions for emission monitoring. Our company philosophy is shaped by the fact that we do not only install the measuring systems but also offer the customer an extensive consulting inlcuding authority engineering as well as a professional after sales service for the measuring equipment. The combination of a research oriented to praxis, a realisation proved for industries and a reliable knowledge of valid environmental legislation in our company allow us to offer each customer an optimum and cost-oriented solution for his respective measuring task.

Your advantage Planning, installation and service from one source Support at the realisation of authority requirements (incl. determination of measuring points) Delivery of reliably-working and low-maintenance systems Compact construction of multi-component measuring systems Limitation to necessary periphery by using complexe measuring systems (e.g. measuring flanges) Professional service independent from brand

Legal bases The concrete requirements on the emission measuring facilities stipulated by authorities are defined by the sub regulations of the Federal Immission Control Act. Various ordinances and administrative regulations contain rules for the separate kinds of plants: 13th BImSchV combustion plants 17th BImSchV plants of single and integrated incineration of wastes 27th BImSchV crematories 30th BImSchV plants for biological treatment of wastes TI Air other plants obliged to monitor

The adjustment of environmental legislations to the present industrial and ecological conditions as well as the implementation of valid EU-norms into German law calls for a steady further development of legal regulations. Our emission measuring systems meet anytime the latest state of law.

4th Edition, Conditions March 2010

73


Product catalog Dr. Födisch AG :: Specifications

4.3.1

References

TI Air: BASF, Schwarzheide Chodovska Plzen (Tschechien) Deutsche Steinkohle AG, Kokerei Prosper DVV Kraftwerk, Dessau ECO Stahl, Eisenhüttenstadt EKOSIS, Ankara (Türkei) Gemeinschaftsklärwerk, Bitterfeld/Wolfen Heraeus Quarzglas, Bitterfeld Mahler AGS, Stuttgart Pfleiderer AG, Delitzsch Rüdersdorfer Zement GmbH Shanxi Textiles (China) SKW Piesteritz GmbH Stadtwerke Frankfurt/O. Stadtwerke Leipzig

74

13th BImSchV: EVO AG, Offenbach EWAG Nürnberg – HKW Langwasser Ibemo GmbH (Kasakhstan) INVISTA Resins & Fibers GmbH, Werk Offenbach Kraftwerke Wolfen GmbH Kraftwerk Laem Chabang, (Thailand) Kronospan Sandebeck Mainova AG, Frankfurt (Main) OPEL AG, Rüsselsheim Omnical Kessel und Apparatebau GmbH, Dietzhölztal Stadtwerke: Cottbus Dessau Frankfurt (Oder) Gotha Leipzig Potsdam Strausberg Wolfen Standard Fasel - Lentjes B.V., Southhampton (UK) Steag Saar Energie AG, Energieversorgung Michelin Reifenwerke Stora Enso GmbH, Eilenburg Südzucker AG, Werk Offenau WACKER Chemie GmbH, Burghausen BMKW Hallein Ante Holz Weiss Umwelttechnik GmbH, BKS Schkölen Dalkia Leinefelde SEV Sömmerda Deutsche Steinkohle AG, Kokerei Prosper Daimler Düsseldorf Uni Marburg Wintershall Emlichheim Dunlop Hanau Probis Bomlitz Enro Ludwigsfelde DEK Kaffee Berlin BMHKW Alperstedt Arcelor Bremen KGU Aalen Technische Werke Kaiserslautern Stadtwerke Leipzig , Standort Piesteritz CopacTec Straubing Pfleiderer AG Werk Nidda Tegra Holz Domat Schweiz BMKW Enkenbach RAG Hanau Heizwerk van Roje Oberhonnefeld BMHKW Zeickhorn, Grub a. Forst

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

17th BImSchV: Zementwerk Dyckerhoff Göllheim Zementwerk Lafarge Wössingen Stadtwerke Giessen PED Energy GmbH Chemiepark Bitterfeld Wolfen (über Baumgarte) RDF Plant Solvaylaan Zandvoord Oostende Belgien (über Baumgarte) Berliner Wasserbetriebe Abfallverwertung Schwedt GmbH Biomasse HKW, Berlin Gropiusstadt Biomasse HKW, Dresden Biomasse HKW, Herbrechtingen Biomasse HKW, Mannheim Biomasse HKW, Monopoli (Italien) Biomasse HKW, Neubrücke Biomasse HKW, Recklinghausen Biomasse HKW, Wicker Biomasse KW, Bergkamen Biomasse KW, Delitzsch Biomasse KW, Emden Biomasse KW, Fechenheim Biomasse KW, Hallein (Österreich) DOW Chemicals, Werk Schkopau Gemeinschaftsklärwerk, Bitterfeld/Wolfen Hauptklärwerk, Mühlhausen MIBRAG: Kraftwerk Deuben Kraftwerk Mumsdorf Kraftwerk Wählitz MHKW, Frankfurt Nordweststadt MHKW, Hameln MHKW, Mainz MHKW, TREA Leuna Muldenhütten Recycling GmbH Müllverbrennung Hameln GmbH Munitionsverbrennung Chi-shan (Taiwan) NIER, Seoul (Korea) Pfleiderer AG, Gütersloh Stora Enso GmbH, Eilenburg Wacker Chemie GmbH, Nünchritz Zellstoffwerk, Stendal

