USER GUIDE

USER GUIDE Version R0312

ECHO-FLEX # 4800 2 and 3-section stretcher

IBIOM Instruments Ltd. 1065, rue du Pacifique, suite 403, Sherbrooke, Quebec, J1H 2G3 Tel.: 450-678-5468 Fax: 450-445-9837


USER GUIDE

SAFETY INSTRUCTIONS…………………………………………………. LIFETIME LIMITED WARRANTY………....….…………………………… ESTABLISHMENT LICENSE……………………………………............... PRODUCT APPROVALS………………………………………………….. 4.1 CSA approval………………………………………………………... 4.2 Electric components approvals……………………………………. 5 CONTACT INFORMATION…………………………………………………

6

3 4 5 6 6 7 8

OPERATION

1 2 3 4

OPERATION…………………………………………………………………. Motion controls………………………………………………………. To use security rail and cardiac opening…………………………. To use IV pole……………………………………………………….. To use stirrups……………………………………………………….. To use individual locking swivel casters………………………….. To use central and direction lock………………………………….. To use rechargeable battery pack………………………….……… 7 SPECIFICATIONS OF ECHO-FLEX STRETCHER…………………..... 7.1 Leather specifications and cleaning instructions…..……………. 8 STRETCHER MODELS…………………………………………………….

9 9 11 12 12 13 14 14 15 16 17

MAINTENANCE

GENERAL

TABLE OF CONTENTS

9 SUGGESTED MAINTENANCE……………………………………………. 10 TROUBLESHOOTING……………………………………………………… 10.1 Pivots greasing procedure-2 and 3-section stretcher…………… 10.2 Central lock lubrication procedure…………………………………. 10.3 Central lock inspection procedure…………………………………. 10.4 Central lock manual release procedure…………………………… 11 SPARE PARTS LIST……………………………………………………….. 11.1 Two-section stretcher……………………………………………….. 11.2 Two-section stretcher – central lock………………………………. 11.3 Three-section stretcher……………………………………………... 11.4 Three-section stretcher – central lock……………………………..

19 20 22 24 26 30 31 31 33 34 36

6.1 6.2 6.3 6.4 6.5 6.6 6.7

IBIOM Instruments Ltd. 1065, rue du Pacifique, suite 403, Sherbrooke, Quebec, J1H 2G3 Tel.: 450-678-5468 Fax: 450-445-9837


USER GUIDE

1 SAFETY INSTRUCTIONS Please read the following safety instructions carefully before using IBIOM’s products. Every person that works with IBIOM’s products must be provided with the necessary information and must have read this guide before using IBIOM’s products.

WARNING – MAXIMUM CAPACITY NEVER ACTIVATE MOVE UP OR DOWN STRETCHER WITH A WEIGHT EXCEEDING 450 POUNDS OR 205 KG. The IBIOM stretchers were conceived to be regularly used with people that weighs 450 pounds (205 kg) or less. If the patient’s weight is OVER 450 pounds (205 kg), the patient can only lay down on stretcher but stretcher cannot be activated.

WARNING - USAGE On three-section stretcher, NEVER sit on foot section. This may cause the stretcher to flip over.

WARNING - MAINTENANCE Before beginning any maintenance tasks on an IBIOM product, refer yourself to this manual. Always unplug the stretcher before performing maintenance tasks on any electric component (motor, control box, remote control, pedals,…). Make sure to do suggested maintenance in this guide to ensure the stretcher works well over the years.

© Copyright IBIOM Instruments ltd 2012

Page 3 of 36


USER GUIDE

2 LIFETIME LIMITED WARRANTY IBIOM Instruments ltd offers a lifetime limited warranty on its products against defects in material and workmanship while under normal use. The following components are excluded from this warranty: mattresses, electric components, all retractable mechanisms, all locking mechanisms, armrest mechanisms, casters and IV pole. Thus, those exclusions are under warranty for a period of 2 years from the date of purchase. These warranties do not apply to any products and components which must be replaced because of normal wear and tear, negligence, abuse or accident, shipping damage or product use other than in accordance with written instructions and warnings. These warranties do not apply to any equipment or component that were repaired or modified by someone not authorised by IBIOM Instruments ltd. Electric components are under warranty for 3 years from the date of purchase under the same condition as those exposed in the first paragraph of this page. All repairs to equipment with expired warranty will be charged at factory prices.

