ENERLUX MP SOFTWARE MODULE USERS’ MANUAL Rev.1

AEM branch of LLC Group 26 Calea Buziasului

Phone: +40-256- 22 22 00 Fax: +40-256-49 09-28

1900 Timisoara

1


CONTENTS GENERAL FOLDERS AND FILES METER PROGRAM Operating program creation, editing, and writing Program preview property page TOU program property page Parameters property page METERS ACCESS Reading dialog behavior Meters reading Meters programming Meter internal clock programming Operating program writing Quick parameters programming Quick clock programming Special commands One step programming Calibrating mode Change of the meter-operating mode Testing the meter basic functions Direct access to the meter internal memories Meter configuration Meter calibration

2


GENERAL EnerluxMP.mpl module is a software library that functions together with Opticom 2k program and enables Enerlux MP single phase meters reading, programming, and calibration. The description is made for the maximum number of the functions that the software can perform. Restrictions are possible regarding the rights of the current user as they are defined in Opticom2K program or in relation with software variant. FOLDERS AND FILES EnerluxMP.mpl library has certain associated types of files, as follows: Default location Extension Aim File type Keeping the current configuration <Opticom 2k installation Configuestablished by the directory >\ EnerluxMP cfg ration user (dlgs.cfg) and Files\ the meter variants (enerluxmp.cfg) Program files <Opticom 2k installation created by the user Program directory>\ EnerluxMP pexmp or read from the Files\ Program\ meter Data files that result <Opticom 2k installation following the meter Data directory>\ EnerluxMP dexmp reading according to Files\Data\ IEC 1107 Files that result following the meter reading, mainly <Opticom 2k installation comprising Status directory>\ EnerluxMP sexmp information Files\Status\ concerning the commands sent to the meter Files that result following the meter <Opticom 2K installation reading comprising Report directory>\ EnerluxMP rexmp cumulated Files\Report\ information of the data, program and status file types

Observations

Encrypted. Can be read and edited with Opticom2K Encrypted. Can be read with Opticom2K

Encrypted. Can be read with Opticom2K

Encrypted. Can be read with Opticom2K

"EnerluxMP Files" folder and the associated sub-folders are created (if they do not exist) every time when library module is called by Opticom2K program.

3


METER PROGRAM The meter operates under multi-tariff duty, enabling energy multi tariff metering in 4 tariff registers. Multi-tariff metering is governed by a time program having the following structure: Seasons: Up to 6 time periods (seasons) of the year characterized by two beginning dates, so that each period is appealed two times a year. Rules: Each season is associated 1 – 14 switching rules (valid within the respective season) characterized by the following: The week time zone for which the rule is valid. The hour and the minute when the tariff switching takes place. Number of the tariff register where the metering takes place. Other programmable parameters: Besides the tariff program, the meter operating program allows the following programmable parameters: The rule of changing for the daylight saving time hour characterized by the month when the change takes place (programmable between March to May for the summer and between September to November for the winter ) and the week of the respective month 1- 4 or the last week. The communication speed as defined in IEC1107 mode C. The number of decimal figures for currency. The program has a certain syntax related to the programming lines, as shown in the table below: Programmable Program line Significance Quantity N000(dd.mm-dd.mm-TTTT) N=season number 1-6 Season dd=day 01-28 mm=month 01-12 These two characterize the two beginning dates of the respective season TTTT the tariffs in which the reactive energy will be metered (4,3,2,1 activated if T=1, not activated if T=0 in the corresponding position)

4


Switching rule

Nnn0(0Z-hh:mm-T)

The rule of 7000(mm.0w-mm.0w) changing for DST

N=the number of the season to which the rule 1-6 is associated nn=the number of the rule within 01-14 season Z=the week time zone for which the rule is valid: Monday-Thursday=0; Monday-Sunday=1; Friday=2; Saturday and Sunday=3; hh:mm=the hour and the minute when the switching takes place (1 min resolution) T=the number of the TOU register where the switching 1-4 takes place mm=changing hour month 0w=changing hour week: 1-4 or 5 (the last week of the month)

Communication 7210(n) n=0,1,2,3,4,5; speed 300,600,1200,2400,4800,9600 Bd Decimal figures 7600(N) N=0,1,2,3 figures after decimal point Besides the program proper, the current date and hour can also be programmed in the meter. Other special commands are also available. The program files proper, edited or read from the meter have pceet15 extension. Their implicit location is <Folder Opticom2K>\ENERLUXMP Files\Program. Programs editing and writing operations are made by means of a properties sheet. Every time it is appealed to for the creation of a new file, its contents will be of a self-created program file type called “template.pexmp�. Since this file is self-creating during the program running, do not save your own programs in this file.