27th BImSchV Crematories Altenburg Berlin- Baumschulenweg Berlin-Wedding Braubach Coburg Dessau Eisenach Eisleben Flamarium Saalkreis Flensburg Fürstenzell Gera Greifswald Görlitz Gröbers Halle Hanau Linz, Österreich Lüdenscheid Magdeburg Meißen Osnabrück Potsdam Reutlingen Rostock Südharz, Nohra Tornesch Verden Weimar Wien Wuppertal Ahrensburg Hannover Schönebeck Aurich

30th BImSchV EGN Entsorgungsgesellschaft Niederrhein mbH, WSAA Neuss MBS Niederlehme MBS Osnabrück MBS Rennerod MBS Wetterau MBS Vogtland, Oelsnitz MBS Wetzlar

4th Edition, Conditions March 2010

75


Product catalog Dr. Födisch AG :: Specifications

Maintenance and Service

MAINTENANCE AND SERVICE

4.4

Periodic maintenance, service of measuring plants Cycle (depending on plant): monthly, quarterly, half-yearly or annually 24 h – reaction time (or on-call duty by agreement) training of operator’s staff remote maintenance and remote diagnosis

In Germany the installation of continuous emission measuring plants is required by diverse environmental regulations as e.g. the TI Air, the 13th BImSchV and 17th BImSchV. With erection, set up and calibration of a measuring plant the corresponding authority requirements are fulfilled at first, however, for some plants the observance of given minimum availabilities is stipulated. Moreover the environmental authorities are interested that the measuring plants in continuous operation guarantee a monitoring of their pollutant emissions being as consistently as possible.

Requirement: According to the "Bundeseinheitliche Praxis bei der Überwachung der Emissionen" (RdSchr. d. BMU v. 13.06.2005; GMBl Nr. 38 of 24.06.2005, page 795ff) it is valid for all plants obliged to monitor that: "The availability of the measuring systems has to reach at least 95 %. The measuring systems have to fulfil an availability of 98% for the determination of the oxygen reference content. The availability for evaluation systems has to be at least 99% in the sense of Nr. 2.5.1.3." Measuring systems for the use in plants of 13th BImSchV have to fulfil additionally the availability pointed out in annex II of the 13th BImSchV in the version of 2004. There it is stated: "Every day, at which more than six half-hourly average values are invalid because of failure or maintenance of the continuous measuring system, will be declared as invalid. If more than 10 days per year are declared as invalid due to such situations, the responsible authority has to oblige the operator to start suitable measures in order to improve the reliability of the continuous monitoring system."

For measuring systems in plants according to 17th BImSchV it is valid according to Art. 11 paragraph 11 of the directive 2000/76/EG of the European Parliament and the Council about the incineration of waste (ABl. of EC of 28.12.2000 Nr. L 332, S. 91, corrected by ABl. EC of 31.05.2001 Nr. L 145, S. 52): " In order to determine the valid daily average value in maximum 5 half-hourly average values of a day are allowed not to be considered for the continuous measurements because of being out of order or maintenance of the system. In maximum 10 daily average values per year are allowed not to be considered because of being out of order or maintenance of the continuous measuring system ."

76

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications In correspondance to the respective customer situation the Dr. Födisch Umweltmesstechnik AG offers an extensive service for the measuring plants which can be built in a customised way from the following elements: periodic maintenance in the respectively necessary rhythm (e.g. monthly, quarterly, annual) execution of annual revision attendance of annual functional tests and/or calibrations by an approved measuring institute provision of test gases within the frame of the plant’s maintenance execution of the cross-sensitivity calibration for steam at multi-component analysers (e.g. MCA 04, CEMAS-FTIR, MCS 100 HW) rapid repair of defective devices provision of replacement device in case of longer lasting repairs 24 h- on call duty (also at week-ends and holidays), observance of short reaction times compilation of the annual emission declaration

