/trz03te_manual_en

Page 41

OPERATION

...............................................................................................................................................................................................................

Current output mode Here you define the measured or default value which is to be outputted. 0: Default value (see above) 1: Flow rate Qm All other values are not permitted! 37

Current damping Here you define how the current output is to be damped. Permissible values for D are 0 to 0.9 ! Output current = (1-D) x value + D x old value

Correction factors for current offset and current rise The current output is adjusted in the factory through two correction factors (offset and rise). In the event that a recalibration should really be necessary, there is a PC-based program available which will define the new offset and rise values for you.

...............................................................................................................................................................................................................