Chapter 9: Maintenance Home Maintenance Schedule Service Time Interval Pre-use check (see Travel Air® Pre-Use Check section) When setting up; weekly Lubrication of connector O-rings and Phasitron® P5-TA Weekly Drain water trap As needed Replace Time Interval Filters Every 6 months If operating in dirty or dusty environments, replace as needed. Multimeter batteries Every 6 months Phasitron® P5-TA breathing circuit Every 6 months Fuses As needed

Lubrication After the Phasitron® P5-TA breathing circuit has been cleaned and sanitized, be sure to lubricate the connector O-rings. Wash hands thoroughly with soap and water or use an alcohol-based hand sanitizer. quick-connect fitting

1. Coat each connector O-ring fitting with the provided Percussionaire® lubricant Lubetube.

2. The Phasitron® P5-TA breathing circuit is ready to be reassembled for use. See Cleaning and Disinfection section for reassembly instructions.