4th Edition, Conditions March 2010

77


Product catalog Dr. Födisch AG :: Specifications

4.4.1

References

Allessa Chemie A.U.G. Neue Aktivkohle und Umweltschutz GmbH, Döberitz AVA Frankfurt Nordweststadt Bilfinger + Berger Bremen Biomasse HKW, Dresden Boehringer Ingelheim Degussa AG, Radebeul DOW BSL, Schkopau EBV, Vogelgesang Eisenmann Maschinenbau KG Energieversorgung Potsdam GmbH Gemeinschaftsklärwerk Wolfen-Bitterfeld GmbH Haindl Papier Schwedt GmbH Heizkraftwerk Cottbus Heizkraftwerk Demmin Heizkraftwerk Dresden Heizkraftwerk Gera Heizkraftwerk Malchin Heizkraftwerk Mühlhausen Heizkraftwerk Neubrücke Industriekraftwerk Leuna Kraftwerk Bitterfeld/Wolfen GmbH Kraftwerk Buschhaus Kraftwerk Wilhelmshaven Krematorium: Bernburg Braubach Dessau Eisleben Eisenach Flensburg Gera Halle Potsdam LAUBAG Senftenberg, HW Jänschwalde Merck KG, Darmstadt MVV Mannheim Muldenhütten Recycling und Umwelttechnik GmbH Nickelhütte Aue Pfleiderer AG, Gütersloh Stadtwerke: Bitterfeld/Wolfen Dessau Frankfurt (Oder) Gera Gotha Leipzig Strausberg Wolfen Resch & Söhne, Schwerte SKW Stickstoffwerke, Piesteritz Stora Enso GmbH, Eilenburg Wacker Chemie GmbH, Nünchritz WESTAG + GETALIT, Rheda-Wiedenbrück Volkswagen AG, Baunatal

78

4th Edition, Conditions March 2010


Product catalog Dr. Fรถdisch AG :: Specifications

Engineering

ENGINEERING

4.5

Authority engineering, e.g.: BImSchG-procedures Environmental impact analysis Immission prognosis Emission declarations Process-related services for filter plants etc. Exchange of information with regard to changes in environmental legislation Execution of seminars

Environmental and Approval management Since 1991 the Dr. Fรถdisch Umweltmesstechnik AG has been working as neutral society in the field of approval management as well as executed numerous approval procedures in the most diverse areas and locations of industry. The services of approval management include: Clarification of authority requirements and the procedures Compilation of the approval applications according to present administrative regulations and federal or authority-specific requirements Compilation and coordination of necessary expertises (e.g.immission prognosis for air pollutants and odours noise reports, measurements on initial level pollution, consideration concerning the regulation on hazardous incidents, landscape conservation support plans) Clarification on necessity od environmental impact analysis according to EIA Law, compilation of general or site-related preliminary examination of the single case or of the environmental impact analysis Check of the congruence of all approval documents to be submitted Scheduling and attendence of authority appointments Preparation of and participation at scoping dates (at projects obliged to EIA) and discussions (in case of procedures with public participation) Work on objections Professional service beyond the approval certificate Due to the long-year experiences of the employees the following advantages result to the customer: Coordination of all project participants Timely strengthening and cost reduction of the procedure Relief of the applicant (no binding of internal resources, no expensive education of own employees) Transfer of present information with regard to amendments in legislation (European, state and federal state law) Procurement of eventually necessary expertises etc. beyond the approval procedure

4th Edition, Conditions March 2010

79


Product catalog Dr. Fรถdisch AG :: Specifications Our complete scope of performance includes:

Authority engineering BImSchG-procedures about erection and operation of plants subject to approval Applications according to ยง 4, ยง 16 BImSchG Announcement according to ยง 15 BImSchG Announcement according to ยง 5 Abs. 2 of 31th BImSchV Compilation of applications concerning legal matters on waste and veterinary Authority agreements Dispatch of environmental impact analysis, preliminary examination of single cases Compilation of waste balances (law on life-cycle resource management) Obtaining of permissions concerning legal matters on water Compilation of hazardous incidents reflections Execution of spread calculations according to TI Air Compilation of immission prognosis for air pollutants, odour and noise Calculations of stack heights Compilation of emission declarations

Process-related services Execution of filter analysis Operational calibrations of continuous dust measuring devices by means of gravimetric reference measurements Compilation of particle size analysis, particle size distributions, dust resistance diagrams Distribution of flow distribution measurements and aerosol measurements Attendance of plant optimisations in the field of power plants and at dryers

Other services Customer service with regard to legal amendments or changes, organisation of seminars Compilation of hazardous substance registers Judgement on odour situations Compilations of documents required by employment law (operation instructions, briefings, control permissions) Compilation of measurement technology concepts for emission measuring plants / Intermediation of installation expertises Provision of documents concerning power plant management Compilation of instruction documentions according to the work and environmental management manual within the frame of the Eco-Audit-Certification

80

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Specifications

4.5.1

References

References (selection): authority engineering, other services EVO AG, Offenbach

Genehmigungsanträge gemäß § 16 BImSchG veterinärrechtlicher Antrag zum Import von Tiermehl aus der EU und der Schweiz Betreuung der Inbetriebnahmephase zur Umfüllung und Verbrennung von Tiermehl Antrag gemäß § 49 Verwaltungsverfahrensgesetz Erarbeitung von Betriebsanweisungen Umsetzung der Öko-Audit-Verordnung Erstellung einer Studie über die Auswirkungen der Neuregelungen TA-Luft; 13.; 17. BImSchV

