Drennen IFPAC Cal Transfer

Page 1

Calibration Transfer: A Pharmaceutical Case Study

James K. Drennen, III DCPT


Duquesne University Center for Pharmaceutical Technology (DCPT)

http:// http://www.pharmacy.duq.edu/DCPT/home.html www.pharmacy.duq.edu/DCPT/home.html

2


DCPT Activities • • • • •

PAT Method Development PAT Validation Pharm. Formulation and Development Biotech. Dosage Form Development Sensor Technology – NIR, Chemical Imaging, Acoustic Spectroscopy

3


Outline • Introduction and Methodology – – – – –

Calibration transfer vs. calibration update Instrumental or product drift Failure detection Investigation/Remediation activities Appropriate transfer standards

• Pharmaceutical Analysis Case Study – Failure Detection – Application of calibration transfer method – Results

4


Calibration Transfer • In general, calibration transfer refers to the derivation of a model to relate two sets of measurements – Calibration transfer from lab. to line – Calibration transfer after instrument maintenance/repair

5


Introduction • Definition of “calibration transfer”? – Calibration transfer is an understatement of the process. In many situations, instrument matching or instrument update is sufficient, while in others, calibration transfer entails complicated models relating one measurement paradigm to another (different sampling geometry, instrument make/model, even transfer between methods)…

6


Continuum of Calibration Transfer Types Preprocessing Methods (scatter correction, derivatives, etc.)

7

Instrument Matching (univariate linear, etc.)

Orthogonal Projection Methods

Complex, black-box Method-Method: PDS, ANN


Calibration Update • Definition: The enhancement of an existing calibration model through the inclusion of additional calibration samples – Calibration update samples are more representative of the current test samples than the original calibration samples

• Calibration update performed following change in product (eg: new supplier) 8


Possible Outcomes of NIR Prediction of Assay Results

9

Valid NIR Result

Not-Valid NIR Result

In Spec

1

2

Out of Spec

NIR Test Result for Tablet:

Evaluation of NIR Test:

4

3

1. Accepted result - pass 2. Investigation required (NIR result not acceptable) 3. Investigation required (NIR result not acceptable) 4. Accepted result – fail or During initial period – Investigation recommended


Failure Detection NIR Instrument

Sample

Potential errors ( B) due to: -Raw material change -Process change

Potential errors ( A) due to: -New instrument -Changed instrument response

NIR Data

Pre treatment

Instrument Matching

Model

Instrument Standardization

Result (Prediction)

Calibration Transfer Processes

NIR Prediction

Historical data and action threshold

Is prediction valid? (Q res and T 2 )

Result requires investigation

Result is valid

10

Final Result

Prediction Validity

Instrument Evaluation and corrective action


Calibration Transfer Data from tablet test

Instrument Evaluation and corrective action

Review internal Performance test results External Instrument Performance test Rescan Tablet

Faulty Initial Scan?

No

Has instrument performance changed significantly

No

Type B problem

Collect data from appropriate samples

Perform calibration update procedure

Yes Yes

11

Verify sample positioning system operation

Type A problem

Resume NIR testing

Recall PDS samples

Validate changes

Collect PDS data

Perform PDS instrument update protocol

Validate changes

Resume NIR testing


•

Potential investigation actions 1. Review daily internal performance test and internal performance test history 2. Run an internal performance test 3. Rescan tablet 4. Review SPC of method assessment 5. Review SPC of tablet test results 6. Perform a parallel laboratory test on tablet

12

•

Potential remediation actions A. Instrument repair, standardization, and external performance test B. Calibration update C. Address as an out-of-specification (OOS) investigation

Instrument evaluation and corrective action (legend)


Instrument evaluation and corrective action

Normal performance test results

Close to limit

Instrument performing correctly

1

Tablet positioning error

A

Normal performance test results Original spectrum not representative of sample

QC NIR measurement indicates in-spec tablet

4-6

QC NIR prediction

OOS tablet

13

6

OOS tablet

In-spec tablet

In-spec tablet

B

• Investigation tests/actions: 1-6 • Remediation actions: A-C • See Exhibit 5 for legend