5


Operating program creation, editing, and writing

The operating program creation, editing and writing can be carried out by means of PROGRAM property sheet. The contents of the property sheet contain either template.pexmp file, or the program file loaded in the view zone of the main window of Opticom 2k program, if it exists. The property sheet contains three pages of properties: Program preview, TOU program, and Parameters.

6


Program preview

This page enables program files loading and saving, current loaded program view, the specification of target meter parameters (address and password) with reference to the current meters list, the specification of what is required to be written into the meter, as well as the communication process starting and control. This page is always present. The table shows the current program contents loaded for editing and/or writing. Description column shows the program line significance, Value column shows the program line according to syntax, and Explanation column contains a concise explanation of the program line. The non-defined program lines have NULL value. Load button enables the loading of a program file other than the current one by means of a standard box, Open. If the user has made modifications in the previously loaded file, a warning will inform about the loss of the modifications that have been carried out. Save button pressing saves the current file loaded by means of a standard box, Save As. Meter group contains the combo box referring to the list of current meters, and General option referring to a generic meter not included in the meters list. In case there is no current list, only General option is present. Description combo-box contains the description of all the meters from the meters list selected in Opticom 2k program. The target meter selection is made from this running list. If All meters check box is selected, the programming operation will be successively applied to all the meters on the list. If the current list is null, the check box is disabled. This communication option shall be used only when the meters are connected to a communication bus. Serial no and Password editing fields reflect the contents of the current list concerning the meter series (used as communication address) and password. For General option, these fields contain General series and NULL password, or the one established in the reading dialogs

7


behaviour. The alteration of these fields is operative only during the existence of PROGRAM property sheet, it not being contained in the meter list used by the program. Communication section allows the communication operation starting and control. The user may choose to enter TOU program, or Parameters either separately or together. Certain program lines enable their exclusive sending, according to Selection option. The option is expressed by marking the chosen option box. The communication operation is started by pressing Start button. The communication process can be watched in WR (no. of transmitted characters) and RD (no. of read characters). The process can be stopped by pressing Stop button. If an error happens to appear during the communication process, the process ends and the user is warned by error message. The properties page cannot be changed and PROGRAM property sheet cannot be closed during the process. Time program

This page enables editing the multi-tariff part of the program. Program table comprises 6 lines corresponding to the program 6 seasons (Season 01-06). The first column (Season) shows the definition of the season, while the following 14 columns (Rule 01-14) show the definitions of the rules attached to the corresponding season. The definition of the seasons and of the rules complies with their syntax, except the absence of the label. Season edit and Rule edit sections comprise the defining controls of the seasons and of the associated switching rules. The carried out modifications are reflected by Program table after pressing Update item and Delete item buttons, respectively. By pressing Delete item button, the current cell contents is deleted from Program table. Season edit section comprises the controls 1st Time and 2nd Time. The two beginning dates can be defined by means of the up-down arrows of the controls, separately, for the month and the day. The section controls are active only when a season cell is selected (Season column of Program table).

8


Rule edit section comprises the following controls: Week days combo box which allows fixing the time zone where the switching rule is valid. Rule type combo box allows fixing the switching type: Time or Power. When time switching selection is carried out, the following controls are activated: The Time control, which allows fixing the time moment when the tariff switching takes place by means of up-down arrows. TOU combo box, which allows fixing the tariff register where the switching will take place. These controls are active only when a rule box is selected (Rule columns) from Program table. The direct filling in of the program table boxes can also be used with the observance of the syntax rules, without specifying the label. When leaving the properties page, the contents of Program preview page will reflect the changes that have been carried out. If there is a data entering error, the user will be warned by message, and he will not be able to quit the page.