GKW Bitterfeld

Antrag gemäß § 49 Verwaltungsverfahrensgesetz zur Änderung der Nebenbestimmung bezüglich der kontinuierlichen HgMessung

Graphischer Großbetrieb Pößneck GmbH

Anzeige gemäß §5 Abs. 2 der 31. BImSchV Erstellung eines Gefahrstoffkatasters Emissionserklärungen

IVO - Power Engineering

BImSchG – Antrag Genehmigungsantrag gemäß § 16 BImSchG

Knauf Deutsche Gipswerke KG

BImSchG – Antrag Genehmigungsantrag gemäß § 16 BImSchG

Rhenus AG & Co. KG

Neugenehmigung entsprechend § 4 BImSchG Betreuung der Inbetriebnahmephase zur Umfüllung von Tiermehl

Riesaer Ölwerke GmbH & Co. KG

Genehmigungsantrag gemäß § 16 BImSchG

Stora Enso GmbH, Eilenburg

Umweltverträglichkeitsstudie Genehmigungsantrag gemäß § 16 BImSchG

References (selection): process-related services EVO AG, Offenbach

Emissionseinzelmessungen Minimierung des Glühverlustes Strömungstechnische Untersuchungen Verfahrenstechnische Untersuchungen im Rahmen der Versuche zur Mitverbrennung von Ersatzbrennstoffen

IBUTEC GmbH, Weimar

Säuretaupunkt-Bestimmung

Infraserv Frankfurt a.M.

Systemanalyse Emissionsmessung an der Reststoffverbrennungsanlage

Knauf Deutsche Gipswerke KG, Rottleberode

Strömungsverteilungs- und Feuchtemessungen zur Trockneroptimierung

MIBRAG GmbH & Co. KG Theissen

Staubverteilungsmessungen nach Dampfkessel vor Filter

Robert Bosch GmbH

Messung der Korngrößenverteilung im Abgasstrom

Scori Leuna

Einzelmessungen von organischen und kanzerogenen Stoffen

Stadtwerke Saarbrücken

Strömungsverteilungsmessungen

SVZ - Schwarze Pumpe

Ermittlung der Geruchsbelästigung durch eine Recyclinganlage

VKH Braunsbedra

Filteranalyse und Überprüfung der Signalwege

4th Edition, Conditions March 2010

81


Product catalog Dr. Födisch AG :: Specifications

Research/Development

RESEARCH/DEVELOPMENT

4.6

Development of triboelectric dust and dust concentration measuring devices (in-situ, extractive, for dry and wet gases, application behind hose and electro filters) Development of odour measuring devices for the recording of olfactory emissions and air monitoring Development of sensory measuring technology on customer's demand Test of own products and products of third-suppliers in own dust channel Calibrations of dust, dust concentration and flow measuring devices Development of special constructions for individual measuring tasks

Dust channel For being able to investigate the triboelectric measuring principle and its influencing factors more in detail it is necessary to simulate in a suitable test facility as many diverse operational states as possible for the single applications and device types. Ad hoc in 1998 a technical facility with a dust channel was erected at the Dr. Födisch Umweltmesstechnik AG. It offers the possibility to produce dust contents within the range of 1...10.000 mg/m³ and gas rates of 1...45 m/s. Apart from the purpose of research the dust channel allows at the same time a precalibration of single devices before delivery (dust and flow measuring devices). The dust channel contains four flow channels, where defined dust can be brought into in dispersed form. The system operates in suction mode. A flow rate of up to 22.000 m³/h is sucked alternating over the round flow channels with the diameters of 200 mm, 400 mm or 560 mm or over a rectangular flow channel with the dimensions 400 mm x 560 mm. The speed control of the fan is done by means of a frequency changer. For the production of defined dust concentrations the dosage of the solid dusts is done by means of a particle dosing and disperging device.

The dust-loaded gas flow is cleaned by a cartridge filter and blown back into the hall. In 2002 the facility was enlarged by an ESP in order to investigate ist effects and influences on the measuring behaviour on triboelectric dust measuring devices. All triboelectric dust concentration measuring devices run a calibration cycle under reall dust conditions and variable gas velocities before their delivery.

82

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Others

Others

5.1

Publications Over the last years several reports were published in the following magazines: Umwelttechnik Umweltmagazin Process Verfahrenstechnik ZKG International Umweltspezial Landtechnik Chemietechnik and the book Staubemissionsmesstechnik [ISBN 3-8169-2296-1]

PUBLICATIONS

5

Publications concerning dust measurement Topic

Author/s

Magazines

Basics of charge difference – Triboelectric dust measurement under theoretical and practical aspects

Dr. H. Födisch, D. Richter

Verfahrenstechnik

Influence of sensor soiling at triboelectric dust measuring devices

Dr. H. Födisch, P. Schengber

Wasser Luft Boden

Timely alternative – Triboelectric filter monitoring and dust concentration measurement

