Ultraview Care Network
Alarm Tone Manager (available on Ultraview UCW and Ultraview 1700 only)
!
•
The Alarm Tone Manager must be set by a system administrator with a Biomed level of privileged access.
The Alarm Tone Manager provides two levels of managing alarm notifications. • •
If this feature is ENABLED, it will not allow individual alarm tones to be permanently disabled. The system administrator can set a minimum volume for alarm tones ensuring that they are always at a safe and audible level.
!
•
The factory default for the Alarm Tone Manager is OFF, which inhibits access to the ALARM TONES keys in the Monitor Setup menu.
•
Contact your system administrator.
Setting Tone Volumes
MINIMUM VOLUME ON OFF
Figure 2-1: MINIMUM VOLUME key (Bedside monitor) To set minimum tone volumes when the Alarm Tone Manager is active: 1.
Set the desired minimum volume on each of the Alarm menus.
2.
Access the BIOMED menu and touch the MORE key until the MINIMUM VOLUME key displays.
3.
Touch MINIMUM VOLUME ON.
Central Alarm Watch Manager (available on Ultraview UCW and Ultraview 1700 only) The Central Alarm Watch Manager automatically activates the Alarm Watch feature for all currently monitored beds.
!
2-10
•
The factory default is Auto Alarm Watch ON and access to the Auto Alarm Watch feature DISABLED.
•
The Central Alarm Watch Manager must be set by a system administrator with a Biomed level of privileged access.
•
Contact your system administrator.