QC NIR measurement indicates OOS tablet

3 In-spec tablet

Spectrum representative of sample

2

Fail

C

Document and continue

C


A

B Instrument evaluation and corrective action (continued)

Type A problem

Type B problem

Recall standardization samples Collect data from appropriate samples Collect standardization data Perform calibration update procedure Perform instrument standardization protocol

Validate changes

Validate changes

Resume NIR testing 14


Pharmaceutical Case Study: Online Tablet Analysis • Why are transfer/update protocols necessary? – Need for a calibrated backup instrument – Eventual expansion to further lines – Transfer-in-time of knowledge from earlier experiments

15


Example: Master and slave instruments 1.4 1.3

Reflectance Ratio

1.2 1.1 1 0.9 0.8 0.7 0.6 1300

1400

1500

1600

1700

Wavelength ( nm )

16

1800

1900

2000


Comparison of new and reference instruments Reference Inst. Absorbance @ 1440 nm

1.05 1 0.95 0.9 0.85 0.8 0.75 0.7 0.65

0.6

0.65

0.7

0.75

0.8

0.85

0.9

0.95

New Instrument Absorbance @ 1440 nm

17

1

1.05


Calibration transfer model 0.6

0.5

Multiplicative Constant

Arbitrary Units

0.4

0.3

0.2

Additive Coefficients 0.1

0

-0.1 1300

18

1400

1500

1600 1700 Wavelength ( nm )

1800

1900

2000


Prediction Following InstrumentInstrument Transfer Predicted Content ( mg, bias corrected)

70 65

N = 78 RMSEP = 2.25 %RSD = 23.2 % r = 0.973

60

r2 = 0.946

55 50 45 40 35 30 30

35

40

45

50

55

Measured Content ( mg )

19

60

65

70


Example: Lamp change 1.4

1.3

Reflectance Ratio

1.2

1.1

1

0.9

0.8

0.7 1300

1400

1500

1600

1700

Wavelength ( nm )

20

1800

1900

2000


Calibration transfer model for correcting lamp change 0.05 0.04 0.03

Reflectance Ratio

0.02 0.01

Additive Calibration Transfer Coefficients

0 -0.01 -0.02 -0.03 -0.04 -0.05 1300

1400

1500

1600

1700

Wavelength ( nm )

21

1800

1900

2000


Justification for baseline subtraction method Following Lamp Change, Uncorrected ( mg )

50 49.5 49 48.5 48 47.5 47 46.5 46 46

22

46.5

47

47.5 48 48.5 Prior to Lamp Change, ( mg )

49

49.5

50


Following Lamp Change, Transfer Corrected ( mg )

Prediction quality with calibration transfer 50 49.5 49 48.5 48 47.5 47 46.5 46 46

46.5

47

47.5

48

48.5

Prior to Lamp Change, ( mg )

23

49

49.5

50


Relative Error (multiple)

How Many Transfer Samples?

1

5

10

15

20

Number of Transfer Samples ( n )

24

25

30


Stability of transfer standards 4 3.8

Transfer SEP ( mg )

3.6 3.4 3.2 3 2.8 2.6 2.4 2.2 2

0

5

10

15

20

Time ( days )

25

25

30

35


Summary • Definition of calibration transfer/calibration update procedure is critical component of method development efforts • Calibration transfer standards may include samples of product

26


Acknowledgements • Carl Anderson • Robert Cogdill • David Molseed

27

• AstraZeneca • Brimrose Corp.


Duquesne University Center for Pharmaceutical Technology (DCPT)

http:// http://www.pharmacy.duq.edu/DCPT/home.html www.pharmacy.duq.edu/DCPT/home.html

28


Issuu converts static files into: digital portfolios, online yearbooks, online catalogs, digital photo albums and more. Sign up and create your flipbook.