EMC Warnings Medical Electrical Equipment needs special precautions regarding electromagnetic compatibility (EMC), and needs to be put into service according to the EMC information provided in this manual. Portable and mobile RF communications equipment can affect Medical Electrical Equipment. The Travel Air® should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, run the Travel Air® before treatment begins to verify the device operates normally in the desired configuration.

Guidance and Manufacturer’s Declaration – Electromagnetic Emissions The Travel Air® is intended for use in the electromagnetic environment specified below. The user of the Travel Air® should make sure it is used in such an environment. Emissions Test

Compliance

Electromagnetic Environment Guidance

RF emissions Group 1 CISPR 11

The device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

RF emissions Class B CISPR 11 Harmonic emissions Class A IEC 61000-3-2 Voltage fluctuations/ Complies Flicker emissions IEC 61000-3-3

The device is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies building used for domestic purpose.

EN-40