One lead to much

Page 1

„Uczyć – bawiąc” - czyli przypowiastki z morałem

One lead to much Even one abandoned lead can create several problems

ICD-Train 2013

Andrzej Kutarski Lublin (a_kutatski@yahoo.com)

www.usuwanieelektrod.pl


One lead to much 2008 implantation VVI system for brady-AF 2011 PM (only PM) removal due to Ca mamme and planned mastectomy & radiotherapy 2011 the new VVI system implanted simultaneously Rt side od the chest 2013 severe trisuspid regurgitation and proper referrence for TLE


One lead to much Clinical problem: lead related severe tricuspid valve dysfunction (III/IV degree tricuspid regurgitation)


One lead to much

???

Clinical problem was solved: abandoned lead was removed and tricuspid regurgitation decreased up to II/III

www.usuwanieelektrod.pl a_kutarski@yahoo.com


One lead to much

www.usuwanieelektrod.pl a_kutarski@yahoo.com


One lead to much

Conclusions: 1. Before mastectmy much better is to remove whole system than to abandon the lead 2. Two ventricular leads: may by one to much 3. Dynamic contact between two leads may to cause silicone tube abrsion – future „anchor” of lead dependent endocatditis


Thanks for your attention Masz problem ? wyślij e-mila na:

a_kutarski@yahoo.com www.usuwanieelektrod.pl