Check that the flow of saline appears straight and focused and uniform when it leaves the orifice. If it is deformed in any way Jetox is defective. Close all taps and replace.

1. Check that the spike in the saline bag is inserted correctly and that the other end of the tube is inserted properly into Jetox. Do not force it. 2. Make sure that the saline tubing clamp is open. 3. Check the gas pressure at the main valve. If the pressure is not between 4 - 6 atmospheres (60 - 90 psi) liquid will not come out. If after these two tests only gas comes out Jetox is defective. Close all taps and replace.

Doesn’t clean the wound

When Jetox is turned on only gas comes out.

Oxygen Tube length

Jetox™ User Instructions

2 meter (6’ 7”)

Saline Tube length

4

0.5 gram / 1/2 liter (30.5 cu in) per second

2 meter (6’ 7”)

Gas requirement during treatment

UNF 9/16 thread

1.5 cc / minute

4 - 6 atmospheres (60 - 90 psi)

Working oxygen pressure

Liquid requirements during treatment

Medical Oxygen

Gas

Oxygen connection to oxygen source

Saline 0.9% sodium chloride

Liquid

Technical Specifications

Close the main oxygen supply and make sure that the oxygen connection to Jetox is correct and fully closed. It should be closed firmly but not tightly. Open the main oxygen supply. If gas continues to escape Jetox is defective. Close all taps and replace.

$

Gas escapes when Jetox is closed

#

&

"

Jetox is defective. Close all taps and replace.

Solution

%

&

!

The saline bag inflates or you see bubbles in it

Problem

Troubleshooting

Note: Discard all disposable articles.

1. Close Jetox. 2. Close the main oxygen supply. 3. Open Jetox to release pressure. 4. Close the saline tubing clamp. 5. Disconnect the spike and saline tubing. 6. Unscrew (counterclockwise) the tubing of the oxygen supply.

When Treatment is Finished

Jetox™ User Instructions

Rev 050400 CE

Note: Size, degree of infection, placement, and skill of the therapist effect the length of the treatment.

The average length of Jetox treatment of a simple wound of two square centimeters (threequarters inch square) lasts between two to three minutes.

Jetox™ is designed for effective cleansing and debridement of pressure ulcers, venous stasis ulcers, diabetic ulcers, surgical wounds and burns. Jetox creates a jet stream from sterile saline and compressed medical oxygen. The drops are between 5 to 200 microns in size and are accelerated up to 200 meters (660 feet) per second. The jet stream creates low pressure on the wound.

Introduction

Jetox is designed for one-time use.

Jetox should only be used by a trained health professional who is dressed appropriately.

Read all instructions before using Jetox.

Safety First

Safety First .................................................................................................................................... 1 Introduction................................................................................................................................... 1 Installation..................................................................................................................................... 2 Starting Treatment........................................................................................................................ 2 Using Jetox ................................................................................................................................... 3 When Treatment is Finished........................................................................................................ 4 Troubleshooting ........................................................................................................................... 4 Technical Specifications.............................................................................................................. 4

Patents Pending: IL 118801 IL 120021 PCT 97/0027 Europe 97900411.6 USA 09/194,044

The contents of this manual are considered accurate at the time of printing.

All rights reserved.

© 2000 Tav-Tech Ltd.

Tav-Tech Ltd. POB 194 Katzrin 12900 Israel Tel. + 972-6-696-3547 Fax. + 972-6-696-4060 email: tavtech@canaan.co.il

Jetox™ User Instructions


2

1. Turn on the main oxygen supply. 2. Bring Jetox to the wound such that the outer opening of the shield touches the treated area. 3. Open the Jetox knob completely. 4. Do not push Jetox towards the wound so that the inner lips of the shield touch the wound.

Note: Make sure that the wound is dry before beginning treatment.

Starting Treatment

5. Remove vented cap and connect the spike to the saline bag. 6. Open the saline tubing clamp. 7. Place the shield on the nozzle. Jetox is designed such that the valve controls only the flow of oxygen, and the saline is drawn by the action of the gas. There is no need to hang the saline solution above the level of Jetox. It is best to place the saline bag on the same level (or slightly higher) that Jetox will be working. In this way you will avoid saline dripping through Jetox by gravitation when it is not in use.

Note: The oxygen pressure should be 4 - 6 atmospheres (60 - 90 psi).

1. Make sure that Jetox valve is closed. 2. Screw male oxygen luer to female luer. Do not force. 3. Connect luer slip to female saline connector. 4. Connect oxygen xmas connector to oxygen source (cylinder or hospital oxygen outlet).

!

%

$

&

%

#

!

'

Jetox™ User Instructions

"

#

The optimal angle for work is 90° to the wound. You can also work at an angle of 70° to 80° by taking advantage of the flexibility of the shield.

In order to examine the wound during treatment turn off Jetox and move it aside. The flexibility of the shield allows you to give treatment in the entire area covered with the shield without having to remove Jetox from the wound.

The transparency of the shield allows you to monitor the process of cleaning continually.

On a flat wound you can work horizontally to the wound without a break.

The active area of the cone of jet sprayed saline is 2 millimeters (0.078”) in diameter. The most effective distance for treatment is 4 millimeters (5/32”) between the Jetox opening and treated area. The shield is designed to maintain the optimum treatment distance.

Open both tear-down kits. "

Using Jetox

Installation

3