GE MAC600 CardioSoft Software  Configuration Details   

On MAC600:  The following steps should be completed before the MAC600 is shipped from Numed to give a  preferred standard setup.  1. Fit the SD Card in to the unit.  2. Fit the paper in to the unit.  3. Switch the unit on and press the F4 More key then the F2 Setup key.  4. Enter the default password 1111 (the enter key is in the centre of the scroll keys). This will  take you to the System Setup menu.  5. Select the ECG option.   a. Select Resting ECG Reports. Select the reports as shown below. Select Return when  finished. 

b. Select Lead Groups. Select the options as shown below. Note Extra rhythm lead must  be set to II. Select Return when finished. 

c. Select ECG Acquisition. Set the Pre‐Acquisition to Yes and other options as shown  below. Select Return when finished. 

MAC600 to CardioSoft link info.doc    18/10/2011 


d. Press Esc on the keypad to return to the System Setup Menu.    6. Select Basic System   a. Select Patient Questions. Select the options as shown below. Select Return when  finished. 

b. Select Option Activation. Ensure the DIAG, EXST and TRAN option codes have been  activated indicated by a * next to them. Select Return when finished. 

c. Press Esc on the keypad to return to the System Setup Menu.  7. Select Country Setup  a. Set Date and Time   b. check Notation is on IEC  c. Press Esc on the keypad to return to the System Setup Menu.  8. Select Storage  a. Select the options as shown below. Select Return when finished. 

9. Select PDF Configuration  a. Set Store PDF format to No  b. Select Return to return to the System Setup Menu. 

MAC600 to CardioSoft link info.doc    18/10/2011 


10. Select Transmission  a. Select the options as shown below. Select Return when finished. 

11. Select Save Setup   a. Select To System  12. Select Exit to leave the System Setup menu.        Set up CardioSoft as shown on the next sheet. 

MAC600 to CardioSoft link info.doc    18/10/2011 


GE CardioSoft Software  MAC600 Communication Configuration Details    CardioSoft v6.61 or later is required for communication with the MAC600. An additional component  of CardioSoft called Connectivity Server is also required. This is included on the CardioSoft  installation CD and must be installed separately after CardioSoft is installed.    1. Install CardioSoft v6.61 or later but do not start up program.  2. Plug in the USB dongle supplied with the software – wait for new hardware to install and be  ready for use.  3. On the installation CD, open the Connectivity Server folder.  a. Run the Setup.exe  b. Do not select the Serial/IP Com Port Redirector  c. When the installation is complete, restart if prompted to do so.  4. Open CardioSoft  5. Go to System Configuration, then the Option Code tab.  6. Ensure the serial number of the key (last six digits) is shown under the Internal Serial Number  section and that this number matches the options code sheet supplied with the software (last  six digits are important – first few may differ).  7. Click the Modify button. Enter the activation codes for the options shown on the option code  sheet. NOTE: the RESB code is required to enable the Receive Data from ECG Device option.  8. When all option codes are entered, click the Check button. Option codes that are correct and  accepted should go grey – check any that don’t as these options will not be enabled.  9. Go to the General tab and click the Export button. 

MAC600 to CardioSoft link info.doc    18/10/2011 


a. Enter a PDF printer name of PDF995  b. Tick Use current Final Report Format  c. Tick Generate automatic file name, then:  i. In box 1 select Patient ID  ii. In the box next to Patient ID put a dash (‐)  iii. In box 2 select Date of Test  This will generate a PDF filename of the format Patient ID‐Date and Time of  test. The date and time of test field is essential to prevent filename duplication  errors when reports are attached to the patient records.  iv. In the Export folder box, enter C:\CSData  v. Tick the PDF box under Export after each examination  d. Click OK to exit the export settings screen.  e. On the General tab, click the Printer... button  i. Tick the Color Printouts box.  ii. Click OK to exit the Printer Settings screen.  10. Go to the Connectivity Server tab. 

a. Tick the Delete Successfully Imported Records box  d. Tick the Perform Automatic XML... box  e. In the Inbound Connections section, click the Add button.  iii. Set the Device Type to Serial  i. Set the COM port to the one in use by the serial cable  ii. Set the Baud Rate to 115200  iii. Click the browse button (three dots) next to the Destination Directory then  select the C:\CSData folder 

MAC600 to CardioSoft link info.doc    18/10/2011 


iv. Click OK to save the settings.  f. Click the Restart Service button at the bottom of the window. The circle next to the  button should go red then green after a couple of seconds.  g. Click OK to save the settings.  11. Configure the PDF printer to prevent the created PDFs from being displayed automatically as  follows.   h. Click Start, All Programs, Software995, pdfedit995  i. Click the Autolaunch tab and deselect Automatically display PDF after printing  option.  j. Click OK to save the settings.       

MAC600 to CardioSoft link info.doc    18/10/2011 


Check Communications:  The Connectivity Server runs as a Windows service and will receive ECGs from the MAC600 even if  CardioSoft is not running. There is no indication on the PC that a communication is in progress.  The MAC600 will display a ‘Transferring’ message when communication with the PC is  established.    1. Record and store and ECG on the MAC600.  2. On the MAC600, select F3 File Manager. You may need to press the F4 More button before  you see this option available.  3. Press the F1 Select button. This will move the cursor to the list of stored ECGs.  4. Move the cursor using the up and down keys until the required ECG is selected. Press the  Enter button in the middle of the scroll keys.  5. Press the F3 Transmit button. Establishing network connection. Please Wait will be  displayed, possibly very briefly if comms with the PC are established immediately.  6. After a few seconds, the MAC600 screen will show Transmitting file 1 of 1.  7. If the transfer completes successfully, the ECG file will be marked as Yes in the Sent column,  or will be deleted if the option to delete transferred ECGs is enabled.  8. If using ECG‐Link, start the ECG‐Link software.  a. ECG‐Link should report ECG files have been detected. Answer Yes to process now.  b. A PDF should be shown with a filename matching the ID number of the transferred  ECG followed by the date and time of the test.  c. If this is seen, the MAC600 is communicating correctly with the PC.  d. This completes the communications test.  9. If using CardioSoft without ECG‐Link, open the CardioSoft software and click Local Database.  The transferred ECG should be shown under the Patient ID column. No patient name will be  shown as none can be entered in the MAC600.   10. If the patient name is required to be shown on the PDF report perform the following:  a. Click on the Patient ID under the Local Database list.  b. Click on the transferred ECG, then click the View button.  c. Under the Patient Information section, click the Modify button.  d. Add the patient name and click OK.  e. You will be asked whether the patient data in the local patient list should be updated.  Click Yes. This will then show the patient details in the local database list and any  subsequent ECGs transferred under that patient ID will automatically have the patient  name displayed on the PDF. 

MAC600 to CardioSoft link info.doc    18/10/2011 

GE MAC600 to CardioSoft Configuration Guide  

On MAC600: c. Select ECG Acquisition. Set the Pre‐Acquisition to Yes and other options as shown below. Select Return when finished. b. Selec...