1 minute read

ALIGNMENT FIXTURES

Some applications require very precise specimen alignment. This is achieved using special fixtures designed to optimize specimen alignment.

Principle of Operation

A variety of fixtures can be used to control both concentricity and angularity. The AlignPro™ system consists of a fixture that is mounted between the load cell and the load frame that allows easy adjustment of both concentricity and angularity. The fixture has been designed for use with a strain-gauged specimen mounted in the grips. The output from the gauges indicates the degree of misalignment, and these can be input to AlignPro software that provides a numerical and pictorial view of the degree of bending in the specimen. Adjustments can be made in real time, and the effect on bending seen immediately. This greatly simplifies and speeds the process of achieving optimum alignment.

Application Range

• Static, fatigue and through-zero tests on materials where optimum alignment is critical

8 alignment cell, 8 gauges, 10 mm diameter - - -

8000-173 Strain-gauged alignment cell, 8 gauges, flat 12 × 6 mm - - -

8000-174 Strain-gauged alignment cell, 8 gauges, thin flat 15 × 2 mm - - -

8000-179 Strain-gauged alignment cell adaptor cable

Note:

1. For NVLAP certified strain-gauged alignment cells, please consult factory.

For use with 8000-170 AlignPro Software and electronics package. Comprises specimen with 8 mm diameter GL and 15 mm gripped diameter.

For use with 8000-170 AlignPro Software and electronics package. Comprises specimen with 10 mm dia in GL and ends compatible with 3117-031 high-temperature reverse stress pullrods.

For use with 8000-170 AlignPro Software and electronics package. Comprises flat specimen 5 × 3 mm in GL and 15 × 3 mm gripped tab dimensions.

For use with 8000-170 AlignPro Software and electronics package. Comprises thin flat specimen 15 × 1 mm

For use with older style strain-gauged alignment cells with current AlignPro system