Page 16

Product Characteristics: • Conical implant with Morse Taper interface; • NeoPoros Surface; • The installation is performed with specific drivers to the index of the implant, providing the implant capture directly on packaging, offering more safety and speed; • Double thread (faster installation and less trauma); • Suitable for bones type III and IV and pos-extracting installation; • Cover screw not included; • Cervical diameter equal to implant diameter; • Unique prosthetic interface for all implant diameters; • Observe the specific Pilot Drill (function countersink); • Best prosthetic resolution with CM Implant 2mm infraosseous; • Perforation rotation: 500-800 rpm; • Insertion rotation: 30 rpm.

CM Pilot 4.3/5.0

Alvim Drill 5.0

CM Pilot 3.6/4.3

Alvim Drill 4.3

CM Pilot 2.8/3.5

Alvim Drill 3.5

Alvim Drill 2.0

Initial Drill

Drill Sequence

3.5 mm 4.3 mm 5.0 mm

Code 8.0 mm

10.0 mm

11.5 mm

13.0 mm

CM Cover Screw

16.0 mm

Ø 3.5

109.657

109.658

109.659

109.621

109.660

Ø 4.3

109.647

109.648

109.649

109.622

109.651

Ø 5.0

109.652

109.653

109.654

109.655

109.656

2 mm

117.013

117.017

•Use Digital Driver 1.2mm (104.007) for installation; •Do not exceed torque of 10 N.cm.

Healings

16

0.8 mm

1.5 mm

2.5 mm

3.5 mm

4.5 mm

5.5 mm

Ø 3.3

106.182

106.168

106.169

106.170

106.183

106.184

Ø 4.5

106.175

106.171

106.172

106.173

106.174

106.180

•Use Digital Driver 1.2mm (104.007) for installation; •Do not exceed torque of 10 N.cm.