Š Copyright IBIOM Instruments ltd 2012

Page 4 of 36


USER GUIDE

3 ESTABLISHMENT LICENSE

© Copyright IBIOM Instruments ltd 2012

Page 5 of 36


USER GUIDE

4 PRODUCT APPROVALS 4.1 CSA approval All IBIOM’s products are CSA certified unless otherwise indicated. The CSA approval sticker is located at the bottom of the stretcher in back of the motor.

Two-section stretcher

Serial Number

Serial Number

Three-section stretcher

© Copyright IBIOM Instruments ltd 2012

Page 6 of 36


USER GUIDE

4.2 Electric components approvals

Š Copyright IBIOM Instruments ltd 2012

Page 7 of 36


USER GUIDE

5 CONTACT INFORMATION Before contacting us, please be sure you can provided us with the following information:  Serial number of the stretcher (as indicated in section 4 of this guide)  Model number of the stretcher (as described in section 8 of this guide)

To reach us:

@

Telephone

450-678-5468

Fax

450-445-9837

E-mail

info@ibiom.com

Website

www.ibiom.com

Address

IBIOM Intruments Ltd. 1065 du Pacifique St, Suite 403 Sherbrooke, Québec J1H 2G3 Canada

© Copyright IBIOM Instruments ltd 2012

Page 8 of 36


USER GUIDE

6 OPERATION 6.1 Motion controls Standard hand control A

B

C

D

Foot control

D

A

C

Button A B C D

Function Tilt up back rest section Tilt down back rest section Move up stretcher Move down stretcher

B

Button A B C D

Function Move up stretcher Move down stretcher Tilt up back rest section Tilt down back rest section

Standard hand control – 3 section stretcher A C

D F

E

Button A B C D E F

Š Copyright IBIOM Instruments ltd 2012

B

Function Tilt up back rest section Tilt down back rest section Move up stretcher Move down stretcher Tilt up foot section Tilt down foot section

Page 9 of 36


USER GUIDE

Trendelenburg hand control A

B

C

D

E

Buttons

C

A

Button A B C D E F

D

F

Function Tilt up back rest section Tilt down back rest section Move up stretcher Move down stretcher Tilt up Trendelenburg Tilt down Trendelenburg

Button A B C D E F

B

E

F

Function Tilt up back rest section Tilt down back rest section Move up stretcher Move down stretcher Tilt up foot section Tilt down foot section

To avoid damaging cable on hand control, use clip located under fowler back mattress on both side of the stretcher.

Š Copyright IBIOM Instruments ltd 2012

Page 10 of 36


USER GUIDE

6.2 To use security rail and cardiac opening

To open cardiac cutout, pull red cable.

To close cardiac cutout, lift up cutout and it will engage automatically.

To lower security rail, pull red cable with one hand and hold security rail with other hand.

To move up security rail, lift up rail and it will engage automatically.

Cardiac cutout opened. Security rail lowered down.

© Copyright IBIOM Instruments ltd 2012

Cardiac cutout opened. Security rail lowered down.

Page 11 of 36


USER GUIDE

6.3 To use IV pole 

To remove IV pole, simply remove from hanger.

6.4 To use stirrups

2 3

1

Š Copyright IBIOM Instruments ltd 2012

1- To remove stirrups, simply remove from hanger. 2- Stirrups can be moved forward or backward. Simply unscrew knob underneath stirrup and move it to desired position then tighten knob. 3- Stirrups can be oriented at different angle. Simply move up stirrup, turn to desired angle and move it back in hanger.

Page 12 of 36


USER GUIDE

6.5 To use individual locking swivel casters Use lever to lock or unlock every wheel of the stretcher.

Š Copyright IBIOM Instruments ltd 2012

Page 13 of 36


USER GUIDE

6.6 To use central and direction lock

Directional wheel

Central and direction lock can be activated by pressing down one of the 2 pedals located on both side of the stretcher. Pedal can be moved to one of three positions: directional (green pedal), neutral or total lock (red pedal). DIRECTION (GREEN)

NEUTRE

LOCK (RED)

3 free wheels 1 directional wheel

4 free wheels

4 locked wheels

6.7 To use rechargeable battery pack 

Battery pack comes fully charged on delivery of the stretcher.

After every day of use, make sure to charge battery for a period of 8 hours. To charge battery, simply plug electric cable into wall outlet.

If battery is not used every day, charge battery as required.