Dr. H. Födisch, G. Dietrich

Chemie Technik, ZKG International, Umweltmagazin, Process

Adhesive and wet – Extractive dust concentration measurement with triboelectric measuring devices

Dr. H. Födisch, P. Schengber

Verfahrenstechnik, Process, Immissionsschutz

Publications concerning gas analysis Topic

Autor/en

Zeitschrift

Emissions – measure safely and analyse exactly

Dr. H. Födisch, S. Schramm

Umweltmagazin

Availability first – continuous emission monitoring under harmonisation of European guidelines

Dr. H. Födisch, S. Schramm

P&A

Publications concerning environmental engineering Topic

Author/s

Magazine

Co-combustion of carcass meal

D. Filipczyk, B. Bergmann

VGB PowerTech

Topic

Author/s

Magazines

Monitoring and reduction of odours at open plants

Dr. H. Födisch, P. Schengber

Vortragsband Achema, Sensormagazin

Application examples for continuous odour emission measurements

Dr. H. Födisch, P. Schengber

VDI- Berichte

Publications concerning odour

4th Edition, Conditions March 2010

83


Product catalog Dr. Födisch AG :: Others

5.2

Terms of delivery

General conditions for the for the supply of products and services of the electrical and electronics industry I. GENERAL PROVISIONS 1. Legal relations between Supplier and Purchaser in connection with supplies and/or services of the Supplier (hereinafter referred to as “Supplies”) shall be solely governed by the present GL. The purchaser’s general terms and conditions shall apply only if expressly accepted by the supplier in writing. The scope of delivery shall be determined by the congruent mutual declarations. 2. The Supplier herewith reserves any industrial property rights and/or copyrights pertaining to its cost estimates, drawings and other documents (hereinafter referred to as “Documents”). The Documents shall not be made accessible to third parties without the Supplier’s prior consent and shall, upon request, be returned without undue delay to the Supplier if the contract is not awarded to the Supplier. Sentences 1 and 2 shall apply mutatis mutandis to the Purchaser’s Documents; these may, however, be made accessible to those third parties to whom the Supplier has rightfully subcontracted Supplies. 3. The Purchaser has the non-exclusive right to use standard software and firmware, provided that it remains unchanged, is used within the agreed performance parameters, an on the agreed equipment. Without express agreement the Purchaser may make one back-up copy of standard software. 4. Partial deliveries are allowed, unless they are unreasonable to accept for the Purchaser. 5. The term “claim of damages” used in the present GL also includes claims for indemnification for useless expenditure.

II. PRICES, TERMS OF PAYMENT, AND SET-OFF 1. Prices are ex works and excluding packaging; value added tax shall beaded at the then applicable rate. 2. If the Supplier is also responsible for assembly or erection and unless otherwise agreed, the Purchaser shall pay the agreed remuneration and any incidental costs required, e.g. for traveling and transport costs as well as allowances. 3. Payments shall be made free Supplier’s paying office. 4. The Purchaser may set off only those claims which are undisputed or non-appealable.

III. RETENTION OF TITLE 1. The items pertaining to the Supplies (“Retained Goods”) shall remain the Supplier’s property until each and every claim the Supplier has against the Purchaser on account of the business relationship has been fulfilled. If the combined value of the Supplier’s security interest exceeds the value of all secured claims by more than 10 %, the Supplier shall release a corresponding part of the security interest if so requested by the Purchaser; the Supplier shall be entitled to choose which security interest it wishes to release. 2. For the duration of the retention of title, the Purchaser may not pledge the Retained Goods or use them as security, and resale shall be possible only for resellers in the ordinary course of their business and only on condition that the reseller receives payment from its costumer or makes the transfer of property to the costumer dependent upon the costumer fulfilling its obligation to effect payment. 3. The Purchaser shall inform the Supplier forthwith of any seizure or other act of intervention by third parties. 4. Where the Purchaser fails to fulfil its duties, fails to make payment due, or otherwise violates its obligations the Supplier shall be entitled to rescind the contract and take back the Retained Goods in the case of continued failure following expiry of a reasonable remedy period set by the Supplier; the statutory provisions providing that a remedy period is not needed shall be unaffected. The Purchaser shall be obliged to return the Retained Goods. The fact that the Supplier takes back Retained Goods and/or exercises the retention of title, or has the Retained Goods seized, shall not be construed to constitute a rescission of the contract, unless the Supplier so expressly declares.