9


Parameters

Parameters property page comprises the following groups: Daylight saving time: Month and Week pairs of combo boxes enable fixing the month, and the week of the respective month when the change to the summer hour and the return to the winter hour takes place. Disable DST option results in ignoring the rule. Communication baud rate radio buttons enable the selection of the communication speed as defined in IEC 1107 mode C. The Decimal figures for currency can be selected from the up down controls. When quitting the properties page, the contents of Program preview is updated with the modified values. METERS ACCESS Communication with the meters The communication with the meters is based on IEC1107 module C protocol provisions. The type of transmission is asynchronous, half duplex, with the following format of the communication word: one start byte, 7 data bytes, even parity, one stop byte. Communication starts with a foreword, initiated by the computer which comprises the selection of the meter through its address, the meter type reading, and the negotiation of the baud speed at which the communication will take place, and the type of communication (data reading, program(s) reading). After this initial phase the reading or writing operation is carried out. The address of the meter is the specified one in the current list selected in Opticom 2k program, and it has to correspond to the meter series + the last two digits of the manufacturing year. The address is compulsory only for the meters connected to this communication bus. In case of meters read through the optical port, the address is not needed to be specified, the address used being General. On the other hand, the addresses 10


of Pxx cu xx=00…39 form are reserved to be used together with the parallel multiplexing interface. They shall not be used if this interface is not present at the meter parallel port that has been previously selected. The communication speed is proposed by the meter, and it is the speed previously preprogrammed in the meter. The program may ignore this proposal and forces another communication speed. Data reading requires only the knowledge of the communication protocol. An automatic safety algorithm ensured by the program shall be fulfilled in program reading mode. The programming mode requires the knowledge of the password previously entered into the meter. If no password has been previously programmed, the chosen password shall be void (equivalent with NULL password). Certain reading or programming functions are available only when the meter is in the CALIBRATING MODE. This mode operates only if a hardware modification is carried out in the meter after its unsealing. EnerluxMP.mpl module accepts only Enerlux MP type meters. If another type of device is accessed, even by observing IEC1107 requirements, error is signaled out. Reading dialogs behavior

The dialog box enables the configuration of the meter reading dialog boxes behaviour by selecting or not Starts automatically and Close when communication ends options. Default password editing field enables password configuration for “General” options, as well for the quick programming. Meters reading The meters reading functions refer to the following possibilities: The data reading as they are supplied by the meter. The reading of the meter work program as it has been previously programmed. The reading of the meter status data: clock, display mode, working mode, price/tariff. The reading of a complete report of the contents of the meter memory which sums up the values of all the previous reading types.

11


The result of the successful reading operations is the storage of the data read in the files with names of the following type: MeterSerialNo @yymmddhhmmss and extensions of dexmp, pexmp, sexmp types, and rexmp type, respectively (for program files @yymmddhhmmss does not appear). After each successful reading operation, the corresponding file will be automatically loaded in the view zone of Opticom 2K program. The reading of the meter program and the meter status is carried out by using the safety algorithm, while the data reading is possible only on the basis of the data reading procedure, according to IEC1107 mode C. All the reading operations are made by means of Read meter dialog box.

This dialog box is displayed after launching any of the reading command. Exit button pressing, closes the dialog box. Meter group contains the data of the current meters list, and also General option referring to a generic meter not contained in the meters list. In case there is no current list, only General option is present. Description combo box contains the descriptions of all the meters from the meters list selected in Opticom 2K program. The selection of the target meter is made from this running list. If All meters check box is selected, the programming operation will be successively applied to all the meters on the list. If the current list is null, the check box is disabled. This communication option shall be used only when the meters are connected to a communication bus. Serial no editing field reflects the contents of the current list according to the meter series (used as a communication address). For general option, this contains General. The alteration of this field is operative only during the presence of the dialog box, it not being comprised in the list of the meters used by the program. If Start automatically check box is marked, the communication process will start at the same time with the launching of the dialog box, and if Close when communication ends check box is marked, the dialog box closes automatically, in case there were no errors in the communication process. The status of these boxes is memorized in dlgs.cfg file. The user can modify the status of these boxes during the communication process. Communication section enables the communication operation start and control. The communication operation starts by pressing Start button. The communication process can be watched in WR (no. of transmitted characters) field and RD (no. of read characters) field. The process can be interrupted by pressing Stop button. If an error

12


appears during the communication process, the process ends and the user is warned by an error message. The dialog box cannot be closed during the communication process. Meters programming Meters programming implies knowing the password which has been previously written in the meter (if it exists). Meter programming changes meter operation. Meter programming includes the following: Meter internal clock programming Operating program writing Special commands Display mode programming Passwords programming Meter internal clock programming Meter internal clock programming is carried out by means of Program meter clock dialog box.

