Solution Manual For Basic & Applied Concepts of Blood Banking and Transfusion Practices, 6th Edition Paula R. Howard, MPH, MS, MLS(ASCP)SBB and Wyenona Hicks ______________________________________________________________________________ Chapters 1-16 EXERCISE 1.1 PEC is a successful and modern blood donor center located in the northeastern region of the United States. Its management style has always stood out among other blood centers in the United States because of its modern and progressive views. The blood center is known for embracing changes in manufacturing with a positive approach by appropriately planning and allocating resources to ensure things are done correctly. However, in the last few years, employees and customers have noticed a steady decline in control and quality. The number of manufacturing errors and product and service complaints reported by customers has drastically increased. PEC has tripled in size in the last 3 years after acquiring several small blood banks in the state. Changes have occurred quickly to accommodate this growth, and thorough planning before implementing these changes has not always been possible. Among its growing pains, PEC employees finally realized that their blood bank computer system was outdated and unable to handle the workload volume increase in the last few years. Management at PEC selected EJO, Inc., as the new software vendor. Although EJO’s experience, reputation, expertise, and client support were well known, its new blood banking software was not. However, the cost of implementing this new software fit well within the approved fiscal budget. With full management support, validation and training efforts began early in the year with a proposed “live” date for September of that year. By June, multiple “bugs” had been identified during system testing. EJO’s technical support was excellent, but management was getting anxious with so many problems to be addressed and corrected. They wanted the new system to go “live” before their next FDA inspection. September came and went, and the system was not ready for implementation. Finally, in March of the following year, parallel testing was completed. Because of time constraints, not all functions were tested, particularly those pertaining to the product issue. Management was aware but decided to go ahead with the implementation of the system. In April, the new system at PEC was implemented with no major problems. Everything was okay until 3 months later when the distribution supervisor noticed that the system had allowed the release of human T-cell lymphotropic virus (HTLV) I/II repeatedly reactive donor units. The PEC QA department was immediately notified, and recall procedures were initiated. By that time, 35 donor units and all their parts had been released. 1. What did PEC fail to do? What were the consequences?
a. PEC failed to thoroughly validate the new blood bank software. PEC’s reputation was damaged. b. The safety of the blood supply was compromised by allowing the release and issue of unacceptable products. c. Recalls and revalidation (rework) are very costly to facilities. PEC had to invest time and effort to fix a preventable event. 2. What are some of the factors that contributed to the release of unacceptable units? PEC underwent rapid changes as it grew. These changes were not controlled, and items started falling through the cracks. A change control system would prevent this event within an organization. To save money and speed things up, PEC’s management forgot to support complete and thorough validation of the blood bank software. 3. What steps could have been taken to prevent this situation? Detailed software validation test plans should be executed to confirm whether the system is able to prevent the release of unacceptable products. Every possible scenario should be tested. It would be logical to go back and review all validations performed for all functions to ensure they were complete and performed and interpreted correctly. Redoing and reviewing work that has already been done costs an organization money and resources. EXERCISE 1.2 AKC Blood Center collects more than 600 platelet apheresis products each month. Platelet counts for donors and these products are determined using KB1, a state-of-the-art automated hematology instrument. This piece of equipment, similar to many others, requires extensive QC and preventive maintenance by the manufacturer. During the last 4 days, Fred, the medical technologist assigned to run the instrument, has failed to notice that the low-level control has consistently fallen below the mean. Fred has been too busy trying to train new personnel in the department and has had no time to plot his results on the QC chart, a departmental procedure requirement. If he performed this task, he would have noticed the obvious shift. Today, the new trainee, Fran, runs platelet counts with Fred’s assistance. Once again, the low-level control falls below the mean. On this run, the control falls outside the manufacturer’s range altogether. Fran notices KB1 flags the value and notifies Fred of the problem. Fred is too busy on the phone handling an irate call from a client hospital and tells Fran to “keep running it until it falls in. You know, sometimes it takes a while for controls to come in.” 1. If you were Fred’s supervisor, what would you do? Fred has a very serious and complex problem. He does not understand the purpose of using controls in the laboratory. As an immediate corrective action, Fred should not operate the KB1.