IV. TIME FOR SUPPLIES; DELAY 1. Times set for Supplies shall only be binding if all Documents to be furnished by the Purchaser, necessary permits and approvals, especially concerning plans, are received in time and if agreed terms of payment an other obligations of the Purchaser are fulfilled. If these conditions are not fulfilled in time, times set shall be extended reasonably; this shall not apply if the Supplier is responsible for the delay. 2. If non-observance of the times set is due to force majeure such mobilization, war, rebellion or similar events, e.g. strike or lockout, such time shall be extended accordingly. The same shall apply if the Supplier does not receive its own supplies in due time or in due form. 3. If the Supplier is responsible for the delay (hereinafter referred to as “Delay”) and the Purchaser has demonstrably suffered a loss therefrom, the Purchaser may claim a compensation as liquidated damages of 0,5 % for every completed week of Delay, but in no case more than a total of 5 % of the price of that part of the Supplies which due to the Delay could not be put to the intended use. 4. Purchaser’s claims for damages due to delayed Supplies as well as claims for damages in lieu of performance exceeding the limits specified in no. 3 above are excluded in all cases of delayed Supplies, even upon expiry of a time set to the Supplier to effect the Supplies. This shall not apply in cases of mandatory liability based on intent, gross negligence, or due to loss of life, bodily injury or damage to health. Rescission of the contract by the Purchaser based on statue is limited to cases where the Supplier is responsible for the delay. The above provisions do not imply a change in the burden of proof to the detriment of the Purchaser. 5. At the Supplier’s request, the Purchaser shall declare within a reasonable period of time whether it, due to the delayed Supplies, rescinds the contract or insists on the delivery of the Supplies. 6. If dispatch or delivery, due to Purchaser’s request, is delayed by more than one month after notification of the readiness for dispatch was given, the Purchaser may be charged, for every additional month commenced, storage costs of 0,5 % of the price of the items of the Supplies, but in no case more than a total of 5 %. The parties to the contract may prove that higher or, as the case may be, lower storage costs have been incurred.

V. PASSING OF RISK 1. Even where delivery has been agreed freight free, the risk shall pass to the Purchaser as follows: a) if the Supplier do not include assembly or erection, at the time when the Supplies are shipped or picked up by the carrier. Upon the Purchaser’s request, the Supplier shall insure the Supplies against the usual risks of transport at the Purchaser’s expense; b) if the Supplies include assembly or erection, at the day of taking over in the Purchaser’s own work or, if so agreed, after a fault-free trial run. 2. The Risk shall pass to the Purchaser if dispatch, delivery, the start or performance of assembly or erection, the taking over in the Purchaser’s own works, or the trial run is delayed for reasons for which the Purchaser is responsible or if the Purchaser has otherwise failed to accept the Supplies.

VI. ASSEMBLY AND ERECTION Unless otherwise agreed in written form, assembly and erection shall be subject to the following provisions: 1. The Purchaser shall provide at its own expense and in due time: a) all earth and construction work and other ancillary work outside the Supplier’s scope, including the necessary skilled an unskilled labor, construction materials and tools, b) the equipment and materials necessary for assembly and commissioning such as scaffolds, lifting equipment and other devices as well as fuels and lubricants, c) energy and water at the point of use including connections, heating and lighting, d) suitable dry and lockable rooms of sufficient size adjacent to the site for the storage of machine parts, apparatus, materials, tools, etc. and adequate working and recreation rooms for the erection personnel, including sanitary facilities as are appropriate in the specific circumstances; further more the Purchaser shall take all measures it would take for the protection of its own possessions to protect the possessions of the Supplier and of the erection personnel at the site, e) protective clothing and protective devices needed due to particular conditions prevailing on the specific site. 2. Before the erection work starts, the Purchaser shall unsolicitedly make available any information required concerning the location of concealed electric power, gas and water lines or of similar installation as well as the necessary structural data. 3. Prior to assembly or erection, the materials and equipment necessary for the work to start must be available on the site of assembly or erection and any preparatory work must have advanced to such a degree that assembly or erection can be started as agreed and carried out without interruption. Access roads and the site of assembly or erection must be level and clear. 4. If assembly, erection or commissioning is delayed due to circumstances for which the Supplier is not responsible, the Purchaser shall bear the reasonable costs incurred for idle times and any additional traveling expenditure of the Supplier or the erection personnel. 5. The Purchaser shall attest to the hours worked by the erection personnel towards the Supplier at weekly intervals and the Purchaser shall immediately confirm in written form if assembly, erection or commissioning has been completed. 6. If, after completion, the Supplier demands acceptance of the Supplier, the Purchaser shall comply therewith within a period of two weeks. In default thereof, acceptance is deemed to have taken place. Acceptance is also deemed to have been effected if the Supplies are put to use, after completion of an agreed test phase, is any.

VII. RECEIVING SUPPLIES

84

4th Edition, Conditions March 2010


Product catalog Dr. Födisch AG :: Others The Purchaser shall not refuse to receive supplies due to minor defects.