The dialog box can be closed by pressing Exit button. Meter group contains the data of the current meters list, as well as General option referring to a generic meter that is not comprised in the meters list. In case there is no such current list, only General option is present. Description combo box contains the description of all the meters in the meters list selected in Opticom 2K program. The target meter will be selected from this combo box. If All meters checkbox is selected, the programming operation will be successively applied to all the meters in the list. If the current list is null, the checkbox is disabled. This communication option shall be used only when the meters are connected to a communication bus. Serial no and Password editing fields contain the series (used as communication address) and the password of the meters in the current list. As far as General option is concerned, these fields contain General series and NULL password, or the password established in the reading dialogs behaviour. The alteration of these fields can be carried out only while the

13


dialog box is present, it not being reflected by the contents of the meters list used by the program. Date and time section comprises the following controls: A time control (date and time) which enables establishing the date and hour to be written in the meter. Date and hour can be established by means of up and down arrows. Synchro button: If this button is pressed after current time modification, time control values are updated. The combo box enables the selection for the following options: The meter manages DST The meter will be passed in summer hour regime The meter will be passed in winter hour regime Communication section enables communication operation start and control. The communication operation begins after pressing Start button. The communication process can be watched up in WR (no. of transmitted characters) and RD (no. of read characters) fields. The process stops if Stop button is pressed. If an error appears during the communication process, the latter ends and the user is warned by an error message. The dialog box cannot be closed during the communication process. Operating program Operating program writing can be carried out by means of Program property sheet, Program preview property page. Operating program writing means the tariff program writing in the meter, either cumulatively or selectively, as well as the other parameters and the display sequence writing in the meter. Quick programming (meter program, clock)

When operating quick programming command, the program file loaded in the view area of Opticom 2K program main window is transmitted to the meter. If quick clock command is operated, the computer clock current value is sent to the meter. The operation of these commands is followed by the display of a dialog box of similar function. The communication starts automatically, and the box closes automatically if the operation has been carried out correctly. The communication process can be viewed up in WR (no. of transmitted characters) and RD (no. of read characters) fields. The process can be stopped by pressing Stop button. In such a case, or when there appears an error, the process stops and it can be resumed by pressing Start button, or the box can be closed by pressing Exit button. The password that will be used in the communication process will be that one specified at the configuration of the communication dialog boxes. 14


Special commands

Special commands section contains the list of the special commands and enables the selection of the required control. If Exit button is pressed, the dialog box closes. Meter group contains data referring to the current meters, as well as General option referring to a generic meter not included in the meters list. In case there is no current list, only General option is present. Description combo box contains the description of all the meters in the meters list selected from Opticom 2K program. The selection of the target meter is made from this combo box. If All meters checkbox is selected, the programming operation will be successively applied to all the meters in the list. If the current list is null, the checkbox is disabled. This communication option shall be used only when the meters are connected to a communication bus. Serial no and Password editing fields contain the series (used as communication address) and the password of the meter in the current list. As far as General option is concerned, these fields contain General series and NULL password, or the password established in the reading dialogues behaviour. The alteration of these fields can be carried out only while the dialog box is present, it not being reflected by the contents of the meters list used by the program. Communication section enables the communication operation start and control. The communication operation begins after pressing Start button. The communication process can be watched up in WR (no. of the transmitted characters) and RD (no. of the read characters) fields. The process stops if Stop button is pressed. If an error appears during the communication process, the latter ends and the user is warned by an error message. The dialog box cannot be closed during the communication process. The right side of the dialog box

15


contains the controls referring to the command that will be sent to the meter and depends on the item chosen from Special controls list. The special commands are as follows: Display mode: it enables the selection of the display operating mode: normal (multiplexed) mode, or only the registered energy in Wh.

Password programming: enables the writing of the password in the meter. Password writing is confirmed. "General" or "NULL" void passwords delete the meter password.

16


One step meter programming The commands can be transmitted through One step programming dialog box.

