ⅠAPPLICATIONø The portable measuring microscope is used widely for inspecting  measurement in production of factory and observing research in laboratory.  It is particularly suitable for spot examination of Machine, Paper making,  Print and Textile industry, etc. This product small and exquisite, nice design  and easy operation. It is also suitable to be used in experiment by students. Ⅱ CONFIGURATIONø

Eyepiece

Stopper

Pen lamp 

Main body  Fixing screw Focusing knob

Objective

Fixing device of the  pen lamp

Ⅲ SPECIFICATION: Specification Total magnification  Mechanical   tube  length Objective  magnification Numerical aperture  (N.A.) Working   distance  (mm) Eyepiece  Focusing range  Diameter of viewing  field in eyepiece Reticule of eyepiece Diameter   of   bottom  plane Height of microscope  Weight   of  microscope Illumination  Instrument size

MX151­100 MX151­50  100×   50×             160mm

MX151­40   40×

MX151­20   20×

 10×

  5×

  4×

  2×

 0.25

 0.12

 0.10

 0.05

  7

  18

  18

  18

WF10×with reticule 30mm Ø18mm Scale value 0.1mm Ø63mm 210mm 0.65kg The pen lamp, uses two 7# dry cell batteries 143mm×63mm×210mm

Ⅳ OPERATION: 1. Remove plastic bags, put the fixing device of the pen lamp to twist in the  spiral hole of microscope main body. Then tight the nut of thread end toward  the microscope main body with spanner which is attached in the instrument, to  lock the fixing device of the pen lamp. Insert the pen lamp into the tube of the  pen lamp, tighten the fixing screw with finger. 


2. Put the measuring microscope on the observing object, turn the  focusing knob, until a clearest image of observing object is obtained. 3. If after the focusing, the observing object is already clear ,but the  reticule of eyepiece is not clear. May rotate the diopter adjusting ring of  eyepiece, also until a clearest image of reticule is obtained . 4. When needs to illuminate, open the pen lamp. Loose the fixing screw,  adjusting the pen lamp into the suitable position, causes the light of the  pen lamp to aim the observing object, then tighten the fixing screw with  finger. 5. In order to prevent objective touch the observing object, adjusting the  stopper to the suitable position. Ⅴ MAINTENANCEø 1.The microscope must be placed in where is shady and dry. Don’t let it  expose under sun light directly. 2. The microscope has be calibrated and inspected strictly before leaving  factory, the users must not knock down the instrument discretionally. 3. If there’s dust on the lens, blow it by rubber ball blower, after that clean  the lens gently with a soft brush pen, carefully wipe off oil or fingerprints  on the lens surface with lens tissue or absorbent cotton moistened with a  few organic solvent (mixture of ether and alcohol 7:3). 4. Keep the mechanical parts clean and wipe regularly. 5. Cover the microscope with a dust cover, when the microscope is not  used. If longer time is not used, put best the microscope into the packing  box.

  OPERATION  INSTRUCTION FOR MX151 SERIES PORTABLE  MEASURING MICROSCOPE

     In order to exert performance of this microscope and to ensure the 

safety, please read operation instruction carefully before use. 



MANUAL MICROSCOPIO PORTATIL