PACMEDTM Server Plus – User Guide 4.1.1 Canister not Empty 1. Identify the reason why the drug is not dropping and resolve the problem; Note: In most of the cases turning the crown and/or shake the canister will resolve the problem.

2. Replace the canister to its position; 3. Close all sliders and doors and touch RETRY (Fig. 4.1-1) on the console. 4.1.2 Defect canister 1. Extract the canister from the machine; 2. Close all sliders and doors and touch DEACTIVATE (Fig. 4.1-3) on the machine console. 3. Confirm through the following prompt:

4. Go to section 4.4 for the remaining of the process. Note: Communicate with our calibration department to report the defect canister.

4.1.3 Empty Canister Drug Available 1. 2. 3. 4.

Extract the canister from the machine; Touch REFILL (Fig. 4.1-2) on the machine console; Touch OK to confirm the refill in the prompt that follows; Go to section 4.2 or 4.3 for the remaining of the process

Drug Out of Stock 1. Extract the canister from the machine; 2. Close all sliders and doors and touch DEACTIVATE (Fig. 4.1-3) on the machine console; 3. Confirm through the following prompt:

4. Go to section 4.4 for the remaining of the process.

Customer Service 1.888.606.6337