VIII. DEFECTS AS TO QUALITY The Supplier shall be liable for defects as to quality (“Sachmängel”, hereinafter referred to as “Defects”,) as follows: 1. Defective parts or defective services shall be, at the Supplier’s discretion, repaired, replaced or provided again free of charge, provided that the reason for the Defect had already existed at the time when the risk passed. 2. Claims for repair or replacement are subject to a statue of limitations of 12 months calculated from the start of the statutory statute of limitations; the same shall apply mutatis mutandis in the case of rescission and reduction. This shall not apply where longer periods are prescribed by law according to Sec. 438 para. 1 No. 2 (buildings and things used for a building), Sec. 479 para. 1 (right of recourse), and Sec. 634a para. 1 No. 2 (defects of a building) German Civil Code (“BGB”), in the case of intent, fraudulent concealment of the Defect or non-compliance with guaranteed characteristics (Beschaffenheitsgarantie). The legal provisions regarding suspension of the statue of limitations (“Ablaufhämmung”, “Hemmung”) and recommencement of limitation periods shall be unaffected. 3. Notifications of Defect by the Purchaser shall be given in written form without undue delay. 4. In the case of notification of a Defect, the Purchaser may withhold payments to an amount that is in a reasonable proportion to the Defect. The Purchaser, however, may withhold payments only if the subject-matter of the notification of the Defect involved is justified and incontestable. The Purchaser has no right to withhold payments to the extent that its claim of a Defect is time-barred. Unjustified notifications of Defect shall entitle the Supplier to demand reimbursement of its expenses by the Purchaser. 5. The Supplier shall be given the opportunity to repair or to replace the defective good (“Nacherfüllung”) within a reasonable period of time. 6. If repair or replacement is unsuccessful, the Purchaser is entitled to rescind the contract or reduce the remuneration; any claims for damages the Purchaser may have according to No. 10 shall be unaffected. 7. There shall no claims based on Defects in cases of insignificant deviations from the agreed quality, of only minor impairment of usability, of natural wear and tear, or damage arising after the passing of risk from faulty or negligent handling, excessive strain, unsuitable equipment, defective civil works, inappropriate foundation soil, or claims based on particular external influences not assumed under the contract, or from non-reproducible software errors. Claims based on defects attributable to improper modifications or repair work carried out by the Purchaser or third parties and the consequences thereof are likewise excluded. 8. The Purchaser shall have no claim with respect to expenses incurred in the course of supplementary performance, including costs of travel, transport, labor, and matertial, to extent that expenses are increased because the subject-matter of the Supplies has subsequently been brought to another location than the Purchaser’s branch office, unless doing so complies with the normal use of the Supplies. 9. The Purchaser’s right of recourse against the Supplier pursuant to Sec. 478 BGB is limited to cases where the Purchaser has not concluded an agreement with its costumers exceeding the scope of the statutory provisions governing claims based on Defects. Moreover, No. 8 above shall apply mutatis mutandis to the scope of the right of recourse the Purchaser has against the Supplier pursuant to Sec. 478 para. 2 BGB. 10. The Purchaser shall have no claim for damages based on Defects. This shall not apply to the extent that a Defect has been fraudulently concealed, the guaranteed characteristics are not complied with, in the case of loss of life, bodily injury or damage to health, restrictions to liberty and/or intentionally or grossly negligent breach of contract on the part of the Supplier. The above provisions do not imply a change in the burden of proof to the detriment of the Purchaser. Any other or addition claims of the Purchaser exceeding the claims provided for in this Article VIII, based on a Defect, are excluded.

IX. INDUSTRIAL PROPERTY RIGHTS AND COPYRIGHT; DEFECTS IN TITLE 1. Unless otherwise agreed, the Supplier shall provide the Supplies free from third parties’ industrial property rights and copyrights (hereinafter referred to as “IPR”) with respect to the country of the place of delivery only. If a third party asserts a justified claim against the Purchaser based on an infringement of an IPR by the Supplies made by the Supplier and used in conformity with the contract, the Supplier shall be liable to the Purchaser within the time period stipulated in Article VIII No. 2 as follows: a) The Supplier shall choose whether to acquire, at its own expense, the right to use the IPR with respect to the Supplies concerned or whether to modify the Supplies such that they no longer infringe the IPR or replace them. If this would be impossible for the Supplier under reasonable conditions, the Purchaser may rescind the contract or reduce the remuneration pursuant to the applicable statutory provisions. b) The Supplier’s liability to pay damages is governed by Article XI. c) The above obligations of the Supplier shall apply only if the Purchaser (i) immediately notifies the Supplier of any such claim asserted by the third party in written form, (ii)does not concede the existence of an infringement and (iii) leaves any protective measures and settlement negotiations to the Supplier’s discretion. If the Purchaser stops using the Supplies in Order to reduce the damage or for other good reason, it shall be obliged to point out to the third party that no acknowledgement of the alleged infringement may be inferred from the fact that the use has been discontinued. 2. Claims of the Purchaser shall be excluded if it is responsible for the infringement of an IPR. 3. Claims of the Purchaser are also excluded if the infringement of the IPR is caused by specifications made by the Purchaser, by a type of use not foreseeable by the Supplier or by the Supplies being modified by the Purchaser or being used together with products not provided by the Supplier. 4. In addition , with respect to claims by the Purchaser pursuant to No. 1 a) above, Article VIII Nos. 4, 5, and 9 shall apply mutatis mutandis in the event of an infringement of an IPR. 5. Where other effects in title occur, Article VIII shall apply mutatis mutandis. 6. Any other claims of the Purchaser against the Supplier or its agents or any such claims exceeding the claims provided for in this Article IX, based on a defect in title, are excluded.