If Exit button is pressed, the dialog box closes. Meter group contains data referring to the current meters, as well as General option referring to a generic meter not included in the meters list. In case there is no current list, only General option is present. Description combo box contains the description of all the meters in the meters list selected from Opticom 2K program. The selection of the target meter is made from this combo box. If All meters checkbox is selected, the programming operation will be successively applied to all the meters in the list. If the current list is null, the checkbox is disabled. This communication option shall be used only when the meters are connected to a communication bus. Serial no and Password editing fields contain the series (used as communication address) and the password of the meter in the current list. As far as General option is concerned, these fields contain General series and NULL password, or the password established in the reading dialogues behaviour. The alteration of these fields can be carried out only while the dialog box is present, it not being reflected by the contents of the meters list used by the program. Items to be programmed section comprises the following controls: Program file combo box enables program file selection. The attached option control validates the operation. Meter clock time control together with the attached option control enables the transmission of the computer current time to the meter. Display mode combo box enables the meter passage to the normal display mode The attached option control enables the validation of the operation. Communication section enables the communication operation starting and control. The communication operation starts by pressing Start button. The communication process can be watched in WR (number of transmitted characters) fields and RD

17


(number of read characters) fields. The process can be stopped by pressing Stop button. If an error appears during the communication process, the latter ends and the user is warned by an error message. The dialog box cannot be closed during the communication process. Calibration mode Calibration mode is a meter special operating mode. The meter does not execute the metering functions in this operating mode, but it enables additional commands through the optical port required by the meter testing, initialization, and calibration. The password checking and the safety algorithms checking are also not executed in this operating mode. The meter operating mode is switched by means of a command through the optical port, but in order to make the calibrating mode effective, an additional hardware operation is required in the meter electronic circuit, operation which cannot be carried out without breaking the meter seal. Changing the meter operating mode Testing the meter basic functions Directly accessing the meter internal memories Meter configuration Meter calibration

18


Changing the meter operating mode

If Exit button is pressed, the dialog box closes. Meter group contains data referring to the current meters, as well as the General option referring to a generic meter not included in the meters list. In case there is no current list, only the General option is present. The Description combo box contains the description of all the meters in the meters list selected from Opticom 2K program. The selection of the target meter is made from this combo box. If All meters checkbox is selected, the programming operation will be successively applied to all the meters in the list. If the current list is null, the checkbox is disabled. This communication option shall be used only when the meters are connected to a communication bus. Serial no and Password editing fields contain the series (used as communication address) and the password of the meter in the current list. As far as General option is concerned, these fields contain General series and NULL password, or the password established in the reading dialogues behaviour. The alteration of these fields can be carried out only while the dialog box is present, it not being reflected by the contents of the meters list used by the program. Working mode section comprises Calibration mode and Normal mode radio buttons enable establishing the meter operating mode following the control release. Even if the command for passing to the Calibration mode has been received by the meter, the meter enters in the Calibration mode only if the hardware modification has been previously executed inside the meter. The passage to the normal operating mode can be executed either with or without the hardware modification inside the meter. Communication section enables the communication operation release and control. The communication operation can be released by pressing Start button. The communication process can be watched in WR (number of transmitted characters) fields and RD (number of read characters) fields. The process can be stopped by pressing the Stop button. If an error appears during the communication process, the communication process ends and the user is warned by an error message. The dialog box cannot be closed during the communication process.

19


The control for the meter passage to the calibration mode shall be password checked, as any of the meter programming. Once the passage to the calibration mode has been carried out, no password or safety algorithm checking will be performed by the meter. Testing the meter basic functions

The testing of some basic functions of the meter internal circuits refers to: The reading of the data supplied by the meter internal measuring chipset The measuring chipset reset Direct command of the meter internal relay. These controls can be transmitted by means of Test Meter test dialog box. If the Exit button is pressed, the dialog box closes. SPE Commands button enables the display of the dialog box with the current generator. The Meter group contains data referring to current meters, as well as General option referring to a generic meter not included in the meters list. In case there is no current list, only the General option is present. The Description combo box contains the description of all the meters in the meters list selected from Opticom 2K program. The selection of the target meter is made from this combo box. As far as the measuring chipset reading command is concerned, the selection of All meters option is not possible. This communication option shall be used only when the meters are connected to a communication bus. Serial no and Password editing fields contain the series (used as communication address) and the password of the meter in the current list. As far as General option is concerned, these fields contain General series and NULL password, or the password established in the reading dialogues behaviour. The alteration of these fields can be carried out only while the dialog box is present, it doesn’t modify the contents of

20


the meter list used by the program. The Command combo box enables the selection of the command that will be applied to the meters. The following controls can be selected: • Chipset reading • Chipset reset • Relay: ON; • Relay OFF The Communication section enables the communication operation release and control. The communication operation can be started by pressing Start button. The communication process can be watched in WR (number of transmitted characters) fields and RD (number of read characters) fields. The process can be stopped by pressing Stop button. If an error appears during the communication process, the transmission ends and the user is warned by an error message. The dialog box cannot be closed during the communication process. The Results section comprises a “read only” editing control. This shows the result of the measuring chipset direct reading or the result of the command applied to the meter. The selection of the control result reading can be made from the Response selection combo list which has the same contents as the Description combo list.