© Copyright IBIOM Instruments ltd 2012

Page 14 of 36


USER GUIDE

7 SPECIFICATIONS OF ECHO-FLEX STRETCHER Dimensions: -

Width: Standard 27 in. (69 cm.), but width can be varied from 23 in. to 30 in. (58 cm. to 76 cm.); Length: Standard 76 in. (193 cm.). Other lengths are also available; Adjustable Height: 21 to 39 in. (53 cm. to 99 cm.); operated by a silent motor; Fowler Back operated by a silent motor: Adjustable from 0 to 60 degrees on 2section stretchers and from 0 to 80 degrees on 3-section stretchers; Trendelenburg option available from 0 to -12 degrees.

Capacity: -

Maximum capacity: 450 lb. (205 kg.)

Security: -

Security rails ergo design; tucks away underneath with single hand operation. Central lock allows the user to lock all 4 wheels at the same time.

Mattress: -

3 in. (7.6 cm.) or 4 in. (10 cm.) Semi-rigid polyurethane foam; Curves in the mattress allow the user to come closer to the patient without having to bend or stretch his/her arm; Covered with a waterproof, heat-conducting synthetic material (blue).

Movements: -

Controls: hand and/or foot control device.

Wheels: -

5 in. (12.7 cm.) individual locking swivel casters. 5 in. (12.7 cm.) with central and direction lock (option).

Advantages and other features: -

The clear/open space at the base of the stretcher lets the user move freely and easily. The patient can easily get on to the EchoFlex stretcher without using a footstool. Handles on each side of the back section make it easier to manoeuvre the stretcher. Jointly developed with the School of Industrial Design, faculty of Development, University of Montreal. All motors are CE and CSA certified (120 VAC, 60Hz). Made and assembled in Canada.

Š Copyright IBIOM Instruments ltd 2012

Page 15 of 36


USER GUIDE

7.1 Leather specifications and cleaning instructions

Clean with disinfectant products containing sodium hypochlorite (common household bleach) diluted 1:10.

Š Copyright IBIOM Instruments ltd 2012

Page 16 of 36


USER GUIDE

8 STRETCHER MODELS Here is the list of the different stretcher models manufactured and sold by IBIOM Instruments Ltd. All models can be identified by their own number. Drawings are intended only as a schematic representation of the stretcher. Actual models can vary from these drawings. Two-section stretchers are represented on this page. 4800-DG-26

4800-DC-26

Two-section stretcher for general examination, with right side curve.

Two-section stretcher for general examination, with two curves, one on each side.

Two-section stretcher for cardiac and general examination with right side curve and 10"x10" cutout, backrest 27" long.

27"

3"

3"

10x10"

3"

4800-DC-20

4800-DY-26

4800-DCY-26

Two-section stretcher for cardiac and general examination with right side curve and 10"x10" cutout, backrest 21" long.

Two-section stretcher for gynaecological examination with right side curve.

Two-section stretcher for gynaecological examination with right side curve and 10"x10" opening.

27"

Š Copyright IBIOM Instruments ltd 2012

Page 17 of 36

10x10"

3"

LEFT SAFETY SIDE

RIGHT SAFETY SIDE

27" 48"

3"

LEFT SAFETY SIDE

RIGHT SAFETY SIDE

3"

48"

10x10"

LEFT SAFETY SIDE

RIGHT SAFETY SIDE

27"

21" 54"

27"

27"

LEFT SAFETY SIDE

RIGHT SAFETY SIDE

48"

3"

RIGHT SAFETY SIDE

27"

27"

27"

LEFT SAFETY SIDE

RIGHT SAFETY SIDE

48"

LEFT SAFETY SIDE

RIGHT SAFETY SIDE

27" 48"

27"

27"

27"

48"

Two-section stretcher for general examination, without curve.

4800-DO-26

LEFT SAFETY SIDE

4800-OO-26


USER GUIDE Three-section stretchers are represented on this page. 4800-GY-26

4800-GY-26-1S

4800-GY-26-2S

Three-section stretcher for general and gynaecological examination, without curve.

Three-section stretcher for general and gynaecological examination, with 1 small curve on right side.

Three-section stretcher for general and gynaecological examination, with 2 small curves, one on each side.

27"

LEFT SAFETY SIDE

RIGHT SAFETY SIDE

19"

20"

RIGHT SAFETY SIDE

LEFT SAFETY SIDE

31" 20" 19"

RIGHT SAFETY SIDE

LEFT SAFETY SIDE

31" 20" 19"

27"

31"

27"

4800-GY-26-1L

4800-GY-26-2L

4800-GYC-26

Three-section stretcher for general and gynaecological examination, with 1 large curve on right side.

