LC_Catalog

Page 300

Biosolutions and Columns for Biologic Characterization

Exceptional Separating Power Column:

Buffer: Flow Rate: Gradient: Detector:

Agilent Bio SCX, stainless steel 5190-2423 4.6 x 50 mm, 3 μm 10 mM Phosphate, pH 6.0 0.5 mL/min 0-1.0 M NaCl, 15 min 280 nm

The hydrophilic, polymeric layer and densely packed ion exchange functional groups provide extremely sharp peak shapes and high resolution of a mixture of proteins with a broad range of isoelectric points (pI).

Resolving Ovalbumin and BSA using Agilent Bio SAX Column:

Agilent Bio SAX, stainless steel 5190-2463 4.6 x 50 mm, 3 μm

Buffer: Flow Rate: Gradient: Backpressure: Detector:

20 mM Tris, pH 8.0 0.5 mL/min 0-0.3 M NaCl (15 min) 1,600 psi 214 nm

Isoforms and impurities of both ovalbumin and BSA can easily be resolved when an Agilent Bio SAX NP3 (3 μm particle) column is used.

1008

LC AND LC/MS

www.agilent.com/chem/lccolumns

mAU

2

1

5

1. Ovalbumin, 4.6 pl 4

40

2. Ribonuclease A, 8.7 pl 3. Cytochrome C, 9.6 pl

3 30

4. Aprotinin, 10.0 pl

20

N > 100,000/50 mm for Lysozyme

5. Lysozyme, 11.0 pl

10

0

2

4

6

8

10

12 min


Issuu converts static files into: digital portfolios, online yearbooks, online catalogs, digital photo albums and more. Sign up and create your flipbook.