The direct access to the meter internal memories • The meter has three internal non-volatile memories, namely: • The calibration memory which stores the meter calibration coefficients; • The clock memory that ensures the meter internal clock operation, and ensures an additional data storage space. • The data memory, which serves for the storage of the metering data, as well as of the meter programming parameters. The meters memories can be directly accessed through Direct memory access dialog box.

21


If Exit button is pressed, the dialog box closes. The Memory type section comprises a combo list which enables the selection of the memory type which will be the object of the communication process: calibration clock or data. The Meter group contains data referring to the current meters, as well as the General option referring to a generic meter not included in the meters list. In case there is no current list, only the General option is present. The Description combo box contains the description of all the meters in the meters list selected from the Opticom 2K program. The selection of the target meter is made from this combo box. If the All meters checkbox is selected, the programming operation will be successively applied to all the meters in the list. If the current list is null, the checkbox is disabled. In case memories reading control is selected, the selection of All meters option is not possible. This communication option shall be used only when the meters are connected to a communication bus. Serial no and Password editing fields contain the series (used as communication address) and the password of the meter in the current list. As far as General option is concerned, these fields contain General series and NULL password, or the password established in the reading dialogues behaviour. The alteration of these fields can be carried out only while the dialog box is present, it not being reflected by the contents of the meters list used by the program. Communication section enables the communication operation release and control. The communication operation can be released by pressing Start button. The communication process can be watched in WR (number of transmitted characters) and RD (number of read characters) fields. The process can be stopped by pressing Stop button. If an error appears during the communication process, the latter ends and the user is warned by an error message. The dialog box cannot be closed during the communication process. Adr Start and No Bytes editing fields enable establishing the starting location and the number of octets for which the reading/writing process is taking place. The program will automatically correct the mistakes referring to the specification of the address and/or 22


number of octets incompatible with the type of the chosen memory, as function of the selected memory type. The Reading and Writing radio buttons, enable the specification of the type of operation that will be carried out with the memory. In case of the writing operation, the Verify option can be selected and, after writing, a memory reading operation, the comparison of the read data with the written data, and error messages will be released, as applicable. Memory content section comprises a table which shows the hexadecimal memory map as it will be written in the meter or as it has been read in the meter. The table enables the contents direct editing, but it is compulsory to write only hexadecimal values and capital letters 00...FF in its cells. The table can be loaded with binary data comprised in bdt type files by means of Load button. The memory image can be saved in bdt type file by pressing the Save button. Bdt files and their location shall be specified by means of some standard boxes of Open and Save as types, respectively. After installing the program, create the following three files: calpr.bdt, clockpr.bdt, datpr.bdt, comprising the initialization memory images for the calibration, clock, and data memories. When changing the type of memory, the file corresponding to the selected memory type will be automatically loaded as memory map. The initialization file corresponding to the current selected type of memory can be loaded any time by pressing Load ini. files button. It is important not to delete or modify these files. The Use system clock for meter clock option overwrites the bytes image over the currently loaded memory map, in case the clock memory type is selected.

23


Meter configuration

Meter configuration refers to the writing of the data concerning the meter series and type related with the voltage, current and frequency option in the non-volatile memory. The meters configuration is carried out by means of the Meter configuration dialog box. The dialog box contains the ListContent table, which shows the “General� meter on its first line, and the contents of the list selected in Opticom 2K program on the other lines. In case no list has been selected, only the first line of the table is present. The user can directly select the New serial column with the series to be written in the meters. If the Serial no ++ button is pressed, the new serial will be automatically filled in on the basis of the first line series. The option selection (reflected by the Meter code, Numerator and Denominator fields) is made from the Meter Types combo box. The frequency option must be filled in the table as 50 or 60 Hz. The meter configuration from the first line of the table can be repeated by pressing the Same config button. The configuration can be achieved only by selecting Select Row or All meters. The following options can be selected: Send serial no and/or Send config. The writing process can start by pressing the Start button. The process stops by pressing Stop button. RD and WR meters enable the communication process supervision. If an error appears during the communication process, the process will be stopped and the user will be warned by an error message. The box can be left by pressing Exit button. The box cannot be left during the communication process. 24


