Isothermal Battery Calorimeter Product Range

Small cell (IBC & IBC-CC) • Coin AA 18650 D • Cell size specific

Prismatic pouch EV cells (IBC-P) • Prismatics

Isothermal Air Calorimeter (IAC) • Complete devices • Lithium ion cells or modules

Offices in ENGLAND, USA, CHINA, INDIA Representation Worldwide


IBC Isothermal Battery Calorimeter Product Range The IBC Isothermal Battery Calorimeter is available in several variants to accommodate different battery sizes. With connection to an external cycler the IBC, IBC-CC and IBC-P provide accurate and easy testing of heat output from charging and discharging a single battery. Heat occurring during the charging and discharging of batteries is largely due to their internal impedance. Characterisation of this heat at different operating temperatures is vital for efficient thermal management of cells in packs or modules. The IAC is an Isothermal Air Calorimeter for measurement of heat flux from electronic devices, lithium ion cells and modules. It’s large sample chamber measures 275 x 275 x 175mm accommodating a wide variety of sample sizes. Power compensation control allows for accurate measurement of heat flux.

IBC Cell format: Cylindrical

Quantify heat release... • under applied load conditions with connection to external cycler.

IBC-CC Features same as IBC above, but with coin cell chamber, and a reference cell chamber.

Voltage 4.5

500

4

-200

3.5

-900

3

-1600 0

8000

16000

24000

2.5 32000

Elapsed Time (s)

1.5 AH Lithium Battery Charge and Discharge Cycle 1A

Heat Rate

Current

Voltage

8

3.2

4

2.8

0

2.4

-4

2

-8 0

12500

25000

37500

1.6 50000

Elapsed Time (s)

Cycling data from ML2032, rechargeable coin cell, 65mAh capacity

Voltage (V)

Cell format: Coin Cell

Heat Rate / Current (mW/mA)

Measure change of heat release with... • Chemistry modification • Applied load • Size and shape • State of charge • Environmental temp. • Age

Current

Voltage (V)

Study battery... • Stability • Efficiency • Performance • Life cycle

Heat Rate / Current (mW/mA)

Heat Rate 1200


Product Range 2000

13500

0 Heat Battery Current

9000

-2000

4500

-4000

0

IBC-P

0

Cell format: Prismatic (small)

Current (mA)

Heat (mW)

18000

1000

2000

3000

-6000 4000

Time (sec)

• Accommodate battery sizes up to: 100x150mm diameter; depth 20mm

Heat generated along with current data during three minute discharges on a prismatic battery at 20°C. 5.0

18000

Heat (mW)

• Temperature range, tailored to automotive operating requirements (-5 to 60°C)

3.8

13500 Heat

9000

2.5

Battery Voltage

1.3

4500 0 0

1000

2000

3000

Voltage (V)

• Heat measurement down to 100mW

0.0 4000

Time (sec)

Heat generated along with voltage data during three minute discharges on a prismatic battery at 20°C.

IAC- Isothermal Air Calorimeter Cell format: Large Sample • mW sensitivity calorimeter to accommodate complete electronic devices, cells or modules of sizes up to: 275x275mm diameter; depth 175mm

Heat (Watts)

0.4 0.2 0 -0.2 -0.4 0

400

800

1200

1600

Time (sec)

Baseline signal over 25min at 25°C.

• Custom build option offering larger internal volumes built to customer requirements Heat Measured

48 36 24

y = 1.0053x

12

R² = 0.9998

0 0

12

24

36

Heat Input

Calibration Steps of 10W, 20W, and 40W (insert) Calibration linearity (main plot)

48


Specifications Calorimeter Type

IBC Cylindrical

IBC-CC Coin Cell

IBC-P Prismatic Small

IAC Power Compensation Calorimeter

Max Temperature (°C)

80

80

60

60

Min Temperature (°C)

-5

-5

-5

-20

Temperature Stability (°C)

0.002

0.001

0.01

0.01

Refrigerated recirculator required

No

No

No

Yes

Calorimeter chamber diameter (mm)

35 (customisable)

25

100 x 150

275 x 275

Calorimeter chamber depth (mm)

70 (customisable)

5

20

175

Number of sample chambers

1

1

1

1

RMS Noise

100 µW

10 µW

1mW

100mW

Max heat detection

9W

500mW

10W

50W

Current resolution

Defined by cycler

Defined by cycler

Defined by cycler

N/A

Internal wiring (force)

8 ga (3.3 mm)

16 ga (1.3 mm)

4 ga (5.2 mm)

N/A

Internal wiring (sense)

20 ga (0.8 mm)

20 ga (0.8 mm)

20 ga (0.8 mm)

N/A

Heat capacity measurement capability

Yes

Yes

Yes

No

Heat cut out

Yes

No

Yes

No

N2 Purge

Yes

Yes

Yes

Yes

External cycler terminals

Yes

Yes

Yes

No

Footprint (Width x Height x Depth)

10 x 12 x 16 Inch 25 x 30 x 41 cm

10 x 12 x 16 Inch 25 x 30 x 41 cm

10 x 12 x 16 Inch 25 x 30 x 41 cm

24 x 16 x 24 Inch 60 x 40 x 60 cm

Weight

22 lbs 10 kg

22 lbs 10 kg

24 lbs 11 kg

32 lbs 15 kg

specification subject to change

HEAD OFFICE 1 North House • Bond Avenue • Bletchley • MK1 1SW • England. Tel: +44 1908 646800 • Fax: +44 1908 645209 • Email: info@thtuk.com • Web: www.thtuk.com US OFFICE Tel: +1 317 222 1904 Fax: +1 317 660 2092 Email: info@thtusa.com Web: www.thtusa.com

CHINA OFFICE Tel: +86 21 5836 2582 Fax: +86 21 5836 2581 Web: www.thtchina.com

INDIA OFFICE Tel: +91 11 4701 0775 Fax: +91 11 4701 0779 Email: info@thtindia.com Web: www.thtindia.com

IBC Isothermal Battery Calorimeter  

Product Range

Read more
Read more
Similar to
Popular now
Just for you