He can perform other job duties until retraining is accomplished. This training, education, and counsel should follow as soon as possible. Fran should be assigned to another qualified trainer. 2. What is the root cause of the problem? Fred does not possess the knowledge and education necessary for this task. A thorough review of previous training records and competency assessments should be done to evaluate whether these have been appropriate. 3. What steps can be taken to prevent this situation? Training programs must be thorough. Employees should be thoroughly trained, and their competency evaluated initially and then periodically to assess whether they can perform their jobs with skill and consistency and in conformance with established standard operating procedures. 4. What is your opinion of Fred’s results since he started using the instrument? Fred’s handling of control results has compromised the accuracy and precision of his laboratory results. In other words, Fred’s results are invalid. EXERCISE 1.3 The transfusion services department at PSB Medical Center has always been happy with the level of service offered by YR Blood Center. However, during the last 3 months, they have noticed quite a few platelet units with visible clumping. If you worked for the QA department at YR Blood Center and were asked to investigate this problem, what areas would you investigate to identify the root cause of the problem?
Step 1. Identify some reasons for platelet clumps in a product. One or more of the following items can usually cause platelets to clump: a. Incorrect spin speed (RPM) and time of centrifugation b. Failure of personnel to follow standard operating procedures in the preparation and storage of blood components c. Lack of or incorrect calibration of centrifuge d. Lack of preventive maintenance Step 2. Investigate and review the YR Blood Center for the following items: a. Verify that current and past quality control and preventive maintenance records are complete and accurate. Look for trends or shifts on quality control charts. Do quality control records on file reflect that the centrifuge is working properly? b. Verify that training and competency assessment records on file and/or in progress are complete and satisfactory.
c. Review units that have clumped platelets. Are these units prepared by only one employee or several employees? In other words, is this incident isolated, or is there a pattern about a certain employee or employees? EXERCISE 1.4 Patient PC needed an irradiated red blood cell (RBC) unit transfusion. ABC Blood Center shipped the unit to the hospital. The unit was labeled as irradiated but did not complete the irradiation process. The RBC unit was issued to the nursing unit for transfusion to Patient PC. The unit was not transfused and was returned to the laboratory. The laboratory was notified of the labeling error by ABC Blood Center later that day. Is this a reportable deviation to the FDA? State your reasons for your decision. Yes, this incident is a reportable biologic product deviation to the FDA. After the release of the irradiated blood product, ABC Blood Center discovered that the irradiated product violated its specifications, standards, or cGMP regulations. The blood bank must report the problem to the FDA and the buyer. In this event, the QA department must issue a biologic product deviation (BPD). BPDs provide information on how a blood product’s safety, purity, or potency may be affected. EXERCISE 1.5 Identify the consequences of the following cGMP violations: 1. Not taking daily temperature readings for a blood refrigerator Red cell hemolysis may result if the temperature drops below 1° C. If the temperature exceeds 4° C, bacterial growth becomes a possibility. 2. Extending test incubation times beyond what the SOP states Depending on the nature of the test, the results could be false-positive or false-negative. The test is technically invalid because it was not performed as specified by the standard operating procedures. 3. Overloading a refrigerator beyond its capacity Equipment must follow the specifications of the manufacturer for use. Overloading the refrigerator can potentially impact the ability of the equipment to maintain required temperatures for the contents. 4. Cutting an employee’s training short The employee will not be able to perform as expected and as specified in the standard operating procedures. This ultimately compromises the employee’s competency and the service or product quality.
5. Running proficiency samples repeatedly until the samples are depleted Proficiency samples must be handled as any other samples tested in the laboratory. They are treated in compliance with established procedures. 6. Using expired reagents Depending on the nature of the test, expired reagents could cause false-positive or false-negative test results. 7. Reporting test results when controls are out of range Quality control ensures the accuracy and precision of test results. In other words, controls indicate whether the test system can be trusted. If controls are out, the test should be considered invalid. 8. Not performing preventive maintenance on an instrument as required by the manufacturer Preventive maintenance avoids unnecessary and costly repairs. It also ensures, in most cases, prolongation of the life of the instrument or equipment and limits downtime of the instrument.
Chapter 02: Immunology: Basic Principles and Applications in the Blood Bank Exercise 2.1 (Obj 3) Identify the following parts of an IgG molecule using the diagram. 1.
Heavy and light chains
2.
Fab - Antigen-binding site
3.
Paratope
4.
Complement-binding region
5.
Macrophage-binding site
6.
Variable region