Three-section stretcher for general and gynaecological examination, with 2 large curves, one on each side.

Three-section stretcher for general, gynaecological and cardiac examination, without curve.

27"

27"

31"

31"

31"

27"

Š Copyright IBIOM Instruments ltd 2012

Page 18 of 36

LEFT SAFETY SIDE

RIGHT SAFETY SIDE

20" 19"

LEFT SAFETY SIDE

RIGHT SAFETY SIDE

20" 19"

LEFT SAFETY SIDE

RIGHT SAFETY SIDE

19"

20"

10x10"


USER GUIDE

9

SUGGESTED MAINTENANCE

We recommend performing simple maintenance tasks to prevent problems on a long term basis. Before performing any maintenance task, read safety instructions at the beginning of this guide. Task

What to do

Cleaning the stretcher.

IBIOM stretcher must be cleaned regularly to remove dust and dirt. All mobile parts must be inspected to detect mechanical damage, wear and breaks. For example, any cracks, missing components, missing fasteners or any other visible damage that can affect the stretcher performance. Central lock must be checked and lubricated if needed. Refer to section 10.2 of this guide for lubrication procedure. Electric components must be inspected at attachment points, cables, wires and plugs for wear and breaks. Visually check mattresses to detect any visible wear, cuts.  Move stretcher completely up and down to detect any problem (noise, visible damage, loose component…).  Move back rest section completely up and down to detect any problem.  Move foot section completely up and down to detect any problem.  Open and close all mobile parts as safety sides and cardiac cutout to detect any problem. If noise can be heard when moving up or down stretcher, pivots must be greased. If so, refer to next section of this guide for greasing procedure.

Inspection of all mobile parts.

Inspection of central lock mechanism. Inspection of all electrical cables. Inspection of mattresses.

Operational test.

Greasing stretcher's pivots.

Weekly

Frequency SemiAnnually annually

X X

X X X

X

Only if stretcher is noisy

NOT DOING MAINTENANCE TASKS MAY CAUSE PROBLEMS OR EVEN INJURIES TO USERS.

© Copyright IBIOM Instruments ltd 2012

Page 19 of 36


USER GUIDE

10 TROUBLESHOOTING Problem Hand/foot switch does not work properly.

One of hand/foot switch button does not work properly.

Cause

Check

Correction

Hand/foot switch may be damaged.

Make sure hand/foot switch is connected in control box.

Hand/foot switch must be replaced. See spare parts list to find spare part number and contact us.

Hand/foot switch may be damaged.

Control box damaged.

may

be

Make sure electric cable of stretcher is connected in wall outlet. Make sure hand/foot switch is connected in control box. Make sure electric cable of stretcher is connected in wall outlet. Make sure all cables are connected properly (control box, electrical outlet, motors cables) Invert motor cables in control box and try hand/foot switch again.

Stretcher makes noises while moving up or down.

Lack of grease stretcher's pivots.

Š Copyright IBIOM Instruments ltd 2012

in

Hand/foot switch must be replaced. See spare parts list to find part number and contact us.

N/A

Page 20 of 36

If problem remains, control box is working properly. Hand/foot switch must be replaced. See spare parts list to find spare part number and contact us. If problem has changed, control box is damaged and must be replaced. See spare parts list to find part number and contact us. Stretcher's pivots must be greased. Refer to next section of this guide for greasing procedure.


USER GUIDE Problem

Cause

Check

Central lock mechanism is noisy.

Lack of oil on mobile parts of central lock mechanism.

N/A

Pedal of central lock do not move.

Mobile part mechanism loosened.

Check all mobile parts and bolting for loose or damaged components.

of has

Broken parts in central lock mechanism. Broken caster. Pedal of central lock move, but does not lock casters.

Part is broken, unaligned or loosened.

Š Copyright IBIOM Instruments ltd 2012

Check all mobile parts and bolting for loose or damaged components.

Page 21 of 36

Correction Mobile parts of mechanism must be lubricated. Refer to section 10.2 of this guide for lubrication procedure. All parts must be checked. Refer to section 10.3 of this guide for inspection procedure or section 10.4 to manually unlock casters. Refer to section 11 of this guide for spare parts. All parts must be checked. Refer to section 10.3 of this guide for inspection procedure or section 10.4 to manually unlock casters. Refer to section 11of this guide for spare parts.


USER GUIDE

10.1 Pivots greasing procedure-2 and 3-section stretcher

WARNING – FOR YOUR SAFETY Secure both ends of the stretcher prior to perform greasing procedure. It is very important to secure the stretcher to prevent it from falling while greasing bolts.

1. Unscrew all 8 Allen set screws with a 1/8" key. Set screws are located underneath steel bushings.

Note: You will find 2 set screws for long pivot bolt and 1 set screw for short pivot bolt. You will find 2 long pivot bolts and 4 short pivot bolts for a total of 6 bolts to be greased.

Š Copyright IBIOM Instruments ltd 2012

Page 22 of 36


USER GUIDE 2. Remove the first long pivot bolt (½"x12"). To remove it, first remove white caps, unscrew nut and then remove bolt.

WARNING – FOR YOUR SAFETY For your safety, remove only one bolt at a time to prevent the stretcher from becoming unstable.

3. Use a white type grease to grease pivot bolt on all its length. 4. Assemble bolt in reverse order (bolt, washer and nut, caps). 5. Repeat steps 3 to 5 for every pivot bolts (total of 6). 6. Once every bolt has been greased, move stretcher up and down a few time. If problem remains, contact us.

© Copyright IBIOM Instruments ltd 2012

Page 23 of 36


USER GUIDE

10.2 Central lock lubrication procedure

WARNING – FOR YOUR SAFETY -

For your safety, do not proceed alone. Make sure no there are no free moving objects on the stretcher before proceeding. Make sure to cover mattresses to avoid damages.

1. Move stretcher to about 30 inches from the ground. 2. Flip stretcher so it lays on mattresses. DO NOT PROCEED ALONE. Fix all mobile parts (security rail for example) before flipping stretcher. 3. Remove all 4 white caps located on both ends of the stretcher.

4x

Š Copyright IBIOM Instruments ltd 2012

Page 24 of 36


USER GUIDE 4. Use a spray penetrating lubricant (Prolab PL-100 for example) to lubricate all mobiles parts in all 4 openings of the base of the stretcher. Once lubricated, move pedal a few times. Put back white caps on holes. IMPORTANT: DO NOT USE WD-40 LUBRICANT.

Mobile parts. 5. Lubricate all mobile parts in the openings underneath the two pedals. Use the same type of lubricant as for step 4. Once lubricated, move pedal a few times.

Š Copyright IBIOM Instruments ltd 2012

Page 25 of 36


USER GUIDE

10.3 Central lock inspection procedure

WARNING – FOR YOUR SAFETY -

For your safety, do not proceed alone. Make sure no there are no free moving objects on the stretcher before proceeding. Make sure to cover mattresses to avoid damages.

1. Move stretcher to about 30 inches from the ground. 2. Flip stretcher so it lays on mattresses. DO NOT PROCEED ALONE. Fix all mobile parts (security rail for example) before flipping stretcher. 3. Remove all 4 white caps located on both ends of the stretcher.

4x

Š Copyright IBIOM Instruments ltd 2012

Page 26 of 36


USER GUIDE 4. While moving pedal, check all mobile parts in all 4 openings of the base. Check for loosened or damaged parts. Check if all set screws are tightened. There is one set screw by opening.

Mobile parts. 5. While moving pedal, check all mobile parts in the openings underneath the two pedals. Check for loosened or damaged parts. Check if all set screws are tightened. There is one set screw by opening.

6. Check if set screws holding pedal are tightened. There is one set screw by pedal.

Š Copyright IBIOM Instruments ltd 2012

Page 27 of 36


USER GUIDE

7. If damaged components are found, refer to spare parts section of this guide and contact us. 8. If everything is normal, check if casters are working properly. a. Remove all 4 screws holding hexagonal rods. b. Move out hexagonal rods and make sure they remain engaged in 2 casters on one side.

Hexagonal rods

Š Copyright IBIOM Instruments ltd 2012

Page 28 of 36


USER GUIDE c. While moving pedal on that side, check if casters work properly (lock, neutral and direction). d. Repeat those 3 steps with the opposite side to check the 2 remaining casters.

Hexagonal rods

e. If one of the caster is damaged, refer to the spare parts section of this guide and contact us.

