1 minute read

TB3 Maintenance

The only routine maintenance required is cleaning. Surface tension changes can occur if proper routine maintenance is not conducted. Isopropyl Alcohol is recommended for surface cleaning.

• Catch Filter

Advertisement

• Syphon

• Interior of Bucket

• Top surface of adjusting screws

• Enclosure locking screws

• All insect screens

TB3 Calibration

• TB3 rain gauge is calibrated at the factory prior to shipment

• TB3 should hold its calibration for 100,000’s of events prior to a calibration being required

• Prior to conducting a calibration check, a thorough cleaning of the rain gauge is required

• most often a cleaning will bring the TB3 back into specified accuracy without having to do a calibration

Specified accuracy is +/- 2% for 0 to 100mm/hr.

This article is from: