Chromatography Column Column Flow Reversal Kit Installation and Instruction Guide
DELTA PRECISION LTD
Chromatography Column
Column Flow Reversal Kit Installation and Instruction Guide
Revision History Revision A B C D
Comments Initial release Labels added to Photos Updated address Updated to reflect Evolve® Trademark status
Date 28/1/2016 30/1/2016 21/06/2018 12/07/2021
Evolve® is a registered trademark of Delta Precision Ltd. Evolve® is a registered trademark in the EU, UK and US.
Column Flow Reversal Kit This kit enables the column flow to be directed downwards or upwards through the column. Reversed flow is useful to remove air in the system or during cleaning procedures. There are two kits AACFR1 for Evolve® 70mm and 100mm columns and AACFR2 for Evolve® 140mm, 200mm and 250mm columns. Kit Contents
COLUMN REVERSAL KIT
1 1 2 2 3 3 4 4 5 6 6 7
DESCRIPTION Flexible hoses length 1.2m, 3mm od Flexible hoses length 1.2m 6mm od Flexible hoses length 0.75m, 3mm od Flexible hoses length 0.75m 6mm od Valve Manifold (small) Valve Manifold (large) 4 port 2 way valve small 4 port 2 way valve large Blanking Cap Ladish Seal 6mm id Ladish Seal 3mm id Clamps
QTY 1 1 2 2 1 1 2 2 2 10 10 10
AACFR2
Number as shown in Figure 1
AACFR1
PRODUCT NUMBER
X X X X X X X X X X
X X X X
Figure 1 Contents of Column Flow Reversal Kit
All materials in contact with the process stream have been selected for their suitability for use in equipment utilised in Biopharmaceutical applications and so either conform to relevant sections of FDA code of Federal Regulations vol. 21 170-199 and/or passed USP class VI test for in vivo toxicity. Wetted Components
Material
Flexible hoses (3mm id and 6mm id) all lengths 4 port 2 way valve small 4 port 2 way valve large Ladish TC Seals Blanking Cap Clamps
Braided PVC hose/ Hytrel ends PVDF valve Body/PTFE Ball/EPDM seals PVDF valve Body/PTFE Ball/EPDM seals Santoprene® Polypropylene Nylon
3
Non Wetted Components Valve manifold (small) Valve manifold (large) Bolts/Rod
Figure 1 Kit contents
Material Acetal Acetal Stainless steel type 316
AACFR1
AACFR2
Installation and Operation
1. Set-up the contents of the kit as shown in the schematic below Downwards Flow
Process In
COLUMN
Column Out
Figure 2a Flow schematic showing set-up of column reversal kit for Downwards flow and photograph showing orientation of 4 port 2way valve. 2. Flexible pipework from and return to the chromatography system is not provided in this kit.
4
3. Before operation the pipework has to be primed with fluid and all air expelled. This is achieved by disconnecting the flexible pipework that is connected to the column and switching on the pumping system. When all air has been expelled from the pipework reconnect to the column. 4. Allow flow to travel through the column and pipework and valves connected to the column bottom and this will expel all air. Upwards Flow Process In
COLUMN
Column Out
Figure 2b Flow schematic showing set-up of column reversal kit for Upwards flow and photographs showing orientation of 4 port 2way valve.
5