Meter calibration The calibration of the meter can be carried out by writing some digital coefficients in its calibrating memory, coefficients according to which corrections are made to the computations additional to metering. There are calibration coefficients for the compensation of the following quantities: • The phase displacement angle between the voltage and the current • The gain for active power • The offset for active power • The gain for reactive power • The offset for reactive power • The gain for the measured voltage • The meter starting current These coefficients can be determined by comparing the target quantity that corresponds to the coefficient to be determined with the value of the same quantity supplied by a standard device. The new coefficient will be determined by applying some computation formulas to the above mentioned values and to the value of the coefficient that is already present in the meter memory.

The calibration of the meter is ensured by means of Manual calibration dialog box. Meter combo box comprises the description “General”, as well as all the descriptions present in the current list of the program (if any). Address field shows the corresponding address. Thus, there takes place the specification of the address of the meter through which the communication takes place. Probe combo box enables the specification of the corresponding calibration coefficient that is to be determined. The voltage and the current applied to the meter have to be correlated with the selected test. The tests that can be selected are as follows: Voltage-voltage gain coefficient determining Zero P-active power offset coefficient determining

25


Zero Q-reactive power offset coefficient determining P gain-active power gain coefficient determining Q gain-reactive power gain coefficient determining Phase -phase displacement coefficient determining. The coefficients calculation is carried out by using the values of an editing controls matrix. The lines of the matrix correspond to the measuring chipset three phases (Ph1, Ph2, Ph3). For single phase meters, only the first row of the matrix is used. The matrix columns are as follows: Meter: the quantity read from the measuring chipset (read only) Calibrator: the quantity supplied by the standard device (it has to be manually entered) Error: calculates the percentage error of the quantity read from the meter as compared to the standard value (read only) Old Coefficient: the value of the coefficient read from the meter and corresponding to the selected test (read only) New coefficient: the value of the new coefficient calculated according to the formula for the selected test. The calculated value can be ignored, and the required value can be manually entered. This value shall range within –128‌127 values. The editing fields matrix that corresponds to Start up, OFFMOD, MCC coefficients, having the values of Old coefficient (read from the meter, read only) and New coefficient (manually edited), enables the manual entering of the value of the coefficients corresponding to the starting current, the offset coefficients and MCC coefficient calculating way. The editing read only control on the right of the dialog box displays the values supplied by the measuring chipset during the acquisition process. Inst values, Average and Freeze radio buttons help choosing the display (and the taking into account) the average and instantaneous values of the quantities supplied by the chipset, the display blocking respectively. Correct calibration requires the use of average values. WR and RD fields show the communication flow by displaying the number of the transmitted and read characters. The buttons that enable starting the operations adherent to the communication and calibration process are as follows: Reset coefficients: writes in the meter all the calibration coefficients of zero value Read coefficients: reads the calibration coefficients values from the meter. The new coefficients will be equal with those written in the meter. Acquisition: starts a reading process of the coefficients in the meter followed by a cyclical acquisition of the values supplied by the measuring chipset. The calculation of the new coefficient dynamically derives in correlation with the specified test. Reset average: resumes from the very beginning the mediation process of the quantities supplied by the chipset. Write coefficients: writes the newly calculated coefficients in the meter. Stop: stops the current communication process. SPE commands: starts-up the dialog box for the command of the voltages and currents generator. It is active only if a communication port with the generator has been previously specified.

26


Store inductive PF– it is active when the phase displacement test is selected. It stores the inductive power factor value. The meter shall have a 60 degrees inductive phase displacement. Store capacitive PF – it is active when the phase displacement test is selected. It stores the capacitive power factor value. The meter shall have a 60 degrees capacitive phase displacement. PCN calculation: carries out a precise calculation of the phase displacement coefficient supposing that the values of the inductive and capacitive power factors have been previously stored. The dialog box quitting is made by pressing Exit button. The dialog box cannot be quit during the communication process.

27