Š Copyright IBIOM Instruments ltd 2012

Page 29 of 36


USER GUIDE

10.4 Central lock manual release procedure By turning clockwise, with a screw driver, screws located at the base of the stretcher, just below casters, it is possible to release central lock. 2-section stretcher

3-section stretcher

Š Copyright IBIOM Instruments ltd 2012

Page 30 of 36


USER GUIDE

11 SPARE PARTS LIST 11.1 Two-section stretcher

© Copyright IBIOM Instruments ltd 2012

Page 31 of 36


USER GUIDE PART # 4500-100 4500-101 4500-102 4500-103 4500-104 4500-105 4500-106 4500-107 4500-108 4500-109 4500-110 4500-111 4500-112 4500-113 4500-114 4500-115 4500-116 4500-117 4500-118 4500-119 4500-120 4500-121 4500-122 4500-123 4500-124 4500-125 4500-126 4500-127 4500-128 4500-129

DESCRIPTION Paper dispenser Fowler back electric motor Linak Up/down electric motor Linak Standard control box Hand switch Foot switch Junction box Power cord Locking swivel caster Fowler back mattress (21" long) 1 Fowler back mattress (27" long) 1 Handle for 21" fowler back mattress Handle for 27" fowler back mattress 48" long mattress for OO-26 stretcher 1 48" long mattress for DO-26 stretcher 1 48" long mattress for DG-26 stretcher 1 48" long mattress for DC-26 stretcher 1 54" long mattress for DC-20 stretcher 1 48" long mattress for DY-26 stretcher 1 48" long mattress for DCY-26 stretcher 1 Cardiac cutout cushion Hanger for IV pole IV pole Security rail Security rail cable assembly Cardiac cutout cable assembly Battery Trendelenburg motor Trendelenburg hand switch Trendelenburg control box

1

IMPORTANT: Refer yourself to section 8 of this guide to choose proper mattress configuration.

Š Copyright IBIOM Instruments ltd 2012

Page 32 of 36


USER GUIDE

11.2 Two-section stretcher – central lock

# PIÈCE 4800-300 4800-301 4800-302 4800-303 4800-304 4800-305 4800-306 4800-307 4800-308 4800-309 4800-310 4800-311

DESCRIPTION Total lock caster Directional and total lock caster Plastic cap Bolt M6x16 mm Transmission rod (kit) Hexagonal rod – wheels Retaining screw Right pedal Left pedal Thick spacer Hexagonal rod – pedal Thin spacer

© Copyright IBIOM Instruments ltd 2012

Page 33 of 36


USER GUIDE

11.3 Three-section stretcher

Š Copyright IBIOM Instruments ltd 2012

Page 34 of 36


USER GUIDE PART # 4500-200 4500-201 4500-202 4500-203 4500-204 4500-205 4500-206 4500-207 4500-208 4500-209 4500-210 4500-211 4500-212 4500-213 4500-214 4500-215 4500-216 4500-217 4500-218 4500-219 4500-220 4500-221 4500-222 4500-223 4500-224 4500-225 4500-226 4500-227 4500-228 4500-229 4500-230 4500-231

DESCRIPTION Paper dispenser Fowler back electric motor Linak Up/down electric motor Linak Standard control box Hand switch Foot switch Junction box Power cord Locking swivel caster Fowler back mattress (31" long) for GY-26 stretcher 1 Fowler back mattress (31" long) for GYC-26 stretcher 1 Handle for 31" fowler back mattress Foot section gas cylinder 19" long foot section mattress 20" long seat section mattress for GY-26 stretcher 1 20" long seat section mattress for GY-26-1S stretcher 1 20" long seat section mattress for GY-26-2S stretcher 1 20" long seat section mattress for GY-26-1L stretcher 1 20" long seat section mattress for GY-26-2L stretcher 1 20" long seat section mattress for GY-26C stretcher 1 Cardiac cutout cushion Hanger for IV pole IV pole Security rail Security rail cable assembly Cardiac cutout cable assembly Battery Trendelenburg motor Trendelenburg hand switch Trendelenburg control box Pair of stirrups Foot section electric motor Linak

1

IMPORTANT: Refer yourself to section 8 of this guide to choose proper mattress configuration.

Š Copyright IBIOM Instruments ltd 2012

Page 35 of 36


USER GUIDE

11.4 Three-section stretcher – central lock

# PIÈCE 4800-400 4800-401 4800-402 4800-403 4800-404 4800-405 4800-406 4800-407 4800-408 4800-409 4800-410 4800-411

DESCRIPTION Total lock caster Directional and total lock caster Plastic cap Bolt M6x16 mm Transmission rod (kit) Hexagonal rod – wheels Retaining screw Right pedal Left pedal Thick spacer Hexagonal rod – pedal Thin spacer

© Copyright IBIOM Instruments ltd 2012

Page 36 of 36