PROCEDURE PR27 - ISOPOTENTIAL pH VALUE SETTING PURPOSE/SCOPE 1 min.
This procedure describes how to configure the isopotential pH value state. This setting is available in the Advanced programming mode and applies to the Custom pH analyzer state. Parts
None.
Tools
None.
PROCEDURE The isopotential point is the potential of the sensor that is not affected by sample fluid temperature changes. □
1. Press the SELECT smart key to modify the ISO.PT state.
□
2. The transmitter displays the current value. a. Use the blinking digit.
smart key to increment the value of the
b. When the first digit value is correct, use the key to move to the next digit.
smart
c. Repeat Steps 2a and 2b for each digit. d. Press the ENTER smart key to accept the new value. □
WBPEEUI520002A1
3. Press the NEXT key to go on to the next configuration parameter.
PURPOSE/SCOPE PR27 - 1