X. IMPOSSIBILITY OF PERFORMANCE; ADAPTATION OF CONTRACT 1. To the extent that delivery is impossible, the Purchaser is entitled to claim damages, unless the Supplier is not responsible for the impossibility. The Purchaser’s claim for damages is, however, limited to an amount of 10 % of the value of the parts of the Supplies which, owning the impossibility, cannot be put to the intended use. This limitation shall not apply in the case of mandatory liability based on intent, gross negligence or loss of life, bodily injury or damage to health; this does not imply a change in the burden of proof to the detriment of the Purchaser. The Purchaser’s right to rescind the contract shall be unaffected. 2. Where unforeseeable events within the meaning of Article IV No. 2 substantially change the economic importance or the contents of the Supplies or considerably affect the Supplier’s business, the contract shall be adapted taking into account the principles of reasonableness and good faith. To the extent this is not justifiable for economic reasons, the Supplier shall have the right to rescind the contract, it shall notify the Purchaser thereof without undue delay after having realized the repercussions of the event; this shall also apply even where an extension of the delivery period has previously been agreed with the Purchaser.

XI. OTHER CLAIMS FOR DAMAGES; STATUE OF LIMITATIONS 1. The Purchaser has no claim for damages based on whatever legal reason, including infringement of duties arising in connection with the contract or tort. 2. The above shall not apply in the case of mandatory liability, e.g. under the German Product Liability Act (“Produkthaftungsgesetz”), in the case of intent, gross negligence , loss of life, bodily injury or damage to health, or breach of a condition which goes to the root of the contract (“wesentliche Vertragspflichten”). However, claims for damages arising from a breach of a condition which goes to the root of the contract shall be limited to the foreseeable damage which is intrinsic to the contract, unless caused by intent or gross negligence or based on liability for loss of life, bodily injury or damage to health. The above provision does not imply a change in the burden of proof to the detriment of the Purchaser. 3. To the extent that the Purchaser has a claim for damages, it shall be time-barred upon expiration of the statue of limitations pursuant to Article VIII No. 2. The same shall apply to the Purchaser’s claim in connection with actions undertaken to avoid any damage (e.g. callback). In the case of claims for damages under the German Product Liability Act, the statutory statue of limitations shall apply.

XII. VENUE AND APPLICABLE LAW 1. If the Purchaser is a businessman, sole venue for all disputes arising directly out of the contract shall be the Supplier’s place of business. However, the Supplier may also bring an action at the Purchaser’s place of business. 2. Legal relations existing in connection with this contract shall be governed by German substantive law, to the exclusion United Nations Convention on contracts for the International Sale of Goods (CISG).

XIII. SERVERABILITY CLAUSE The legal invalidity of one ore more previsions of this Agreement in no ways affects the validity of the remaining provisions. This shall not apply if it would be unreasonable for one of the parties to be obligated to continue the contract.

4th Edition, Conditions March 2010

85


Product catalog Dr. Födisch AG :: Others

5.3

How to find us ? Highway (from direction Nürnberg):

HOW TO FIND US ?

from BAB 9 to exit Kreuz Rippachtal, BAB 38 direction Leipzig to exit Leipzig Südwest, then B 186 direction Markranstädt

Highway (from direction Berlin): BAB 9 to exit Leipzig West, then B 181 direction Leipzig, in Dölzig at the traffic light turn right to B 186 direction Markranstädt

Plane: To airport Leipzig/Halle, then by taxi or rental car (see highway)

Train: By train to Leipzig main station, then by train to Markranstädt or by S-Bahn to final destination Miltitzer Allee

Regional: coming from the centre of Leipzig take B 87 in direction Weißenfels to Markranstädt. At the traffic light turn left and go on B 186 in direction Zwenkau, after approximately 1,5 km there is the company's entry on left side (after old power station)

BERLIN

A9

A14

B2

Schkeuditzer Kreuz

HALLE

A14

DRESDEN

Flughafen Leipzig/Halle

Saale Park

Leipziger Messe

Abfahrt Leipzig-West

B181

Merseburg

Leipzig

Leipzig

B87 B186 Markranstädt Dr. Födisch Umweltmeßtechnik AG MÜNCHEN

Abfahrt Bad Dürrenberg

A38

B87 Abfahrt Leipzig-Südwest

B2/95

Abfahrt Lützen

Kreuz Rippachtal

A38

86

A9

B186

B2

Zwenkau

4th Edition, Conditions March 2010


Issuu converts static files into: digital portfolios, online yearbooks, online catalogs, digital photo albums and more. Sign up and create your flipbook.