How to use Nanoliter Injector?

Page 1

WORLD PRECISION INSTRUMENTS

Instrumenting scientific ideas

QUICK START GUIDE

NANOLITER2010

Nanoliter Injector with SMARTouch™ Controller

Serial No._____________________

www.wpiinc.com 061217


Copyright Š 2017 by World Precision Instruments. All rights reserved. No part of this publication may be reproduced or translated into any language, in any form, without prior written permission of World Precision Instruments, Inc.


Nanoliter 2010 with SMARTouch™ Controller Get started with your Nanoliter 2010 and SMARTouch™ Controller as quickly as possible using these step by step instructions. NOTE: This Quick Start Guide assumes that the operator back filled a micropipette with mineral oil. For instructions on how to prepare the micropipette, see the Nanoliter2010 Instruction Manual page 7. 1. Connect the AC power cord to the power supply unit and connect the power supply unit to the MICRO2T controller. Plug the AC power cord into a wall outlet.

2

Connect the Nanoliter adapter cable to the output of channel 1 of the SMARTouch™ controller and the RJ-11 connector of the Nanoliter2010 to the RJ-11 female connector on the adapter. Make sure the embossed arrow on the connector is facing upwards for proper pin alignment.

Adapter Cable connects to the SMARTouch™ controller

World Precision Instruments

Nanoliter 2010 Injector connection to the adaper. 1


3. Power on the unit. The following screen appears.

4. Touch the Initial Splash Screen anywhere. The MAIN MENU appears. Touch DISPLAY ALL CHANNELS.

5. The syringe status screen appears. Press anywhere in the PUMP1 Display Area to make Pump1 the active pump.

2

World Precision Instruments


Nanoliter 2010 with SMARTouch™ Controller 6. Touch the Configure button.

7. Press the Type box to bring up the Syringe Selection Screen.

8. In the Syringe Selection Screen, press the DWN key until you see the NL (Nanoliter 2010) type and press it. It will become highlighted. Then press Back to return to the Pump Configuration Screen

World Precision Instruments

3


9. The CONFIGURATION screen appears. Touch the Volume Target display.

10. The Volume Target screen appears. Use the numerical keypad to enter 2500 nL. Touch ENTER to set the volume and return to the CONFIGURATION screen.

11. Touch the DELIVERY RATE display.

4

World Precision Instruments


Nanoliter 2010 with SMARTouch™ Controller 12. The Volume Rate screen appears. Use the numerical keypad to enter a desired injection rate in nL /sec. In this example, 500 nl/sec was entered. Touch ENTER to set the delivery rate and return to the CONFIGURATION screen.

13. The Configuration screen shows that a target volume of 2500 nL was entered and a delivery rate of 500 nL/s was entered. Touch BACK to return to the syringe status {Home} screen.

14. Verify that the arrow inside the syringe is yellow and pointing in the injection direction. If the arrow is not red and pointing to the left, touch DIRECTION to change it. Touch the RUN button to start the injection.

World Precision Instruments

5


15. The injection will run until 2500 nL is injected and will then halt automatically. Even though no injection actually happened, this will place the metal needle plunger about 13.6 mm from the home (fully retracted) position. At this point install the back filled micropipette on to the Nanoliter2010 injector as shown below.

16. Press the DIRECTION button to change the direction to Withdraw. The arrow turns green and points towards the back of the syringe depiction.

17. Press the CONFIGURE button.

6

World Precision Instruments


Nanoliter 2010 with SMARTouch™ Controller 18. Press the Volume Target area to bring up the Volume Target screen.

19. Use the keypad to enter the desired aliquot volume to be pulled into the micropipette. We will enter 200 nL and press Enter.

20. Press the Delivery Rate area.

World Precision Instruments

7


21. Use the keypad to set the rate to 50 nL/sec and press Enter.

22. Press Back to go to the main Syringe information display.

23. Next place the micropipette tip into the aliquot to be loaded and press RUN. This will pull 200 nL into the micropipette.

8

World Precision Instruments


Nanoliter 2010 with SMARTouch™ Controller 24. When the injector finishes, press the direction button to set it to Infuse. The arrow will turn Red.

25. Now the micropipette can be placed at the desired injection location and after setting the desired injection volume, the injection can be done by pressing the Run button.

World Precision Instruments

9


USA International Trade Center, 175 Sarasota Center Blvd., Sarasota FL 34240-9258 Tel: (941) 371-1003 • Fax: (941) 377-5428 • E-mail: sales@wpiinc.com UK 1 Hunting Gate, Hitchin, Hertfordshire SG4 0TJ Tel: 44 (0)1462 424700 • Fax: 44 (0)1462 424701 • E-mail: wpiuk@wpi-europe.com Germany Zossener Str. 55, 10961 Berlin Tel: 030-6188845 • Fax: 030-6188670 • E-mail: wpide@wpi-europe.com China & Hong Kong WPI Shanghai Trading Co., Ltd. Rm 25e, No8 Dongfang Rd., Pudong District, Shanghai, 200120 PR China Tel: +86 21 6888 5517 • E-mail:chinasales@china.wpiinc.com Brazil Av. Conselheiro Nébias, 756 sala 2611, Santos-CEP: 11045-002, São Paulo Brazil • Tel: (013) 406-29703 • E-mail: info@brazil.wpiinc.com Internet www.wpiinc.com • www.wpi-europe.com • www.wpiinc.cn