DIB29 1217_ZPower Retrofit Kit Battery Door Removal

Page 1

EVOKE™ AND BEYOND™ FUSION2 RECHARGEABLE HEARING AIDS ZPower® Retrofit Kit Battery Door Removal 1. Place the hearing aid with the Widex logo facing down with the battery door completely closed. 2. Center the battery door hinge pin over the center hole of the gray stability piece. 3. Carefully insert the pin ejection tool STRAIGHT into the hinge pin to push it out into the gray stability piece. If you feel resistance, realign the removal tool to the hinge pin. 4. Remove the battery door by pulling it straight out from the hearing aid. 5. Insert the ZPower battery door with the battery door pin holes lined up to the hearing aid casing to ensure a clear pathway for the hinge pin. 6. Align the battery door hinge pin so that it is perpendicular to the hearing aid. 7. Use the flat side of the battery door pin ejection tool to carefully push the hinge pin back in place.

To see a full video of the battery door replacement process go to: https://usa.widex.pro/en-us/products/rechargeable-hearing-aids

www.widexPRO.com

DIB29/0219

8. Make sure that the hinge pin is completely flush with the hearing aid casing.